文献速递胸部CT

体质指数是否影响人工智能与人类阅片者在低剂量胸部CT中的肺结节检测

Does BMI influence AI and human reader lung nodule detection in low-dose chest CT?
PubMed 文献 · Eur J Radiol · 2025-09-25

研究利用Lifelines队列的低剂量胸部CT,纳入最高BMI组(均值39.8)和最低BMI组(均值18.7)各176人,比较人工智能软件与人类阅片者的结节检测效能;结果显示两组BMI间检测敏感度无显著差异,但人工智能在两组中每扫描假阳性数均高于人类阅片者。

要点

  • 高BMI与低BMI组间,人工智能和人类阅片者的肺结节检测敏感度均无显著差异。
  • 人工智能在高BMI组每扫描假阳性0.30、低BMI组0.55,两组均高于人类阅片者(P<0.001)。
  • 高BMI组人工智能敏感度0.75、人类阅片者0.76;低BMI组分别为0.80和0.84。

核心结论

  • 低剂量胸部CT肺结节检测敏感度在高BMI与低BMI组间无显著差异,人工智能与人类阅片者均如此;人工智能在两组中每扫描假阳性数均高于人类阅片者。

研究设计与样本

  • 研究利用Lifelines队列的胸部低剂量CT扫描,纳入BMI最高1.5%参与者(均值39.8,标准差3.0)和最低1.5%参与者(均值18.7,标准差0.9),两组各176人;由人工智能软件和一名受训人类阅片者进行结节检测,两名胸部放射科医生复核检测差异,分歧由一名专家放射科医生解决。

诊断标准与关键数值

  • 人工智能敏感度:高BMI组0.75(95%CI 0.66-0.82),低BMI组0.80(95%CI 0.72-0.86),P=0.37。
  • 人工智能每扫描假阳性数:高BMI组0.30,低BMI组0.55,P=0.005。
  • 人类阅片者敏感度:高BMI组0.76(95%CI 0.68-0.83),低BMI组0.84(95%CI 0.76-0.89),P=0.17。
  • 人类阅片者每扫描假阳性数:高BMI组0.05,低BMI组0.16,P=0.09。
  • 两组中人工智能每扫描假阳性数均高于人类阅片者(P<0.001)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: Body mass index (BMI) can influence image quality in low dose computed tomography (LDCT) through higher image noise levels. We evaluated whether BMI affects lung nodule detection by artificial intelligence (AI) software and a human reader.

METHOD: The study utilized chest LDCT scans from the Lifelines cohort. We included 1.5 % participants at highest BMI (mean = 39.8, sd = 3.0), and 1.5 % at lowest BMI (mean = 18.7, sd = 0.9). Nodule detection was performed by AI software and by a trained human reader (HR). Two chest radiologists reviewed detection discrepancies, with disagreements resolved by an expert radiologist. Sensitivity and false positives per scan (FP/scan) were compared between BMI groups, for AI versus HR.

RESULTS: There were 176 participants in both groups, with 131 nodules in high BMI, and 136 in low BMI. AI detected 356 nodular findings and HR 251, including 154 nodules found by both. AI's sensitivity was 0.75 (95 % confidence interval 0.66-0.82) in high BMI, and 0.80 (0.72-0.86) in low BMI groups (p = 0.37). FP/scan was 0.30 and 0.55 in high and low BMI, respectively (p = 0.005). HR's sensitivity was 0.76 (0.68-0.83) in high BMI, and 0.84 (0.76-0.89) in low BMI groups (p = 0.17), with FP/scan of 0.05 and 0.16, respectively (p = 0.09). In both BMI groups, AI had more FP/scan than the human reader (p < 0.001).

CONCLUSIONS: Sensitivity for lung nodule detection in LDCT was not significantly different for high versus low BMI, either for AI or human reader. Compared to the human reader, AI had higher FP/scan in both BMI groups.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Sourlos N, van Tuinen M, Sidorenkov G, et al. Does BMI influence AI and human reader lung nodule detection in low-dose chest CT? Eur J Radiol. 2025. PMID: 41033015. doi:10.1016/j.ejrad.2025.112453

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。