文献速递腹部儿科影像MR

深度学习重建改善儿童腹部MRI图像质量:基于三维T1快速扰相梯度回波采集的研究

Deep learning reconstruction for improving image quality of pediatric abdomen MRI using a 3D T1 fast spoiled gradient echo acquisition.
PubMed 文献 · Pediatr Radiol · 2025-07-18

该回顾性研究纳入38例儿童患者,比较三维快速扰相梯度回波腹部MRI常规重建与基于商业深度学习算法的原型重建的图像质量。深度学习重建组各结构图像质量评分均高于非深度学习组(3.8±1.1至4.6±0.6 vs 3.1±1.1至3.9±0.6,均P<0.001),所有放射科医生在多数病例中偏好深度学习重建(32-37/38,P<0.001),信噪比提高2.3倍,平均边缘数减少3.9%。

要点

  • 深度学习重建在多数病例中被放射科医生偏好(32-37/38,P<0.001)。
  • 深度学习组各结构图像质量评分均高于非深度学习组(均P<0.001)。
  • 深度学习重建信噪比提高2.3倍,平均边缘数减少3.9%(均P<0.001)。

核心结论

  • 在行对比增强Dixon腹部MRI的广泛儿童患者中,原型深度学习重建通常优于常规方法,并在广泛结构上改善了图像质量。

研究设计与样本

  • 回顾性研究,纳入2023年10月至2024年3月期间接受腹部MRI的38例儿童患者(年龄<18岁,平均8.6±5.7岁,23例男性),采用钆增强加速三维SPGR 2点Dixon采集(LAVA-Flex,GE HealthCare),分别用常规重建和基于商业深度学习算法的原型重建(AIR™ Recon DL,GE HealthCare,75%噪声降低设置)生成图像,由三名放射科医生独立评分。

关键影像表现/征象

  • 深度学习组各评估结构的平均图像质量评分为3.8±1.1至4.6±0.6,非深度学习组为3.1±1.1至3.9±0.6(均P<0.001)。
  • 所有放射科医生在多数病例中偏好深度学习重建(32-37/38,P<0.001)。
  • 深度学习图像较非深度学习图像信噪比提高2.3倍,平均边缘数减少3.9%(均P<0.001)。
  • 除脊柱和非深度学习组肾上腺外,所有评分解剖结构的观察者间一致性为中等至较高度(W=0.41-0.74,均P<0.01)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Deep learning (DL) reconstructions have shown utility for improving image quality of abdominal MRI in adult patients, but a paucity of literature exists in children.

OBJECTIVE: To compare image quality between three-dimensional fast spoiled gradient echo (SPGR) abdominal MRI acquisitions reconstructed conventionally and using a prototype method based on a commercial DL algorithm in a pediatric cohort.

MATERIALS AND METHODS: Pediatric patients (age < 18 years) who underwent abdominal MRI from 10/2023-3/2024 including gadolinium-enhanced accelerated 3D SPGR 2-point Dixon acquisitions (LAVA-Flex, GE HealthCare) were identified. Images were retrospectively generated using a prototype reconstruction method leveraging a commercial deep learning algorithm (AIR™ Recon DL, GE HealthCare) with the 75% noise reduction setting. For each case/reconstruction, three radiologists independently scored DL and non-DL image quality (overall and of selected structures) on a 5-point Likert scale (1-nondiagnostic, 5-excellent) and indicated reconstruction preference. The signal-to-noise ratio (SNR) and mean number of edges (inverse correlate of image sharpness) were also quantified. Image quality metrics and preferences were compared using Wilcoxon signed-rank, Fisher exact, and paired t-tests. Interobserver agreement was evaluated with the Kendall rank correlation coefficient (W).

RESULTS: The final cohort consisted of 38 patients with mean ± standard deviation age of 8.6 ± 5.7 years, 23 males. Mean image quality scores for evaluated structures ranged from 3.8 ± 1.1 to 4.6 ± 0.6 in the DL group, compared to 3.1 ± 1.1 to 3.9 ± 0.6 in the non-DL group (all P < 0.001). All radiologists preferred DL in most cases (32-37/38, P < 0.001). There were a 2.3-fold increase in SNR and a 3.9% reduction in the mean number of edges in DL compared to non-DL images (both P < 0.001). In all scored anatomic structures except the spine and non-DL adrenals, interobserver agreement was moderate to substantial (W = 0.41-0.74, all P < 0.01).

CONCLUSION: In a broad spectrum of pediatric patients undergoing contrast-enhanced Dixon abdominal MRI acquisitions, the prototype deep learning reconstruction is generally preferred to conventional methods with improved image quality across a wide range of structures.

来源:Pediatr Radiol,查看原文。版权归原作者所有。

引用

  1. Zucker EJ, Milshteyn E, Machado-Rivas FA, et al. Deep learning reconstruction for improving image quality of pediatric abdomen MRI using a 3D T1 fast spoiled gradient echo acquisition. Pediatr Radiol. 2025. PMID: 40679617. doi:10.1007/s00247-025-06313-3

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。