文献速递心血管MR

一站式心脏MRI:统一非增强与增强成像的连续采集检查

All-in-One Cardiac MRI: Unified Non-Contrast-enhanced and Contrast-enhanced Imaging for a Continuous-Acquisition Examination.
PubMed 文献 · Radiol Cardiothorac Imaging · 2025-01-01

本研究评估了一种新型连续采集心脏MRI方法,在单次20分钟自由呼吸、非心电门控扫描中整合非增强与增强成像。前瞻性研究纳入10名健康志愿者和4只再灌注损伤猪,多任务T1与T2测量与参考方法高度相关,心肌T1和T2值高于参考方法,ECV无显著差异,猪缺血损伤区T1和ECV升高与参考测量一致。

要点

  • 20分钟自由呼吸非心电门控一站式心脏MRI可同时获得心脏功能、T1、T2、ECV和晚期钆增强。
  • 多任务心肌T1和T2值高于MOLLI和T2-prep FLASH,ECV与MOLLI无显著差异。
  • 猪缺血损伤区增强前T1和ECV升高,与参考测量一致。

核心结论

  • 新型三维多任务方法可实现20分钟一键式心脏MRI检查,无需心电门控或屏气,提供心脏功能、T1、T2、ECV和晚期钆增强测量。

研究设计与样本

  • 前瞻性研究于2021年9月至2022年2月纳入10名健康志愿者(平均年龄37岁±21[SD],男8例女2例)和4只再灌注损伤猪,采用三维堆叠星轨T2准备快速低角度激发脉冲序列联合三维多任务重建与多参数映射框架,参考标准为数字体模和体内参考方法。

关键影像表现/征象

  • 猪缺血损伤区增强前T1和ECV升高,与参考测量一致。

诊断标准与关键数值

  • 体模研究中多任务T1与二维反转恢复自旋回波T1高度相关(R2>0.99),T2与二维自旋回波T2相关(R2>0.960)。
  • 体内多任务心肌T1高于MOLLI(1383毫秒±37.8 vs 1217毫秒±16.3;P<.001)。
  • 体内多任务心肌T2高于T2-prep FLASH(43.9毫秒±1.4 vs 40.0毫秒±1.6;P<.001)。
  • 多任务与MOLLI的ECV值无差异证据(27.5%±1.4 vs 27.9%±1.8;P=.52)。

临床应用与阅片要点

  • 该20分钟自由呼吸非心电门控一站式方案可同时获得心脏功能、T1、T2、ECV和晚期钆增强,用于日常阅片和报告。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Purpose To evaluate a novel continuous-acquisition cardiac MRI method that integrates non-contrast-enhanced and contrast-enhanced imaging within a single 20-minute, free-breathing, non-electrocardiographically (ECG) gated scan. Materials and Methods A three-dimensional (3D) stack-of-stars T2-prepared fast low angle shot (FLASH) pulse sequence, combined with a 3D multitasking reconstruction and multiparametric mapping framework, was developed to jointly reconstruct coregistered precontrast and postcontrast images. A prospective study conducted from September 2021 to February 2022 involved 10 healthy human volunteers (mean age, 37 years ± 21 [SD]; eight male, two female) and four pigs with reperfusion injury. Quantitative and qualitative MRI measurements were verified in a digital phantom and in vivo. Statistical analysis included intraclass correlation coefficient, Bland-Altman analysis, and paired t tests to compare the proposed method against reference methods. Results Multitasking T1 measurements were highly correlated with two-dimensional (2D) inversion recovery spin-echo-T1 (R2 > 0.99) and T2 measurements with 2D spin-echo-T2 (R2 > 0.960) in the phantom study. In vivo, multitasking provided higher myocardial T1 values than did modified Look-Locker inversion recovery (MOLLI) (1383 msec ± 37.8 vs 1217 msec ± 16.3; P < .001) and higher myocardial T2 values than did T2-prep FLASH (43.9 msec ± 1.4 vs 40.0 msec ± 1.6; P < .001). There was no evidence of a difference between multitasking and MOLLI extracellular volume (ECV) values (27.5% ± 1.4 and 27.9% ± 1.8; P = .52). In pigs, the proposed method depicted increased precontrast T1 and ECV in ischemic injury regions, aligning well with reference measurements. Conclusion The novel 3D multitasking method enables a comprehensive, 20-minute, push-button cardiac MRI examination without ECG gating or breath holding, providing cardiac function, T1, T2, ECV, and late gadolinium enhancement measurements. Keywords: Cardiac MRI, Cardiac MRI Multitasking, Continuous-Acquisition Cardiac MRI, All-in-One Cardiac MRI Supplemental material is available for this article. © RSNA, 2025.

来源:Radiol Cardiothorac Imaging,查看原文。版权归原作者所有。

引用

  1. Mao X, Chen Z, Hu Z, et al. All-in-One Cardiac MRI: Unified Non-Contrast-enhanced and Contrast-enhanced Imaging for a Continuous-Acquisition Examination. Radiol Cardiothorac Imaging. 2025. PMID: 40996331. doi:10.1148/ryct.240383

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。