优化囊性纤维化中的氙-129通气MRI:螺旋成像与翻转角校正

Optimizing Xenon 129 Ventilation MRI in Cystic Fibrosis with Spiral Imaging and Flip-Angle Correction.
PubMed 文献 · Radiol Cardiothorac Imaging · 2025-01-01

这项前瞻性研究在轻度囊性纤维化患者与年龄匹配健康对照中,比较二维螺旋超极化氙-129通气MRI与常规笛卡尔采集,并用翻转角校正替代N4ITK校正。螺旋与笛卡尔采集的VDP无差异且强相关;翻转角校正去除了线圈相关不均匀性并保留生理性重力梯度,缺陷掩膜与读者分割的一致性更高。

要点

  • 螺旋与笛卡尔采集的N4校正VDP无差异(9.3% vs 9.1%,P=.97),r²=0.95。
  • 翻转角校正保留重力梯度(健康人斜率0.026 SNorm/cm),N4校正将其去除(0.002 SNorm/cm)。
  • 翻转角校正图像缺陷掩膜Dice评分较N4校正高17%(0.48 vs 0.41,P=.001)。

核心结论

  • 二维螺旋采集结合翻转角校正与mGLB分析可实现快速氙-129通气MRI,有效减轻不均匀性并保留生理性异质性,在健康人和囊性纤维化患者中均能准确高效量化通气异常。

研究设计与样本

  • 前瞻性研究,最终分析纳入38例轻度囊性纤维化患者(平均年龄16岁±6,20例女性)和25例健康对照(平均年龄18岁±7,13例男性),使用螺旋(约3秒)与笛卡尔(约10秒)两种采集方式行氙-129通气MRI。

关键影像表现/征象

  • 笛卡尔与螺旋采集产生的区域通气图像定性相似。
  • 翻转角校正去除了线圈相关不均匀性,同时保留生理性异质性,包括被N4校正去除的重力梯度。
  • 翻转角校正图像的缺陷掩膜较N4校正图像更好地描绘区域通气不足和通气过度。

诊断标准与关键数值

  • N4校正VDP在笛卡尔采集为9.1%±8.1,螺旋采集为9.3%±8.7,两者无差异(P=.97),强相关(r²=0.95,P<.001),平均差异-0.2%(95%CI:3.6,-3.9)。
  • 健康人重力梯度平均斜率:翻转角校正为0.026 SNorm/cm±0.013,N4校正为0.002 SNorm/cm±0.001,P<.001。
  • N4校正VDP与读者分割VDP的r²=0.94,翻转角校正VDP与读者分割VDP的r²=0.95,均P<.001。
  • 缺陷掩膜平均Dice评分:N4校正为0.41±0.31,翻转角校正为0.48±0.29,翻转角校正高17%(P=.001)。

临床应用与阅片要点

  • 螺旋采集(约3秒)与常规笛卡尔采集(约10秒)的N4校正VDP无差异且强相关,可快速完成通气评估。
  • 翻转角校正可去除线圈相关不均匀性并保留重力梯度等生理性异质性,避免N4校正对生理梯度的去除。
  • 翻转角校正图像的缺陷掩膜与读者分割一致性更高,有助于更准确描绘区域通气不足和通气过度。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Purpose To implement and evaluate two-dimensional spiral hyperpolarized xenon 129 (129Xe) ventilation MRI with flip-angle (FA) correction, as compared with conventional N4ITK (N4) correction, in healthy individuals and those with cystic fibrosis (CF). Materials and Methods In this prospective study, participants with mild CF and age-matched healthy control participants underwent 129Xe ventilation MRI using both rapid spiral (approximately 3 seconds) and conventional Cartesian (approximately 10 seconds) acquisitions. Images were corrected using N4 bias correction, and ventilation defect percentage (VDP) was calculated using median-anchored generalized linear binning (mGLB). Separately, B1 inhomogeneities in spiral images were FA-corrected and analyzed using mGLB. Gravitational gradients in ventilation were quantified from uncorrected and N4- and FA-corrected images in healthy participants. VDP from N4-corrected (VDPN4) and FA-corrected (VDPFA) images were compared between participant groups and against reader-segmented VDP (VDPRS). Statistical analyses included Wilcoxon signed rank test, Pearson correlation, and Bland-Altman analysis. Results The final analysis included 38 participants with CF (mean age, 16 years ± 6 [SD]; 20 female) and 25 healthy controls (mean age, 18 years ± 7; 13 male). Qualitatively, Cartesian and spiral acquisitions produced similar regional ventilation images. There was no evidence of a difference in VDPN4 between acquisition types (Cartesian = 9.1% ± 8.1; spiral = 9.3% ± 8.7; P = .97) with strong correlation (r2 = 0.95; P < .001) and no systemic bias (mean difference, -0.2%; 95% CI: 3.6, -3.9). FA correction removed coil-related inhomogeneities while preserving physiologic heterogeneity, including gravitational gradients that were removed by N4 correction (mean slope in healthy participants: FA-corrected = 0.026 SNorm/cm ± 0.013; N4-corrected = 0.002 SNorm/cm ± 0.001; P < .001). VDPN4 and VDPFA were strongly correlated with VDPRS (r2 = 0.94 and 0.95, respectively; P < .001 for both). Defect masks from FA-corrected images showed better agreement with reader segmentations compared with N4-corrected image-based defect masks (17% higher Dice score from FA-corrected images; mean Dice score: N4-corrected, 0.41 ± 0.31; FA-corrected, 0.48 ± 0.29; P =.001) and better depicted regional hypo- and hyperventilation. Conclusion Two-dimensional spiral acquisition combined with FA correction and mGLB analysis enabled rapid 129Xe ventilation MRI, effectively mitigating inhomogeneities while preserving physiologic heterogeneity. This approach provided accurate and efficient quantification of ventilation abnormalities in both healthy individuals and individuals with CF. Keywords: MRI, Pulmonary, Lung, Xenon, Ventilation Supplemental material is available for this article. © RSNA, 2025.

来源:Radiol Cardiothorac Imaging,查看原文。版权归原作者所有。