腹部计算扩散加权成像的软件间一致性:表观扩散系数与图像质量评估
回顾性纳入154例患者,比较扫描仪控制台软件与商用软件生成的腹部计算DWI(b=1500和2000)的ADC值一致性和图像质量。两种软件ADC偏倚和一致性界限分别为-0.2%至0.7%和0.3%至2.1%,ICC为0.991至1.000;部分组织SNR和CNR存在显著差异。作者认为两种软件生成的ADC值和图像质量一致,SNR和CNR差异的临床影响似乎很小。
要点
- 两种软件生成的腹部cDWI的ADC值高度一致(ICC 0.991–1.000),可互换使用。
- b1500时肝脏、脾脏、肾脏SNR有显著差异,b2000时肝脏、胆汁SNR有显著差异。
- b1500时肝脏、胰腺、肾脏CNR有显著差异,b2000时除脾脏外CNR有显著差异。
核心结论
- 两种软件程序生成的cDWI在ADC值上具有强一致性和高可靠性,尽管SNR和CNR存在显著差异,但其临床影响似乎很小。
研究设计与样本
- 回顾性研究,纳入154例患者(男64例,女90例;平均年龄70.1±10.7岁),2023年11月至2024年10月间接受腹部DWI,使用3.0T扫描仪采集b=0和1000 s/mm²,由扫描仪控制台软件和商用软件分别生成b=1500和2000的cDWI,以Bland-Altman分析和ICC评估ADC一致性,配对t检验比较SNR和CNR。
关键影像表现/征象
- 在正常组织中测量ADC和SNR,并以椎旁肌为参考测量正常组织的CNR。
- 两种软件生成的cDWI在ADC值上显示强一致性和高可靠性。
- 部分组织在b1500和b2000时SNR和CNR存在显著差异。
诊断标准与关键数值
- ADC偏倚范围为-0.2%至0.7%,一致性界限范围为0.3%至2.1%。
- 所有组织的ADC值ICC范围为0.991至1.000(p<0.05)。
- b1500时肝脏、脾脏、肾脏SNR有显著差异;b2000时肝脏、胆汁SNR有显著差异。
- b1500时肝脏、胰腺、肾脏CNR有显著差异;b2000时除脾脏外CNR有显著差异。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To investigate the agreement in apparent diffusion coefficient (ADC) values and the image quality of abdominal computed diffusion-weighted imaging (cDWI) generated by two software programs.
METHODS: This retrospective study included 154 patients (64 males, 90 females; mean age: 70.1 ± 10.7 years) who underwent abdominal diffusion-weighted imaging (DWI) between November 2023 and October 2024. DWI was acquired with a b = 0 and 1000 s/mm2 using a 3.0T scanner. cDWIs with b 1500 and 2000 were generated by a scanner console software and a commercially available software program. The ADC and signal-to-noise ratio (SNR) were measured in normal tissues. The contrast-to-noise ratio (CNR) of normal tissues relative to the paraspinal muscle was also measured. Agreement and reliability of ADC value were assessed using Bland-Altman analysis and intraclass correlation coefficients (ICC), respectively. SNR and CNR were compared using paired t-tests.
RESULTS: ADC bias and limits of agreement ranged from - 0.2 to 0.7% and 0.3-2.1%, respectively. ICCs for ADC values ranged within 0.991 and 1.000 (p < 0.05) for all tissues. Significant SNR differences were observed in b1500 (liver, spleen, kidney) and b2000 (liver, bile). Significant CNR differences were noted in b1500 (liver, pancreas, kidney) and b2000 (excluding spleen).
CONCLUSION: There is strong agreement and high reliability in ADC values between the two software programs. Although significant differences in SNR and CNR were observed, their clinical impact appears minimal. These findings indicate that the ADC values and image quality of cDWI are consistent between the two software programs.
引用
- Yoshida T, Urikura A, Nakashima K, et al. Inter-software agreement in abdominal computed diffusion-weighted imaging: apparent diffusion coefficient and image quality assessment. Abdom Radiol (NY). 2025. PMID: 40938381. doi:10.1007/s00261-025-05198-y
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。