活检针喷砂改良在同轴芯针肝活检中的随机试验:超声可见性、疼痛或组织收获量均无显著改善
这项随机三盲临床试验纳入70例计划接受芯针肝活检的成人患者,比较喷砂时间加倍改良针与常规针在超声可见性、疼痛强度和组织收获量上的差异。结果显示两组在超声可见性(P=0.12)、各时间点疼痛强度(P>0.05)和门管区数量(P=0.74)上均无显著差异,改良针未见显著优势。
要点
- 喷砂时间加倍改良的活检针在超声可见性上与常规针无显著差异(P=0.12)。
- 改良针与常规针在各时间点疼痛强度上均无显著差异(P>0.05),两组疼痛评分均低。
- 两组活检标本门管区数量无显著差异(P=0.74),改良针未提高组织收获量。
核心结论
- 改良活检针在超声可见性、患者疼痛和组织充分性方面较常规针无显著改善,针具选择更多取决于临床偏好、操作者熟悉度和成本效益而非性能差异。
研究设计与样本
- 随机三盲临床试验,70例计划接受芯针肝活检的成人患者随机分入常规针组或改良针组,在伊朗马什哈德的Imam Reza医院进行;改良针通过将喷砂时间加倍制造,表面粗糙度和针尖锐度增加(2.74至5.28牛顿);因严重疼痛需静脉镇痛的患者被排除后续分析。
诊断标准与关键数值
- 常规针组33例与改良针组34例完成全部评估,两组年龄(P=0.85)、性别(P=0.38)、活检指征(P=0.50)无显著差异。
- 超声可见性两组比较P=0.12,无显著差异。
- 各时间点疼痛强度(术中、术后10分钟、术后30分钟及出院时)两组比较P>0.05,无显著差异。
- 活检标本门管区数量两组比较P=0.74,无显著差异。
临床应用与阅片要点
- 作者认为活检针的选择是临床偏好、操作者熟悉度和成本效益的问题,而非基于性能的显著差异。
局限性
- 因严重疼痛需静脉镇痛的患者被排除进一步分析。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: Optimizing biopsy needle design could enhance procedural performance and patient comfort during core-needle liver biopsy (CNLB). This study aimed to compare a modified biopsy needle with a conventional one in terms of ultrasound visibility, patient-reported pain intensity, and diagnostic tissue yield.
METHODS: In this randomized, triple-blind clinical trial, 70 adult patients scheduled for CNLB at Imam Reza Hospital (Mashhad, Iran) were enrolled and randomly assigned to either a conventional or modified needle group, both of which were commercially available. The modified needle was manufactured by doubling the sandblasting duration, which increased surface roughness and tip sharpness (2.74 to 5.28 Newtons). Pain intensity was measured using the Visual Analogue Scale (VAS) at four time points: during the procedure, 10 min after, 30 min after, and at discharge. The ultrasound visibility and the number of portal tracts in biopsy specimens were assessed independently. Patients requiring intravenous analgesics due to severe pain were excluded from further analysis.
RESULTS: 33 participants in the conventional biopsy needle group and 34 in the modified biopsy needle group completed all evaluations. There were no significant differences between groups in terms of age (P = 0.85), sex (P = 0.38), biopsy indication (P = 0.50), ultrasound visibility (P = 0.12), pain intensity at any time point (P > 0.05), and the number of portal tracts in biopsy specimens (P = 0.74). Despite the non-significant differences, both groups demonstrated low pain scores.
CONCLUSION: The modified biopsy needle showed no significant improvement over the conventional needle regarding visibility, patient pain, and tissue adequacy. Therefore, the choice between biopsy needles remains a matter of clinical preference, user familiarity, and cost-effectiveness rather than significant differences in performance.
TRIAL REGISTRATION: IRCT20250422065439N1.