基于临床患者对中国乳腺锥形束CT平均腺体剂量的评估
本研究回顾性分析2020年10月至2023年11月间不同地区1590例中国女性非增强乳腺锥形束CT的三维影像数据,量化非压迫乳腺特征并计算平均腺体剂量。典型扫描参数下平均腺体剂量范围为3.9至5.1 mGy(中位数4.9 mGy),约为中国2次数字乳腺摄影视图累积辐射剂量的1.81倍。
要点
- 乳腺锥形束CT典型平均腺体剂量中位数为4.9 mGy,范围3.9至5.1 mGy。
- 典型扫描参数为管电压固定49 kVp、管电流50 mA。
- 该剂量约为中国2次数字乳腺摄影视图累积剂量的1.81倍。
核心结论
- 非增强乳腺锥形束CT扫描的典型平均腺体剂量约为中国2次数字乳腺摄影视图累积辐射剂量的1.81倍。
研究设计与样本
- 回顾性分析1590例中国女性非增强乳腺锥形束CT扫描的三维乳腺影像数据,扫描时间为2020年10月至2023年11月,涵盖不同地区。
诊断标准与关键数值
- 乳腺特征总体中位数:胸壁有效直径12.6 cm、胸壁至乳头长度6.8 cm、乳腺体积370.6 cm³、腺体体积46.7 cm³、体积乳腺密度14.4%;管电压固定49 kVp,典型管电流50 mA;平均腺体剂量范围3.9至5.1 mGy(中位数4.9 mGy)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To quantify the non-compressed breast characteristics of females and the typical scanning parameters of breast cone-beam computed tomography (CBCT), and to provide a preliminary assessment of the average glandular dose (AGD) of breast CBCT in China.
METHODS: This study retrospectively analysed 3D breast imaging data from 1590 Chinese women who underwent non-enhanced breast CBCT scans in different regions between October 2020 and November 2023. Initially, a semiautomatic measurement tool integrated into the breast CBCT system was used to quantify the breast characteristics for each patient, including effective diameter at the chest-wall (Deff), chest-wall to nipple length (CNL), breast volume (BV), glandular volume (GV), and volumetric breast density (VBD). Then, the AGD for each patient was calculated using the dose conversion factor, combining breast characteristics and scanning parameters.
RESULTS: The overall median values of Deff, CNL, BV, GV, and VBD reflecting breast characteristics were 12.6 cm, 6.8 cm, 370.6 cm3, 46.7 cm3, and 14.4%, respectively. The tube voltage was fixed at 49 kVp, and the typical tube current was 50 mA. Under the typical scanning parameters, the AGD ranged from 3.9 to 5.1 mGy (median 4.9 mGy).
CONCLUSIONS: The typical AGD for non-enhanced breast CBCT scanning is ∼1.81× higher than the cumulative radiation dose from 2 digital mammography views in China.
ADVANCES IN KNOWLEDGE: This study assessed the AGD of patients undergoing breast CBCT examination based on a large amount of clinical data, providing valuable guidance to clinicians in choosing appropriate breast imaging modalities.