基于多能CT数据训练的深度学习模型在腰椎CT成像中显示更好的金属伪影减少效果
研究纳入93例腰椎植入物患者行多能CT扫描并重建70和100 keV虚拟单能图像,训练三种深度学习金属伪影减少模型(model70、model100、modelmix)。结果显示modelmix在40至140 keV范围内峰值信噪比和结构相似性指数与model70和model100相当或更高,作者推荐在多能CT数据中训练深度学习金属伪影减少模型用于术后腰椎CT扫描。
要点
- 多能CT数据训练的深度学习模型在宽能量范围内图像质量与单能模型相当或更高。
- 在70 keV和100 keV水平,modelmix的衰减校正优于单一能量模型。
- 三种深度学习模型在100 keV水平的椎管内图像噪声低于传统模型金属伪影减少。
核心结论
- 多能CT数据训练的深度学习金属伪影减少模型在更宽能量水平上图像质量更好,推荐将多能CT数据纳入深度学习金属伪影减少模型的训练用于术后腰椎CT扫描。
研究设计与样本
- 纳入93例腰椎植入物患者行多能CT扫描,重建70和100 keV虚拟单能图像,其中10例随机选取40至140 keV水平;训练model70、model100和modelmix三种深度学习金属伪影减少模型,以传统模型金属伪影减少处理的原始图像为参考标准。
诊断标准与关键数值
- 在40至140 keV范围内,modelmix的峰值信噪比和结构相似性指数与model70和model100相当或更高。
- 在70 keV水平modelmix衰减校正优于model100(P<0.010),在100 keV水平优于model70(P<0.010),但与modelMAR在两种水平相当(P>0.050)。
- 三种深度学习模型在100 keV水平的椎管内图像噪声低于modelMAR(P<0.010)。
- modelmix的主观图像质量评分与model70和model100相当或更高。
临床应用与阅片要点
- 作者建议将多能CT数据纳入深度学习金属伪影减少模型的训练,用于术后腰椎CT扫描。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIM: To develop different deep learning-based metal artifact reduction (MAR) models (deep-MAR) based on virtual monochromatic images (VMIs) at both multiple energy levels and single-energy level, and compare their performance under wider energy levels.
MATERIALS AND METHODS: We enrolled 93 patients with lumbar implants who underwent multi-energy CT scans and then reconstructed into VMIs at energy levels of 70 and 100 keV (10 randomly selected cases at levels ranging from 40 to 140 keV). Original images processed by modelMAR were served as the established reference. Deep-MAR models were trained using diverse datasets at energy levels of 70 KeV, 100 KeV, and two levels (model70, model100, and modelmix). Afterwards, original images were processed using three deep-MAR models, and peak signal-to-noise ratio (PSNR) and structural similarity index (SSIM) were calculated. Objective and subjective image qualities of all processed images were also compared.
RESULTS: From 40 to 140 keV, PSNR and SSIM of modelmix were comparable to or higher than those of model70 and model100. For attenuation correction, modelmix performed better than model100 at 70 keV level and model70 at 100 keV level (P<0.010) but comparably to modelMAR at both levels (P>0.050). Meanwhile, image noise in the spinal canal of three deep-MAR models at 100 keV level were lower than these of modelMAR (P<0.010). The scores of subjective image quality for modelmix were comparable to or higher than those of model70 and model100.
CONCLUSION: With better image quality across broader energy levels, multiple energy CT data are recommended to be comprised in the training of deep-MAR model for postoperative lumbar CT scanning.