文献速递腹部泌尿生殖技术与安全CT

CT肾脏、输尿管和膀胱扫描中体型特异性剂量估算的评估:水等效直径与有效直径的中心层面法与全层面法比较

Evaluation of size-specific dose estimates in CT kidneys, ureters, and bladder scans: comparison of central-slice and full-slices methods using water-equivalent and effective diameters.
PubMed 文献 · Br J Radiol · 2026-01-01

研究纳入203例成人CT KUB扫描,用IndoseCT软件比较基于水等效直径(Dw)与有效直径(Deff)的体型特异性剂量估算(SSDE),并评估单中心层面代替全层面的准确性。Dw,mean与Deff,mean的SSDE分别为10.3±2.9 mGy与10.2±2.9 mGy;单中心层面与全层面SSDE的R²>0.97,Dw一致性略优。作者认为Deff可作为Dw的替代,单中心层面法可高效估算SSDE。

要点

  • Dw,mean与Deff,mean的SSDE分别为10.3±2.9 mGy与10.2±2.9 mGy,二者可比。
  • 单中心层面与全层面SSDE高度相关(R²>0.97),Dw一致性略优于Deff。
  • 单中心层面法估算SSDE的平均百分比差异Dw为-1.8±3.9%、Deff为-1.1±4.6%,均在±11%内。

核心结论

  • Dw对患者体型估算更准确,但两者差异极小,提示无Dw专用工具时Deff是合适的替代。
  • 使用单中心层面估算SSDE实用高效,可显著降低计算需求。

研究设计与样本

  • 研究获伦理批准后从当地医院采集203例成人患者的CT图像,使用IndoseCT软件计算患者体型指标及相应SSDE值。

诊断标准与关键数值

  • 基于Dw,mean的SSDE为10.3±2.9 mGy,基于Deff,mean的SSDE为10.2±2.9 mGy。
  • 单中心层面与全层面SSDE的平均百分比差异:Dw为-1.8±3.9%,Deff为-1.1±4.6%,所有值均在±11%内。
  • 单中心层面与全层面SSDE的相关系数R²>0.97,Dw的一致性略优于Deff。

临床应用与阅片要点

  • 作者认为,无Dw专用工具时可用Deff替代Dw估算SSDE。
  • 作者认为,单中心层面法可实用高效地估算CT KUB的SSDE,显著减少计算量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To evaluate the size-specific dose estimates (SSDE) in adult patients undergoing kidneys, ureters, and bladder (CT KUB) scans using 2 approaches: water-equivalent diameter (Dw) and effective diameter (Deff). The study also aimed to assess the accuracy of using a single central image slice for SSDE estimation.

METHODOLOGY: Ethical approval was obtained to collect patient data from a local hospital. CT images from 203 adult patients were retrieved and processed using IndoseCT software to calculate patient size metrics and corresponding SSDE values.

RESULTS: SSDE values calculated using the Dw,mean and Deff,mean were comparable: 10.3 ± 2.9 mGy and 10.2 ± 2.9 mGy, respectively. When using a single central slice, the mean percentage differences were -1.8 ± 3.9% for Dw and -1.1 ± 4.6% for Deff, with all values falling within ±11%. A strong correlation was observed between SSDE values derived from single-slice and full-slice measurements (R2 > 0.97), showing slightly better agreement for Dw.

CONCLUSION: While Dw offers a more accurate estimation of patient size, the minimal differences observed suggest that Deff is a suitable alternative when Dw specific tools are unavailable. Additionally, using a single central slice is a practical and efficient method to estimate SSDE, significantly reducing computational demands.

ADVANCES IN KNOWLEDGE: This study provides clinical validation that SSDE can be reliably estimated using a single-slice method in CT KUB examinations, offering a substantial reduction in processing time. It also demonstrates that Deff is a viable substitute for Dw when access to advanced imaging analysis tools is limited.

来源:Br J Radiol,查看原文。版权归原作者所有。

引用

  1. Alenazi K, Abuhaimed A, Alanazi A, et al. Evaluation of size-specific dose estimates in CT kidneys, ureters, and bladder scans: comparison of central-slice and full-slices methods using water-equivalent and effective diameters. Br J Radiol. 2026. PMID: 40845160. doi:10.1093/bjr/tqaf218

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。