手持式即时超声(POCUS)筛查男性腹主动脉瘤的评估:一项比较性队列研究

Evaluation Of Handheld Point-of-Care-Ultrasound (POCUS) in Screening for Abdominal Aortic Aneurysm in Men: A Comparative, Cohort Study.
PubMed 文献 · Ultrasound Med Biol · 2025-07-30

这项比较性队列研究纳入230名既往筛查时肾下主动脉直径25–29 mm的男性,在5–12年后用POCUS(Lumify,Philips)与标准超声(LOGIQ 9E,GE)各测三次肾下主动脉直径。POCUS较标准超声低估0.3 mm,95%一致性界限±3.0 mm,重复性分别为±2.5 mm与±3.2 mm,POCUS诊断腹主动脉瘤(直径≥30 mm)的敏感度94%、特异度98%。

要点

  • POCUS测量肾下主动脉直径较标准超声低估0.3 mm,95%一致性界限±3.0 mm。
  • POCUS诊断腹主动脉瘤(直径≥30 mm)敏感度94%、特异度98%。
  • 标准超声重复性±2.5 mm,POCUS重复性±3.2 mm。

核心结论

  • POCUS测量肾下主动脉直径未显示有临床意义的系统偏倚,但重复性区间较标准超声略宽,提示可考虑将POCUS设备用于腹主动脉瘤筛查。

研究设计与样本

  • 比较性队列研究,纳入230名既往腹主动脉瘤筛查时肾下主动脉直径25–29 mm的男性,5–12年后复查。
  • 同时使用POCUS(Lumify,Philips)与标准超声(LOGIQ 9E,GE)各测量肾下主动脉直径三次。
  • 评估设备间偏倚与一致性界限、设备内重复性,以及POCUS对比标准超声诊断腹主动脉瘤(直径≥30 mm)的敏感度与特异度。

诊断标准与关键数值

  • 腹主动脉瘤定义为肾下主动脉直径≥30 mm。
  • POCUS诊断腹主动脉瘤的敏感度94%、特异度98%(对照标准为标准超声)。
  • POCUS较标准超声低估肾下主动脉直径0.3 mm,95%一致性界限±3.0 mm。

临床应用与阅片要点

  • 作者认为POCUS测量肾下主动脉直径未显示有临床意义的系统偏倚,提示可考虑将POCUS设备用于腹主动脉瘤筛查。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Screening for abdominal aortic aneurysm (AAA) using ultrasound in men reduces aneurysm-related mortality. Ambulatory screening and surveillance procedures facilitated by point-of-care ultrasound (POCUS) may improve adherence to screening programs. This proof-of-concept study evaluated whether infrarenal aortic diameter (IAD) can be accurately and interchangeably measured using handheld POCUS-an ultrasound probe connected to a tablet-compared to standard cart-based ultrasound equipment.

METHODS: This comparative cohort study included men with an IAD of 25-29 mm during AAA screening and re-examined them after 5-12 years, anticipating most would approach the 30 mm AAA threshold. Three IAD measurements were taken using both POCUS (Lumify™, Philips) and standard US (LOGIQ™ 9E, GE). Bias, limits of agreement between devices and repeatability within devices were evaluated. Sensitivity and specificity of POCUS for AAA detection (IAD ≥30 mm) compared to standard ultrasound were also assessed.

RESULTS: In total, 230 men participated. The median age was 75.0 years (IQR: 72.9-76.5), BMI 27.7 (26.0-30.5), and IAD 27.1 mm (23.7-32.8) respectively. POCUS underestimated IAD by 0.3 mm compared to standard US, with 95% limits of agreement ±3.0 mm. Repeatability was ±2.5 mm for standard ultrasound and ±3.2 mm for POCUS. The sensitivity and specificity of POCUS for AAA diagnosis, compared to standard ultrasound, are 94% and 98%, respectively.

CONCLUSION: POCUS showed no systematic bias of clinical importance in measuring IAD although with a slightly wider repeatability interval compared to standard US implying the possibility to consider POCUS devices for AAA-screening purposes.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。