增强新生儿臂丛神经MR神经成像:压缩敏感编码与敏感编码的对比分析
本研究纳入45例临床怀疑臂丛神经损伤的新生儿,在3T MRI上比较压缩敏感编码(CS)与传统敏感编码(SENSE)3D NerveVIEW序列的图像质量与扫描时间。CS较SENSE缩短采集时间约30%(3分36秒 vs 5分08秒),信噪比与神经-肌肉对比噪声比无显著差异,主观评分无显著差异,两种方法观察者间一致性均较强。
要点
- CS较SENSE缩短采集时间约30%(3分36秒 vs 5分08秒),图像质量无显著差异。
- SNR与神经-肌肉CNR在CS与SENSE间无显著差异(SNR P=.002;CNR P=.03)。
- 两名放射科医师主观评分无显著差异,观察者间一致性CS κ=0.773、SENSE κ=0.617。
核心结论
- 在新生儿臂丛神经3D NerveVIEW成像中应用CS技术可行且有效,图像质量与SENSE相当,同时显著缩短扫描时间,可能通过提高检查成功率改善患者结局。
研究设计与样本
- 45例临床怀疑臂丛神经损伤的新生儿(男18例、女27例,平均年龄14.82±9.62天),在3T MRI上同时行CS与SENSE 3D NerveVIEW序列,SENSE并行采集加速因子1.3、CS为6,由2名资深儿科放射科医师进行5分半定量法主观评估。
诊断标准与关键数值
- 图像质量定量评估采用SNR和神经-肌肉CNR:CS与SENSE间SNR无显著差异(t(44)=3.109,P=.002),神经-肌肉CNR无显著差异(t(44)=1.984,P=.03);主观评估采用5分半定量法,两方法间图像质量评分无显著差异,观察者间一致性CS κ=0.773、SENSE κ=0.617。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Neonatal brachial plexus imaging faces challenges with extended scan times and motion artifacts. This study assessed whether compressed sensitivity encoding (CS) acceleration could achieve image quality comparable with conventional sensitivity encoding (SENSE) while significantly reducing scanning time, potentially enhancing diagnostic accuracy and success rates in neonatal brachial plexus imaging.
MATERIALS AND METHODS: Forty-five neonates (18 male, 27 female; mean age 14.82 ± 9.62 days) with clinical suspicion of brachial plexus nerve injury were examined by using both CS and SENSE 3D NerveVIEW sequences on a 3T MRI scanner. The parallel acquisition acceleration factor was 1.3 for SENSE and 6 for CS. Image quality was evaluated quantitatively by using SNR and nerve-to-muscle contrast-to-noise ratio (CNR), and qualitatively through a 5-point semiquantitative scale assessment by 2 senior pediatric radiologists.
RESULTS: CS reduced acquisition time by approximately 30% (3:36 versus 5:08 minutes) compared with SENSE, without compromising image quality. No significant differences were found in SNR and nerve-to-muscle CNR between CS and SENSE, with equivalence testing confirming comparable image quality (SNR: t(44) = 3.109, P = .002; nerve-to-muscle CNR: t(44) = 1.984, P = .03). Radiologists' subjective evaluations revealed no significant difference in image quality scores between CS and SENSE, with strong interrater agreement for both methods (CS: κ = 0.773; SENSE: κ = 0.617).
CONCLUSIONS: Implementation of CS technology in 3D NerveVIEW sequences for neonatal brachial plexus imaging is feasible and effective, providing image quality comparable to SENSE while significantly reducing scanning time. This advancement potentially improves patient outcomes through higher success rates in imaging examinations.
引用
- Zhu B, Guo Y, Peng X, et al. Enhanced Neonatal Brachial Plexus MR Neurography: A Comparative Analysis of Compressed Sensitivity Encoding versus Sensitivity Encoding. AJNR Am J Neuroradiol. 2026. PMID: 40623827. doi:10.3174/ajnr.A8915
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。