本研究基于5例患者DSA锥形束源图像重建个体化几何模型,用计算流体动力学分析48丝与64丝血流导向支架(孔隙率均为65%)植入后支架丝间动脉瘤颈处的剪切应力与剪切应变率。结果显示支架植入后支架丝间剪切应力与应变率显著升高,部分患者超过阈值;两种支架设计间无显著差异;动脉瘤体积与支架丝间最大及平均剪切力呈显著正相关。
要点
- 血流导向支架植入后动脉瘤颈支架丝间剪切应力与剪切应变率显著升高,部分患者超过阈值。
- 48丝与64丝支架设计间剪切力无显著差异。
- 动脉瘤体积与支架丝间最大及平均剪切力呈显著正相关。
核心结论
- 血流导向支架植入持续在动脉瘤颈支架丝间诱导超生理性剪切应力与剪切应变率,所测剪切力与动脉瘤体积显著相关。
研究设计与样本
- 基于数字减影血管造影(DSA)锥形束源图像重建5例患者个体化几何模型的计算流体动力学(CFD)研究,含3例远端颈内动脉侧壁动脉瘤与2例大脑中动脉分叉部动脉瘤,模拟孔隙率均为65%的48丝与64丝血流导向支架设计。
诊断标准与关键数值
- 支架植入前动脉瘤入口处剪切应力与应变率在整个心动周期内一般低于既定阈值。
- 支架植入后支架丝间所测值显著升高,部分患者在整个周期或仅在收缩期超过阈值。
- 48丝与64丝支架模型的剪切力均显著高于无支架模型,但两种支架设计间无显著差异。
- 动脉瘤体积与通过支架丝产生的最大及平均剪切力呈统计学显著正相关。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: This study investigates the hemodynamic effects of flow diverter stent (FDS) implantation on intracranial aneurysms, focusing on shear stress and shear strain rates at the neck region between the stent struts.
MATERIALS AND METHODS: Patient-specific geometries were reconstructed from Digital Subtraction Angiography (DSA) cone-beam source images and processed to accurately model arterial anatomy and aneurysm morphology. FDS designs with 48- and 64-wire configurations were modeled while maintaining consistent porosity (65%) across all stents. Blood flow simulations were performed using computational fluid dynamics (CFD) to analyze shear stress and strain rates between the stent struts after FDS implantation.
RESULTS: CFD analysis using DSA images from five patients-three with distal internal carotid artery sidewall aneurysms and two with middle cerebral artery bifurcation aneurysms-demonstrated that, prior to stent implantation, shear stress and strain rate values at the aneurysm inlet were generally below established thresholds throughout the cardiac cycle. Following FDS implantation, those values measured between stent struts increased significantly, with some patients exceeding thresholds either throughout the entire cycle or only during systole. Both 48- and 64-wire FDS models exhibited significantly higher shear forces compared to the non-stent models; however, no significant differences were observed between the two stent designs. Aneurysm volume showed a statistically significant positive correlation with the maximum and average shear forces generated through the stent struts.
CONCLUSION: FDS implantation consistently induces supra-physiologic shear stress and shear strain rates across the aneurysm neck between the stent struts. The measured shear forces are significantly correlated with aneurysm volume.