本研究对34名健康受试者先后进行无牵引与有牵引的腕关节MR成像,评估直接牵引装置的牵开效果、固有结构可见度及佩戴舒适度。牵引后关节间隙平均增宽0.7±0.34 mm(校正P<0.001),桡月、舟头及月头关节软骨可见度改善(均校正P<0.001),固有韧带及三角纤维软骨复合体病变检出趋于改善;21名(62%)受试者无腕部疼痛,13名(38%)腕痛VAS平均增加2.6分,无肩颈部疼痛。作者认为该装置能有效牵开腕关节,可能改善关节软骨可见度及固有韧带和TFCC病变的检出。

要点

  • 牵引后腕关节间隙平均增宽0.7±0.34 mm(校正P<0.001)。
  • 牵引后桡月、舟头及月头关节软骨可见度改善(均校正P<0.001)。
  • 21名(62%)受试者无腕痛,13名(38%)腕痛VAS平均增加2.6分。

核心结论

  • 该直接牵引装置能有效牵开腕关节,可能改善关节软骨可见度以及固有韧带和TFCC病变的检出。

研究设计与样本

  • 研究纳入34名健康受试者,先后进行无牵引和有牵引的腕关节MR成像。
  • 评估指标包括关节间隙宽度、关节软骨可见度、固有韧带及三角纤维软骨复合体(TFCC)可见度和病变检出。
  • 采用10分视觉模拟量表(VAS)问卷评估MRI期间的疼痛。
  • 使用配对样本t检验和边际同质性检验分别比较牵引前后的关节间隙宽度以及关节软骨和固有韧带可见度。

关键影像表现/征象

  • 牵引后关节间隙平均增宽0.7±0.34 mm(校正P<0.001)。
  • 牵引后桡月、舟头及月头关节软骨可见度改善(均校正P<0.001)。
  • 牵引后固有韧带和TFCC病变检出趋于改善,可见度提高。

临床应用与阅片要点

  • 作者认为该直接牵引装置可能改善关节软骨可见度以及固有韧带和TFCC病变的检出。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

AIM: To evaluate the distraction effect, intrinsic structure visibility, and wearability of a novel direct wrist traction device for wrist magnetic resonance (MR) imaging.

MATERIALS AND METHODS: A total of 34 healthy subjects underwent wrist MR imaging initially without traction and then with traction using a direct traction device and were enrolled in this study. Joint space width, articular cartilage visibility, intrinsic ligaments and triangular fibrocartilage complex (TFCC) visibility, and lesion detection were assessed before and after traction. A 10-point visual analogue scale (VAS) questionnaire was used to assess pain during MRI. Paired sample t-test and marginal homogeneity test were used to compare joint space width, articular cartilage and intrinsic ligament visibility before and after traction, respectively.

RESULTS: Mean joint space width was 0.7 ± 0.34 mm wider after traction (adjusted P < 0.001). Articular cartilage visibility improved at the radiolunate, scaphocapitate, and lunocapitate joints after traction (all adjusted P < 0.001). Intrinsic ligaments and TFCC lesion detection tended to improve following traction with improved visibility. Twenty-one (62%) of 34 subjects had no wrist pain during traction while 13 (38%) had an average increase of 2.6 VAS points in wrist pain. No subjects had shoulder or neck pain.

CONCLUSION: This direct traction device effectively distracts the wrist joint, which may potentially improve articular cartilage visibility as well as the detection of intrinsic ligament and TFCC lesions.

来源:Clin Radiol,查看原文。版权归原作者所有。

引用

  1. Zhang M, Yeung DKW, Wu S, et al. Direct traction device for wrist magnetic resonance (MR) imaging. Clin Radiol. 2025. PMID: 40992354. doi:10.1016/j.crad.2025.107050

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。