新型自动化Alberta卒中项目早期CT评分(ASPECTS)软件的验证:可靠性、磁共振成像(MRI)相关性及早期缺血核心判别
本研究回顾性纳入三个综合卒中中心的急性缺血性卒中患者,验证新型自动化ASPECTS软件(JLK ASPECTS)与专家阅片及商用软件(RAPID ASPECTS)的评分一致性,并分析其与DWI梗死体积的相关性。结果显示JLK ASPECTS与RAPID ASPECTS及专家的ICC为0.692-0.797,软件间一致性高于专家间(0.797 vs 0.631,P=0.039),尤其发病3小时内(0.751 vs 0.434,P=0.018);自动化ASPECTS与DWI梗死体积显著相关(ρ=-0.532,P<0.001),判别梗死≥70 mL的截断值为5,敏感度88.2%、特异度95.5%。
要点
- JLK ASPECTS与RAPID ASPECTS及专家的ICC为0.692-0.797,软件间一致性高于专家间。
- 发病3小时内软件间一致性0.751 vs 专家间0.434(P=0.018)。
- 自动化ASPECTS≤5判别DWI梗死≥70 mL的敏感度88.2%、特异度95.5%。
核心结论
- 新型自动化ASPECTS软件在早期时间窗内软件间一致性高于专家间,且自动化ASPECTS与DWI梗死体积相关。
研究设计与样本
- 回顾性研究,纳入三个综合卒中中心2022年6月至2023年12月的急性缺血性卒中患者;一致性分析纳入74例前循环大血管闭塞并成功接受血管内取栓、24小时内行随访DWI的患者,相关性分析纳入306例在NCCT后30分钟内行DWI的急性缺血性卒中患者(不论是否行血管内取栓);参考标准为专家阅片、RAPID ASPECTS及DWI梗死体积。
诊断标准与关键数值
- JLK ASPECTS与RAPID ASPECTS及专家的ICC范围为0.692-0.797;软件间一致性0.797 vs 专家间0.631(P=0.039);发病3小时内软件间一致性0.751 vs 专家间0.434(P=0.018);自动化ASPECTS与DWI梗死体积相关(ρ=-0.532,P<0.001);判别DWI梗死≥70 mL的截断值为5,敏感度88.2%、特异度95.5%。
临床应用与阅片要点
- 自动化ASPECTS软件可用于早期缺血改变的评分,其与DWI梗死体积相关,尤其发病3小时内软件间一致性高于专家间。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIM: The Alberta Stroke Program Early Computed Tomography Score (ASPECTS) on noncontrast computed tomography (NCCT) is widely used to assess early ischaemic changes, but its consistency is limited by inter-rater variability. We aimed to validate a novel automated software (JLK ASPECTS) by assessing its scoring agreement against expert readers and a validated commercial software (RAPID ASPECTS), and by correlating its scores with diffusion-weighted imaging (DWI) infarct volumes.
MATERIALS AND METHODS: We retrospectively included patients with acute ischaemic stroke (AIS) admitted to three comprehensive stroke centres between June 2022 and December 2023. For agreement analysis, we included 74 patients with anterior circulation large vessel occlusion (LVO) who achieved successful endovascular thrombectomy (EVT) and underwent follow-up DWI within 24 hours, allowing comparison between pretreatment NCCT ASPECTS and posttreatment DWI ASPECTS. For correlation analysis, 306 AIS patients who underwent DWI within 30 minutes of NCCT were included, regardless of EVT. Agreement was assessed using weighted kappa and intraclass correlation coefficient (ICC), and infarct volume correlation was analysed using Spearman's coefficient and receiver operating characteristics (ROC) analysis.
RESULTS: JLK ASPECTS showed good agreement with both RAPID ASPECTS and experts (ICC range: 0.692-0.797), with higher concordance between software than between experts (0.797 vs 0.631; P = 0.039), particularly within 3 hours of onset (0.751 vs 0.434; P = 0.018). Automated ASPECTS correlated significantly with DWI infarct volume (P = -0.532, P < 0.001), effectively identifying infarcts ≥70 mL (cutoff = 5; sensitivity 88.2%, specificity 95.5%).
CONCLUSION: We validated the newly developed automated ASPECTS software, demonstrating higher inter-rater agreement between software than experts in early time windows. Furthermore, automated ASPECTS correlated with infarct volume on DWI.
引用
- Ha HJ, Lee M, Lee BC, et al. Validation of a novel automated Alberta Stroke Program Early Computed Tomography Score (ASPECTS) software: reliability, magnetic resonance imaging (MRI) correlation, and early ischaemic core discrimination. Clin Radiol. 2025. PMID: 41259964. doi:10.1016/j.crad.2025.107138
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。