成人腹部光子计数CT协议:腹部放射学会光子计数探测器CT新兴技术委员会多机构共识
腹部放射学会光子计数探测器CT新兴技术委员会通过调查问卷、采用共识减一法由九家机构的九名放射科医生投票,就成人腹部光子计数CT的门静脉期、多期主动脉CTA和多期胰腺协议达成20项协议特征的共识,包括推荐70-keV虚拟单能图像用于门静脉期和多期胰腺PCCT的主要判读,以及多期主动脉CTA应查看低虚拟单能级图像。
要点
- 门静脉期和多期胰腺光子计数CT推荐以70-keV虚拟单能图像作为主要判读图像。
- 多期主动脉光子计数CTA应查看低虚拟单能级图像。
- 共识覆盖扫描模式、管电压、图像质量水平、重建核及光谱重建数据归档等20项协议特征。
核心结论
- 该多机构共识由腹部放射学会认可,建立了标准化的腹部光子计数CT协议。
研究设计与样本
- 腹部放射学会光子计数探测器CT新兴技术委员会创建调查问卷,采用共识减一法由来自九家机构的九名放射科医生投票生成共识声明,涵盖门静脉期CT、多期主动脉CTA和多期胰腺CT协议。
诊断标准与关键数值
- 门静脉期和多期胰腺光子计数CT推荐以70-keV虚拟单能图像作为主要判读图像。
- 多期主动脉光子计数CTA应查看低虚拟单能级图像。
- 共识就20项协议特征达成一致,包括扫描模式、管电压、图像质量水平、主要判读虚拟单能级、重建核、发送至PACS的额外虚拟单能或其他光谱重建,以及特殊光谱成像数据集的归档细节。
局限性
- 原文指出标准化腹部光子计数CT协议缺乏,限制了临床采用、机构间可比性和多中心研究。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Photon-counting CT (PCCT) offers new opportunities for abdominal imaging by enabling higher spatial resolution, improved image quality, and reduced radiation exposure compared with conventional energy-integrating detector CT systems. However, standardized PCCT protocols for abdominal imaging are lacking, which limits clinical adoption, comparability across institutions, and multicenter research studies. The Society of Abdominal Radiology (SAR) Photon-Counting Detector CT Emerging Technology Commission created a survey to achieve consensus on common adult abdominal PCCT protocols for FDA-approved PCCT, including portal venous phase CT, multiphase aortic CTA, and multiphase pancreas CT protocols. The survey included the following items for each protocol: scan mode, tube potential, image quality level, primary viewing virtual monoenergetic imaging (VMI) level, reconstruction kernel, additional VMI or other spectral reconstructions sent to PACS, and details of archiving the special spectral imaging dataset. Consensus statements were generated based on the survey and voted on by nine radiologists from nine institutions using the consensus-minus-one method. Consensus was reached for 20 protocol features, including that 70-keV VMI is recommended for primary interpretation of portal venous and multiphase pancreas PCCT and that, for multiphase aortic CTA, low virtual monoenergetic levels should be viewed. This multiinstitutional consensus endorsed by SAR establishes standardized abdominal PCCT protocols.
引用
- Dane B, Ananthakrishnan L, Marin D, et al. Adult Abdominal Photon-Counting CT Protocols: A Multiinstitutional Consensus of the Society of Abdominal Radiology Photon-Counting Detector CT Emerging Technology Commission. AJR Am J Roentgenol. 2025. PMID: 41060190. doi:10.2214/AJR.25.33625
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。