探究CT血管造影中有效剂量与基于体重指数的体型特异性剂量估算之间的相关性
该回顾性研究纳入422例接受冠状动脉CT血管造影的成年患者,比较两种有效剂量(ED103与EDDLP)和两种基于体重指数的体型特异性剂量估算(SSDEBMI)的相关性。结果显示ED103与SSDEBMI呈强正相关,且显著强于EDDLP与SSDEBMI的相关性;在肥胖II级女性患者中相关性最高。作者认为SSDEBMI可作为CCTA中可靠的个体化剂量指标。
要点
- ED103与SSDEBMI呈强正相关,且相关性显著强于EDDLP与SSDEBMI。
- 在不同BMI和性别患者中,ED103与SSDEBMI均呈高度显著相关。
- 肥胖II级女性患者中ED103与SSDEBMI相关性最高。
核心结论
- ED103与SSDEBMI呈强正相关,且显著强于EDDLP与SSDEBMI的相关性;SSDEBMI可作为CCTA中可靠的个体化剂量指标,用于个体化辐射风险评估和协议优化。
研究设计与样本
- 回顾性分析,纳入422例成年患者(男性226例、女性196例),平均年龄51.28岁,均接受冠状动脉CT血管造影(CCTA);计算两种有效剂量(ED103和EDDLP,分别基于剂量长度乘积DLP和ICRP转换因子)及两种SSDEBMI(基于CTDIvol和患者BMI)。
诊断标准与关键数值
- ED103与SSDEBMI呈强正相关,且相关性显著强于EDDLP与SSDEBMI。
- 在不同BMI和性别患者中,ED103与SSDEBMI均呈高度显著相关。
- 肥胖II级女性患者中ED103与SSDEBMI相关性最高,归因于为补偿X线衰减增加而进行的比例剂量调整。
临床应用与阅片要点
- 作者认为SSDEBMI可作为CCTA中可靠的个体化剂量指标,用于个体化辐射风险评估和协议优化。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: Computed tomography angiography (CTA) is a widely used diagnostic tool, but concerns about radiation exposure necessitate accurate dose estimation. This study investigates the correlation between Effective Dose (ED) and Size-Specific Dose Estimate (SSDE) estimated from the patient's body mass index (BMI) in CTA to evaluate the utility of SSDE-BMI as a patient-specific dosimetry parameter.
METHODS: A retrospective analysis of 422 adult patients (226 males and 196 females) with an average age of 51.28 years undergoing Coronary Computed Tomography Angiography (CCTA) was conducted. Two values of ED were derived (ED103 and EDDLP derived using the dose-length product (DLP) and ICRP conversion factors), while the two values of SSDEBMI were calculated from the volume-computed tomography dose index (CTDIvol) and the patient's BMI.
RESULTS: Overall, statistical analysis revealed a strong positive correlation between ED103 and SSDEBMIs, significantly stronger than correlations between traditional metrics EDDLP and SSDEBMI. A highly significant correlation between these two dose indices was observed across various patients' BMI and gender. The highest correlation was reported for Obese Class II female patients, which is attributed to proportional dose adjustments made to compensate for increased X-ray attenuation.
CONCLUSIONS: These findings underscore SSDEs' importance in personalized radiation risk assessment and protocol optimization. This supports the use of SSDEBMI as a reliable, patient-specific dose metric in radiation protection strategies for CCTA.
引用
- Mohammed HA, Bari DS, Yacoob Aldosky HY, et al. Exploring the correlation between effective dose and BMI-based size-specific dose estimate in CT angiography. Eur J Radiol. 2025. PMID: 40749323. doi:10.1016/j.ejrad.2025.112342
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。