文献速递儿科影像骨肌DR

新加坡多族裔儿科队列中自动骨龄工具与人工评估的对比分析

Comparative analysis of an automated bone age tool with manual assessment in a multiethnic Southeast Asian paediatric cohort in Singapore.
PubMed 文献 · Pediatr Radiol · 2025-08-28

该回顾性队列研究纳入新加坡KK妇女儿童医院2012年2月至2021年9月内分泌门诊的200例0-17岁儿科患者,比较BoneXpert自动骨龄与放射科医生及儿科内分泌医生采用Greulich-Pyle法的人工评估。BoneXpert与两位内分泌医生的组内相关系数分别为0.983和0.974,均方根误差分别为0.652年和0.852年;与放射科医生的均方根误差较高,为1.178年。结论认为BoneXpert是可靠且高效的人工骨龄评估替代方法,与内分泌医生评估高度一致。

要点

  • BoneXpert与内分泌医生1的ICC为0.983(95%CI 0.977-0.987),与内分泌医生2的ICC为0.974(95%CI 0.967-0.980)。
  • BoneXpert与内分泌医生1的RMSE为0.652年,与内分泌医生2的RMSE为0.852年。
  • BoneXpert与放射科医生的RMSE较高,为1.178年。

核心结论

  • BoneXpert提供了一种可靠且高效的人工骨龄评估替代方法,与内分泌医生的评估高度一致,可能显著减少评估者间变异并提升临床效率,尤其是在多族裔人群中。

研究设计与样本

  • 回顾性队列研究,纳入200例0-17岁儿科患者,来自新加坡KK妇女儿童医院内分泌门诊,就诊时间范围为2012年2月至2021年9月;由放射科医生和儿科内分泌医生独立采用Greulich-Pyle法评估骨龄,并与BoneXpert自动评估进行比较。

诊断标准与关键数值

  • BoneXpert与内分泌医生1的组内相关系数为0.983(95%CI 0.977-0.987),均方根误差为0.652年。
  • BoneXpert与内分泌医生2的组内相关系数为0.974(95%CI 0.967-0.980),均方根误差为0.852年。
  • BoneXpert与放射科医生的均方根误差为1.178年。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Bone age assessment is essential for evaluating growth and pubertal disorders, but manual methods such as the Greulich-Pyle atlas are subject to inter-observer variability. Automated tools like BoneXpert may improve accuracy and consistency.

OBJECTIVE: To validate the accuracy and reliability of BoneXpert, an automated bone age assessment tool, compared to traditional manual methods performed by paediatric endocrinologists and radiologists in a multiethnic paediatric cohort in Singapore.

MATERIALS AND METHODS: This retrospective cohort study was conducted at KK Women's and Children's Hospital and included paediatric patients aged 0 years to 17 years who presented to the endocrine clinic between February 2012 and September 2021. Bone age was assessed independently using the Greulich and Pyle (GP) method by radiologists and paediatric endocrinologists and compared with BoneXpert's automated assessments.

RESULTS: The study included 200 paediatric patients with a mean age of 9.4 years (± 3.5 years). BoneXpert demonstrated excellent agreement with endocrinologists, with intraclass correlation coefficients (ICC) of 0.983 (95% CI: 0.977-0.987) and 0.974 (95% CI: 0.967-0.980) for Endocrinologists 1 and 2, respectively. The root mean square error (RMSE) between BoneXpert and the endocrinologists was 0.652 years and 0.852 years, indicating high accuracy. However, the RMSE between BoneXpert and radiologists was higher at 1.178 years, reflecting greater variability. Bland-Altman plots showed good agreement between BoneXpert and manual assessments.

CONCLUSION: BoneXpert provides a reliable and efficient alternative to manual bone age assessments and aligns closely with endocrinologists' evaluations. This tool could significantly reduce inter-rater variability and enhance clinical efficiency, particularly in diverse ethnic populations.

来源:Pediatr Radiol,查看原文。版权归原作者所有。

引用

  1. Chan D, Tan CN, Cheah PV, et al. Comparative analysis of an automated bone age tool with manual assessment in a multiethnic Southeast Asian paediatric cohort in Singapore. Pediatr Radiol. 2025. PMID: 40874978. doi:10.1007/s00247-025-06374-4

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。