光子计数探测器CT血管造影中虚拟单能成像联合金属伪影减少算法对颅内动脉瘤手术夹金属伪影的减少效果
纳入2023至2025年动脉瘤夹闭术后行光子计数探测器CT血管造影的患者28例,比较常规120kVp图像与40至190keV虚拟单能图像联合或不联合金属伪影减少算法的定性与定量表现。定量显示高密度伪影在≥70keV降低、低密度伪影与伪影指数在≥100keV降低,对比噪声比随能量升高及联合金属伪影减少算法而下降;定性显示载瘤血管显示在70与100keV改善、联合金属伪影减少算法后下降。作者结论认为减少金属伪影同时保留血管信息的最佳设置为100keV。
要点
- 定量高密度伪影在≥70keV(68.4H.U.)较常规120kVp(82.7)降低(P<0.001)。
- 低密度伪影与伪影指数在≥100keV降低(分别为-2.3H.U.与21.6)(P<0.001)。
- 载瘤血管显示在70与100keV改善、联合金属伪影减少算法后下降(P<0.001)。
核心结论
- 减少金属伪影同时保留血管信息的最佳设置为100keV。
研究设计与样本
- 研究类型为2023至2025年动脉瘤夹闭术后行光子计数探测器CT血管造影患者的影像重建比较研究,纳入28例(中位年龄64岁[53-72],男15例)。
- 图像重建包括常规多能120kVp图像(T3D)及40、70、100、130、160、190keV虚拟单能图像,并对每种图像应用金属伪影减少算法。
- 定量测量由两名神经放射科医师完成,定性评估由四名神经放射科医师完成。
关键影像表现/征象
- 高密度伪影在≥70keV(68.4H.U.)较T3D(82.7)降低(P<0.001)。
- 低密度伪影与伪影指数在≥100keV降低(分别为-2.3H.U.与21.6)(P<0.001)。
- 对比噪声比在≥70keV(20.1)降低,联合金属伪影减少算法后进一步降低(13.6)(P<0.001)。
- 定性评估显示高密度与低密度伪影在≥100keV降低,联合金属伪影减少算法后进一步轻度降低(P<0.001)。
- 噪声在≥100keV降低;载瘤血管显示在70与100keV改善、联合金属伪影减少算法后下降(P<0.001)。
诊断标准与关键数值
- 定量指标包括高密度/低密度伪影、伪影指数与对比噪声比;定性指标包括高密度/低密度伪影、噪声、载瘤血管显示、 Blooms与金属伪影减少算法产生的新伪影。
- 夹闭后载瘤血管显示在70与100keV改善,联合金属伪影减少算法后下降(P<0.001)。
- 夹闭物 Blooms在≥100keV未出现(P<0.001),与是否使用金属伪影减少算法无关(P>0.500)。
- 新伪影在≥100keV联合金属伪影减少算法较T3D联合金属伪影减少算法减少(P<0.001)。
临床应用与阅片要点
- 原文结论认为减少金属伪影同时保留血管信息的最佳设置为100keV。
- 定性评估显示联合金属伪影减少算法虽进一步减少伪影,但载瘤血管显示下降(P<0.001)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To identify conditions that reduce metal artifacts from surgical clips for intracranial aneurysms while preserving parent vessel visualization in CT angiography (CTA) by using a reconstruction method combining virtual monoenergetic image (VMI) and a metal artifact reduction (MAR) algorithm.
MATERIALS & METHODS: Patients who underwent CTA after clipping surgery using photon-counting detector CT (PCD-CT) in 2023-2025 were included. Images were reconstructed as conventional polychromatic 120kVp images (T3D) and 40, 70, 100, 130, 160, and 190 keV using VMI; we created images applying the MAR algorithm to each. Hyperdense/ hypodense artifacts, the artifact index (AI), and the contrast-to-noise ratio (CNR) were quantitatively measured by two neuroradiologists. Hyperdense/ hypodense artifacts, noise, parent vessel visualization, blooming, and new artifacts generated with the MAR algorithm were qualitatively assessed by four neuroradiologists.
RESULTS: 28 patients (median 64 [53-72] years, 15 men) were included. Quantitatively, hyperdense artifacts decreased at ≥ 70 keV (68.4H.U.) versus T3D (82.7) (P < 0.001). Hypodense artifacts and AI decreased at ≥ 100 keV (-2.3H.U. and 21.6) (P < 0.001). CNRs decreased at ≥ 70 keV (20.1) and further decreased when the MAR was applied (13.6) (P < 0.001). Qualitatively, hyperdense/ hypodense artifacts decreased at ≥ 100 keV and further decreased slightly with the MAR (P < 0.001). Noise decreased at ≥ 100 keV. The parent vessel visualization after clipping improved at 70 and 100 keV but decreased with the MAR (P < 0.001). Clip blooming did not occur at ≥ 100 keV (P < 0.001) regardless of the use/nonuse of the MAR (P > 0.500). New artifacts decreased at ≥ 100 keV-MAR versus T3D-MAR (P < 0.001).
CONCLUSION: The optimal setting for reducing metal artifacts while preserving vascular information was 100 keV.
引用
- Nakashima M, Kawai T, Matsumoto K, et al. Metal artifact reduction from surgical clips for intracranial aneurysms in photon-counting detector CT angiography. Eur J Radiol. 2025. PMID: 40939397. doi:10.1016/j.ejrad.2025.112411
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。