文献速递神经MR

C-FLAIR:脑MRI中可控伪影抑制的液体衰减反转恢复序列

C-FLAIR: Fluid-attenuated Inversion Recovery with Controlled Artifact Suppression in Brain MRI.
PubMed 文献 · Radiology · 2025-01-01

本前瞻性研究设计了一种对B0和射频场不均匀性稳健的最优控制反转脉冲,并在体模和14名参与者中于3 T下采集FLAIR与C-FLAIR图像。C-FLAIR在体模中几乎完全反转并消除伪影信号,在人体中脑脊液抑制、伪影及病灶可见性与常规FLAIR相同,平均SNR和CNR无显著差异,但比吸收率升高13.8%、时间平均射频升高6.8%。

要点

  • C-FLAIR在B0和射频场不均匀时实现近乎完美反转,消除不完全反转所致伪影信号。
  • C-FLAIR对多发性硬化病灶和脑白质高信号的对比度与常规FLAIR相同。
  • C-FLAIR比吸收率0.033 W/kg较FLAIR 0.029 W/kg升高13.8%,时间平均射频0.78 µT较0.73 µT升高6.8%。

核心结论

  • 采用稳健射频反转的C-FLAIR可实际消除因不完全反转引起的伪影。

研究设计与样本

  • 前瞻性研究,在3 T下于体模和14名参与者(平均年龄36.1岁±11.5,9名男性,2024年10月至2025年8月于单一学术医疗中心入组)中采集FLAIR与C-FLAIR图像,体模设计为呈现B0和射频场强不均匀。

关键影像表现/征象

  • C-FLAIR在B0场不均匀存在时表现出近乎完美的反转,导致伪影信号被去除。
  • C-FLAIR与常规FLAIR对脱髓鞘多发性硬化病灶和脑白质高信号的图像对比度相同。
  • 由放射科医师视觉评估脑脊液信号抑制、伪影存在及多发性硬化病灶和脑白质高信号可见性。

诊断标准与关键数值

  • 平均SNR:FLAIR 26.5±3.5 vs C-FLAIR 26.4±4.7,两序列间无差异证据(Student t检验)。
  • 平均CNR:FLAIR 18.5±2.1 vs C-FLAIR 17.9±2.4,两序列间无差异证据(Student t检验)。
  • C-FLAIR比吸收率0.033 W/kg vs FLAIR 0.029 W/kg,C-FLAIR升高13.8%。
  • C-FLAIR时间平均射频0.78 µT vs FLAIR 0.73 µT,C-FLAIR升高6.8%。

临床应用与阅片要点

  • 在B0和射频场不均匀区域,C-FLAIR可消除不完全反转所致伪影信号,避免伪影模拟或掩盖病变。
  • C-FLAIR对多发性硬化病灶和脑白质高信号的对比度与常规FLAIR相同,SNR和CNR无显著差异。
  • 使用C-FLAIR时需注意其比吸收率和时间平均射频较常规FLAIR升高。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Background The T2-weighted fluid-attenuated inversion recovery (FLAIR) sequence is part of the routine brain MRI protocol. In regions with strong inhomogeneities in the static magnetic (B0) field and/or the radiofrequency (RF) field, inadequate magnetization inversion results in artifacts that may mimic or obscure pathologic features. Purpose To reduce artifacts on FLAIR images by using an optimized inversion pulse that is robust to inhomogeneities in both the B0 and RF fields. Materials and Methods In this prospective study, a FLAIR inversion pulse was designed using optimal control. FLAIR and FLAIR with controlled inversion (C-FLAIR) images were acquired at 3 T in a phantom designed to exhibit strong inhomogeneities in B0 and in 14 participants (mean age, 36.1 years ± 11.5 [SD]; nine male participants) enrolled between October 2024 and August 2025 at a single academic medical center: nine healthy participants, two with relapsing-remitting multiple sclerosis, one with persistent concussion symptoms, and two with asymptomatic white matter hyperintensities. In the phantom, water signal suppression was assessed visually. In human participants, cerebrospinal fluid signal suppression, presence of artifacts, and visibility of multiple sclerosis lesions and white matter hyperintensities were assessed visually by a radiologist. In eight healthy volunteers, mean signal-to-noise ratio (SNR) and mean contrast-to-noise ratio (CNR) were computed for FLAIR and C-FLAIR, with differences between the sequences evaluated using the Student t test. Results C-FLAIR exhibited nearly perfect inversion in the presence of inhomogeneities in the B0 field, resulting in the removal of artifactual signal. The image contrast of demyelinating multiple sclerosis lesions and white matter hyperintensities was identical for the optimized pulse and the conventional pulse. There was no evidence of a difference in mean SNR (26.5 ± 3.5 for FLAIR vs 26.4 ± 4.7 for C-FLAIR) or mean CNR (18.5 ± 2.1 for FLAIR vs 17.9 ± 2.4 for C-FLAIR) between sequences. C-FLAIR had 13.8% higher specific absorption rate (0.033 vs 0.029 W/kg) and 6.8% higher time-averaged RF (0.78 vs 0.73 µT). Conclusion C-FLAIR with robust RF inversion showed practical elimination of artifacts caused by incomplete inversion. © RSNA, 2025 See also the editorial by Casselman and Bowen in this issue.