文献速递心血管MRSPECT

使用基于同位素的参考标准验证临床T1 mapping方法估算的病理细胞外容积分数

Verification of Pathologic ECV Estimated with Clinical T1 Mapping Methods Using an Isotope-based Reference.
PubMed 文献 · Radiol Cardiothorac Imaging · 2025-01-01

研究在六只猪中诱导急性心肌缺血再灌注7天后,用三种MOLLI和一种SASHA序列在体测量细胞外容积分数(ECV),并与离体SPECT参考标准对照。所有序列ECV与参考标准高度相关(r²范围0.71-0.99),但MOLLI 5(3)3和MOLLI 5s(3s)3s在梗死区高估ECV,所有序列在远隔区均高估ECV。

要点

  • 所有T1 mapping序列ECV与SPECT参考标准高度相关,r²范围0.71-0.99。
  • MOLLI 5(3)3和MOLLI 5s(3s)3s在梗死区高估ECV,偏倚分别为8.2%±8.5和4.9%±8.7。
  • 所有序列在远隔区均高估ECV,偏倚范围3%-14%点。

核心结论

  • 研究使用独立的放射性同位素三维采集参考标准,验证了临床T1 mapping序列在一系列病理值范围内测量ECV绝对值的准确性。

研究设计与样本

  • 动物实验,六只猪于2018年10月至2018年11月通过左前降支血管内球囊闭塞诱导急性心肌缺血,再灌注7天后在体测量ECV,使用三种MOLLI序列和一种SASHA序列,参考标准为注射锝99m二乙烯三胺五乙酸后的离体SPECT。

诊断标准与关键数值

  • 所有T1 mapping序列ECV测量值与参考标准高度相关,r²范围0.71-0.99。
  • MOLLI 5(3)3在梗死区高估ECV,偏倚±2SD为8.2%±8.5,P=.005。
  • MOLLI 5s(3s)3s在梗死区高估ECV,偏倚±2SD为4.9%±8.7,P=.04。
  • 所有序列在远隔区高估ECV,偏倚范围3%-14%点,所有MOLLI序列P值≤.001,SASHA序列P=.04。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Purpose To verify the accuracy of clinically used cardiac MRI T1 mapping sequences for measuring extracellular volume fraction (ECV) throughout the spectrum of pathologic values using an independent reference standard. Materials and Methods Acute myocardial ischemia was induced in six pigs using endovascular balloon occlusion in the left anterior descending artery between October 2018 and November 2018. After 7 days of reperfusion, ECV was measured in vivo using four sequences: three modified Look-Locker inversion recovery (MOLLI) and one saturation recovery single-shot acquisition (SASHA) sequences. ECV was also measured using ex vivo SPECT following injection of technetium 99m diethylenetriamine pentaacetic acid. ECV was calculated for corresponding regions of interest of necrosis on cardiac MR and SPECT images and in myocardium remote from injury. Mixed model analysis was used to test for correlation, while the paired t test was used to test for differences in T1 and ECV. Results ECV measurements with all T1 mapping sequences showed high correlation to the reference standard (r2 range, 0.71-0.99). Measurements with MOLLI 5(3)3 and MOLLI 5s(3s)3s overestimated ECV in the infarct region compared with the reference standard (bias ± 2 SDs: 8.2% ± 8.5 and 4.9% ± 8.7; P = .005 and P = .04, respectively). All sequences showed overestimation of ECV in the remote region (range of bias, 3%-14% points, all MOLLI P values ≤ .001, SASHA P = .04). Conclusion This study verified the accuracy of clinically used T1 mapping sequences in measuring ECV in absolute values over a spectrum of pathologic values, using an independent radioisotope-based three-dimensionally acquired reference standard. Keywords: Animal Studies, MR Imaging, Cardiac, Edema, Imaging Sequences, Ischemia/Infarction Supplemental material is available for this article. © RSNA, 2025.

来源:Radiol Cardiothorac Imaging,查看原文。版权归原作者所有。

引用

  1. Nordlund D, Kopic S, Jablonowski R, et al. Verification of Pathologic ECV Estimated with Clinical T1 Mapping Methods Using an Isotope-based Reference. Radiol Cardiothorac Imaging. 2025. PMID: 40772845. doi:10.1148/ryct.240441

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。