基于全自动人工智能的超声心动图左心室质量定量:多模态验证
该研究在218例同时接受超声心动图和CT的患者中,比较AI与专家手动测量的左心室质量,以CT为参考标准。AI混合法一致性最高(ICC 0.76,MAPE 0.16),诊断左心室质量增加的AUC高于专家(混合法0.84 vs 0.77),重测信度系数7.8%和8.1%。
要点
- AI混合法测左心室质量与CT一致性最高(ICC 0.76,MAPE 0.16)。
- AI较专家手动测量减少约20%测量误差,主要来自室间隔和 posterior wall 测量更一致。
- AI重分类34%的相对室壁厚度增加(RWT>0.42)病例。
核心结论
- 与专家评估相比,AI超声心动图在左心室质量评估中提供更优的准确性、诊断性能和可重复性,支持其临床整合以改善心脏影像标准化。
研究设计与样本
- 在218例接受超声心动图和CT的患者中,分析左心室质量、室间隔直径、舒张末期内径、后壁直径和舒张末期容积;左心室质量采用常规线性法和结合舒张末期描记与平均壁厚的新混合法评估;以CT为参考标准,用组内相关系数和平均绝对百分比误差比较AI与专家超声测量;在床旁队列(n=40)中用变异系数评估重测信度。
诊断标准与关键数值
- AI混合法与CT一致性最高(ICC 0.76,MAPE 0.16),线性法一致性较低(ICC 0.49,MAPE 0.44);检测左心室质量增加的AUC:AI线性法0.78 vs 专家线性法0.71,AI混合法0.84 vs 专家混合法0.77;重测信度变异系数:AI两种质量法分别为7.8%和8.1%。
临床应用与阅片要点
- AI超声心动图可减少左心室质量测量误差并提高诊断准确性,混合法较线性法与CT一致性更高;AI重分类34%的RWT>0.42病例,提示手动评估可重复性随操作者技能变化。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Accurate assessment of left ventricular (LV) myocardial mass is critical for guiding treatment decisions. Measurements based on echocardiography are limited by operator variability. Artificial intelligence (AI)-based methods promise improved precision and reproducibility, but they require validation against high-resolution reference standards such as computed tomography (CT).
OBJECTIVES: The aim of this study is to compare the accuracy of AI-based vs expert manual echocardiography LV mass measurements using CT as the reference and to evaluate real-world test-retest reliability.
METHODS: In 218 patients undergoing echocardiography and CT, the authors analyzed LV mass, interventricular septal diameter (IVSD), end-diastolic diameter, posterior wall diameter (PWD), and end-diastolic volume. LV mass was assessed by the conventional linear method as well as a novel hybrid method combining end-diastolic tracings with mean wall thickness. AI- and expert-based (Expert) echocardiographic measurements were compared with CT using intraclass correlation coefficient (ICC) and mean absolute percentage error (MAPE). Increased relative wall thickness (RWT >0.42;+) and abnormal LV mass was evaluated using receiver-operating characteristic analysis (area under the curve [AUC]). Test-retest reliability was assessed in a bedside cohort (n = 40) using coefficient of variation.
RESULTS: AI echocardiography showed strongest agreement with CT for LV mass assessment using the hybrid method (ICC: 0.76; MAPE: 0.16) and lower agreement for the linear method (ICC: 0.49; MAPE: 0.44). Nevertheless, compared with Expert echocardiography, AI reduced measurement error by ∼20%, mainly due to more consistent IVSD and PWD measurements, and reclassified 34% of RWT+ cases. Diagnostic accuracy for detecting increased LV mass was higher with AI echocardiography than with Expert echocardiography (AUC: 0.78 vs 0.71 for linear; 0.84 vs 0.77 for hybrid). Test-retest reliability was highest with AI echocardiography for both mass methods (coefficient of variation: 7.8% and 8.1%), and reproducibility of manual assessment varied with operator skills.
CONCLUSIONS: AI echocardiography provides superior accuracy, diagnostic performance, and reproducibility for LV mass assessment compared with expert evaluation, supporting its clinical integration to improve standardization in cardiac imaging.
引用
- Sveric KM, Platzek I, Jellinghaus S, et al. Fully Automatic AI-Based Quantification of LV Mass in Echocardiography: A Multimodality Validation. JACC Cardiovasc Imaging. 2025. PMID: 41165630. doi:10.1016/j.jcmg.2025.09.014
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。