儿童高螺距心脏CT血管造影中光子计数探测器CT与能量积分探测器CT的辐射剂量比较分析
该回顾性研究纳入200例儿童患者,比较全光谱光子计数探测器CT与双源能量积分探测器CT行高螺距心脏CT的图像质量与辐射剂量。结果显示光子计数探测器CT在两组年龄中辐射剂量均显著更高(P<0.0001),≤12月龄婴儿中信噪比显著更低(P<0.0001),对比噪声比无显著差异。
要点
- 光子计数探测器CT在≤12月龄和>12月龄患儿中辐射剂量均显著高于能量积分探测器CT(P<0.0001)。
- ≤12月龄患儿中光子计数探测器CT的小圆肌信噪比显著低于能量积分探测器CT(P<0.0001)。
- 两组患儿对比噪声比在≤12月龄(P=0.39)和>12月龄(P=0.17)均无显著差异。
核心结论
- 使用全光谱处理的光子计数探测器CT行高螺距心脏CT,与能量积分探测器CT相比,辐射剂量更高,且婴儿信噪比更低。
研究设计与样本
- 回顾性、经IRB批准的研究,分析2021年1月至2023年10月期间接受高螺距心脏CT的200例儿童患者(<18岁),其中100例使用全光谱技术(「QuantumPlus」)的光子计数探测器CT,100例使用能量积分探测器CT;辐射剂量以CT剂量指数和剂量长度乘积测量,计算信噪比和对比噪声比,图像质量采用5点Likert量表评估,统计方法包括非配对t检验、Shapiro-Wilk检验、Mann-Whitney检验及kappa系数评价观察者间一致性。
诊断标准与关键数值
- ≤12月龄患儿对比噪声比:光子计数探测器CT组31.61 vs 能量积分探测器CT组32.14(P=0.39)。
- >12月龄患儿对比噪声比:光子计数探测器CT组30.07 vs 能量积分探测器CT组25.27(P=0.17)。
- ≤12月龄患儿小圆肌信噪比:光子计数探测器CT组显著低于能量积分探测器CT组(P<0.0001)。
- >12月龄患儿小圆肌信噪比:光子计数探测器CT组与能量积分探测器CT组无显著差异(P=0.89),两扫描仪间信噪比相似。
- 辐射剂量:光子计数探测器CT组在两个年龄组中均显著高于能量积分探测器CT组(P<0.0001)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Previous studies have shown improved image quality in pediatric cardiac imaging using photon-counting detector CT (PCDCT). However, these studies did not evaluate image quality and radiation dose when utilizing the full spectral capabilities of PCDCT scanners. The full spectral capability of PCDCT scanners allows the generation of the entire array of mono-energetic reconstructions, virtual non-contrast (VNC) images, and iodine maps, which have potential advantages in evaluating complex congenital heart disease. For example, following complex congenital cardiac repairs, when distinguishing intraluminal or soft tissue calcifications from contrast, or when evaluating intrastent thrombus.
OBJECTIVE: To compare image quality and radiation dose between high-pitch cardiac CT using full spectral PCDCT and dual-source energy-integrating detector CT (EIDCT).
MATERIALS AND METHODS: This retrospective, IRB-approved study analyzed high-pitch cardiac CTs from January 2021 to October 2023 in pediatric patients (< 18 years). Patients were scanned using either PCDCT with full spectral technique ("QuantumPlus") or EIDCT. Radiation doses were measured by CT dose index (CTDI) and dose-length product (DLP). Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were also calculated, and image quality was assessed using a 5-point Likert scale. Statistical analysis included unpaired T-test, Shapiro-Wilk test, Mann-Whitney test, and kappa coefficients for interrater agreement.
RESULTS: Two hundred patients were evaluated, with 100 scanned on PCDCT and 100 on EIDCT. In the PCDCT scanner, 47/100 (47%) were male and 53/100 were female (53%) (P = 0.01). In the EIDCT scanner, 65/100 were male (65%) and 35/100 (35%) were female (P = 0.01). In the PCDCT scanner, 68/100 (68%) (P = 0.05) were ≤ 12 months, and 32/100 (32%) (P = 0.05) were > 12 months. In the EIDCT, 80/100 (80%) (P = 0.05) were ≤ 12 months, and 20/100 (20%) (P = 0.05) were > 12 months. In patients ≤ 12 months, the CNR were 31.61 in the PCDCT group and 32.14 in the EIDCT group (P = 0.39). For those > 12 months, CNR were 30.07 for PCDCT and 25.27 for EIDCT (P = 0.17). In patients ≤ 12 months, SNR was significantly lower (P < 0.0001) in PCDCT, compared to EIDCT for the teres minor muscles, while in patients > 12 months, SNR was not significantly lower (P = 0.89); SNR was similar between scanners. Radiation doses were significantly higher for PCDCT across both age groups (P < 0.0001).
CONCLUSION: High-pitch cardiac CT with PCDCT using spectral processing resulted in higher radiation doses and lower SNR in infants compared to EIDCT.