利用高帧率多普勒超声监测弹性组织中的碎裂
本研究在胶原蛋白和弹性树脂体模及离体牛浅屈肌腱中,比较被动空化成像(PCI)与气泡诱导彩色多普勒(BCD)监测组织粉碎术碎裂的效果。结果显示在胶原蛋白样品中BCD与液化相关性优于PCI(r=0.75±0.22 vs r=-0.40±0.48),在肌腱中BCD回归系数绝对值高于PCI(|r|=0.63±0.19 vs |r|=0.26±0.21)但方向不一致,提示BCD或PCI与BCD联合可能用于监测高弹性组织的聚焦超声碎裂。
要点
- 在胶原蛋白体模中,BCD与液化相关性优于PCI(r=0.75±0.22 vs r=-0.40±0.48)。
- 在肌腱中,BCD回归系数绝对值高于PCI(|r|=0.63±0.19 vs |r|=0.26±0.21),但BCD回归方向不一致。
- BCD或PCI与BCD联合可能用于监测高弹性组织的聚焦超声碎裂。
核心结论
- 与多普勒指标相关的回归趋势提示BCD或PCI与BCD联合可能用于监测聚焦超声诱导的高弹性组织碎裂。
研究设计与样本
- 制作端胶原蛋白和弹性树脂体模(各n=3)用于BCD验证;另制作5个端胶原蛋白凝胶与离体牛浅屈肌腱直接比较,肌腱浸泡于磷酸盐缓冲盐水(健康)或胶原酶(肌腱病性;各3例)1周。所有样品使用1.5 MHz聚焦超声,10 ms脉冲以1 Hz重复(p+=127 MPa,p-=35 MPa),用Verasonics Vantage-128超声系统和ATL L7-4换能器以PCI或BCD监测治疗进展。
关键影像表现/征象
- 在肌腱中,组织粉碎术在磷酸盐缓冲盐水浸泡的12个治疗位置中的4个和胶原酶浸泡的12个治疗位置中的10个产生肉眼可见的碎裂。
- 在聚焦超声脉冲后6.7 ms,BCD与端胶原蛋白样品液化的相关性优于PCI(r=0.75±0.22 vs r=-0.40±0.48)。
- 在肌腱中,BCD监测显示碎裂产生的回归系数绝对值(|r|=0.63±0.19)高于PCI(|r|=0.26±0.21),但BCD的回归方向不一致。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Passive cavitation imaging (PCI) and bubble-induced color Doppler (BCD) have been used successfully to monitor liquefaction of soft tissues during histotripsy. However, it is unclear whether PCI and BCD can be similarly used in highly elastic tissues, which fractionate rather than liquefy. Our objective was to compare PCI with BCD for monitoring histotripsy fractionation of elastic tissues and tissue-mimicking phantoms.
METHODS: Telocollagen and elastic resin phantoms (n = 3 each) were fabricated for BCD validation; an additional 5 telocollagen gels were fabricated for direct comparison to ex vivo bovine superficial digital flexor tendons soaked in phosphate-buffered saline (healthy) or collagenase (tendinopathic; 3 each) for 1 week. All samples were treated with 1.5 MHz focused ultrasound (FUS) using 10 ms pulses repeated at 1 Hz (p+ = 127 MPa, p- = 35 MPa) and treatment progression was monitored with PCI or BCD using a Verasonics Vantage-128 ultrasound system and an ATL L7-4 transducer.
RESULTS: At 6.7 ms post-FUS pulse, BCD better correlated with liquefaction in telocollagen samples than PCI (r = 0.75 ± 0.22 and r = -0.40 ± 0.48, respectively). In tendon, histotripsy produced grossly observable fractionation in 4/12 phosphate-buffered saline- and 10/12 collagenase-soaked treatment locations. Monitoring with BCD indicated that fractionation produced higher regression coefficient magnitudes in BCD (|r| = 0.63 ± 0.19) than PCI (|r| = 0.26 ± 0.21); however, regression directions in BCD were inconsistent.
CONCLUSION: Regression trends associated with Doppler metrics suggest the potential for BCD or perhaps a combination of PCI and BCD for monitoring FUS-induced fractionation in highly elastic tissues.