文献速递乳腺乳腺X线

乳腺植入物患者乳腺X线摄影中自动与手动曝光技术平均腺体剂量比较

Comparison of average glandular dose in mammography for patients with breast implants when using automatic or manual exposure technique.
PubMed 文献 · Eur J Radiol · 2025-09-23

回顾性分析536例乳腺植入物患者的乳腺X线摄影,比较自动曝光控制与手动曝光在标准及Eklund技术下的平均腺体剂量和图像质量;自动曝光在标准体位剂量更高而在植入物移位体位剂量更低,手动曝光在标准体位图像质量更优。

要点

  • 乳腺植入物患者标准体位(CC、MLO)摄影建议使用手动曝光设置。
  • 植入物移位体位(CCID、MLOID)摄影建议使用自动曝光控制并维持足够图像质量。
  • 手动曝光在CC、MLO和MLOID体位图像质量更优,CCID图像质量无显著差异。

核心结论

  • 自动曝光控制在植入物移位体位降低平均腺体剂量,但在标准体位增加平均腺体剂量;手动曝光在CC、MLO和MLOID体位图像质量更优,使用自动曝光时图像质量也处于可接受水平。

研究设计与样本

  • 回顾性分析536例来自国家筛查项目的乳腺植入物患者,使用单一厂商(Hologic Selenia Dimensions)乳腺X线摄影设备,每例患者拍摄8张图像(4张标准CC、MLO和4张Eklund CCID、MLOID);其中35例患者(每人8张图像)接受两种曝光设置(自动曝光控制和手动曝光)的连续两次乳腺X线摄影,由3名有乳腺X线摄影报告经验的放射科医生采用标准化PGMI分类评估图像质量。

诊断标准与关键数值

  • 图像质量采用标准化PGMI分类评估,由3名有经验的放射科医生判定。
  • CCID图像质量在自动曝光与手动曝光之间无显著差异。
  • 在6项质量标准中的4项上,MLOID图像手动曝光优于自动曝光。

临床应用与阅片要点

  • 作者认为乳腺植入物患者的标准体位乳腺X线摄影应采用手动曝光设置。
  • 作者认为植入物移位体位摄影应采用自动曝光控制并维持足够图像质量。
  • 使用自动曝光控制时图像质量也处于可接受水平。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: The aim of this study was to evaluate the average glandular dose (AGD) in mammography screening of women with breast implants and to compare the AGD and image quality obtained with automatic exposure control (AEC) versus manual exposure settings with standard and Eklund techniques (implant displacement).

METHODS: A retrospective analysis of 536 patients with breast implants from a national screening program using mammography equipment from a single vendor (Hologic Selenia Dimensions) was performed. Each patient underwent eight images; four standard (CC, MLO) and four Eklund (CCID, MLOID) projections. Exposure parameters, breast thickness, compression force and AGD were assessed. A subset of 35 patients (eight images each) who underwent two consecutive mammograms with different exposure settings (AEC and manual) underwent image quality assessment using a standardized PGMI classification, which was evaluated by three radiologists experienced in mammography reporting.

RESULTS: AGD was significantly higher with AEC in standard projections (CC, MLO), but lower in Eklund projections (CCID, MLOID) than with manual exposure (p < 0.001). Manual exposure settings resulted in better image quality in standard projections, while no significant differences were observed in CCID images. For four out of six quality criteria, MLOID images were superior to manual settings.

CONCLUSION: AEC leads to lower AGD in implant displacement projections, but increases AGD in standard projections. Manual exposure settings resulted in better image quality for CC, MLO and MLOID projections, while IQ was at an acceptable level even when using AEC. The results emphasize that standard projection mammography in patients with breast implants should be performed with manual exposure settings, while ID projections should be performed with AEC while maintaining sufficient IQ.