双低剂量腹部盆腔CT用于急性阑尾炎:图像质量与诊断性能的早期探索性分析
这项前瞻性单臂研究的探索性分析纳入49例疑似急性阑尾炎的急诊患者,采用自动kVp选择和按kVp及体重调整对比剂用量的双低剂量CT方案,并与53例既往采用固定碘量和低kVp单低剂量CT的历史对照比较。双低剂量CT显著降低碘剂量(0.383±0.049 vs 0.589±0.085 g I/kg,P<0.001),图像噪声和阑尾可见度相当,急性阑尾炎和替代诊断的阳性预测值无显著差异。结论认为按管电压和体重调整的降低碘剂量双低剂量CT方案在急诊可行,可大幅节省对比剂剂量而不影响诊断性能。
要点
- 双低剂量CT同时降低碘对比剂剂量和管电压,碘剂量0.383±0.049 g I/kg,较单低剂量CT的0.589±0.085 g I/kg显著降低(P<0.001)。
- 双低剂量CT组信噪比和对比噪声比低于单低剂量CT组,但图像噪声和阑尾可见度保持相当。
- 急性阑尾炎阳性预测值双低剂量CT组63.6% vs 单低剂量CT组72.7%,替代诊断阳性预测值100% vs 86.7%,两组间无显著差异。
核心结论
- 按管电压和体重调整碘剂量的双低剂量CT方案在急诊疑似急性阑尾炎中可行,可大幅节省对比剂剂量而不损害对急性阑尾炎或替代诊断的诊断性能。
研究设计与样本
- 前瞻性单臂研究的探索性分析,双低剂量CT组49例(年龄18-44岁,急诊疑似急性阑尾炎),历史对照组53例(相同纳入标准,研究启动前接受单低剂量CT),采用自动kVp选择和按kVp及体重调整对比剂用量的双低剂量CT方案,参考标准为手术所见、临床随访或Alvarado评分≥7。
关键影像表现/征象
- 双低剂量CT组信噪比和对比噪声比低于单低剂量CT组,但图像噪声和阑尾可见度保持相当。
- 双低剂量CT方案采用自动kVp选择和按kVp及体重调整对比剂用量,目标为kVp自120 kVp每降低10 kVp碘剂量减少10%(参考标准0.521 g I/kg at 120 kVp)。
诊断标准与关键数值
- 急性阑尾炎阳性预测值:双低剂量CT组63.6% vs 单低剂量CT组72.7%,两组间无显著差异。
- 替代诊断阳性预测值:双低剂量CT组100% vs 单低剂量CT组86.7%,两组间无显著差异。
- 碘剂量:双低剂量CT组0.383±0.049 g I/kg vs 单低剂量CT组0.589±0.085 g I/kg,双低剂量CT组显著降低(P<0.001)。
临床应用与阅片要点
- 双低剂量CT方案采用自动kVp选择和按kVp及体重调整对比剂用量,目标为kVp自120 kVp每降低10 kVp碘剂量减少10%(参考标准0.521 g I/kg at 120 kVp)。
- 诊断性能以手术所见、临床随访或Alvarado评分≥7为参考标准,双低剂量CT组急性阑尾炎阳性预测值63.6%、替代诊断阳性预测值100%。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To evaluate the feasibility of a standardized double low-dose CT (DLCT) protocol that simultaneously reduces both iodine contrast media (CM) dose and tube voltage, and to compare its image quality and diagnostic performance with a single low-dose CT (SLCT) protocol using fixed iodine dose and reduced tube voltage, representative of institutional low-radiation dose imaging practice in suspected acute appendicitis.
METHODS: In this exploratory analysis of a prospective single-arm study, 49 participants aged 18-44 years presenting to the emergency department (ED) with suspected acute appendicitis underwent DLCT using automated kVp selection and a kVp- and weight-adjusted CM dosing protocol. This protocol targeted a 10% iodine dose reduction for every 10 kVp decrease from a reference standard of 0.521 g I/kg at 120 kVp. A historical control group of 53 patients with the same inclusion criteria, who had undergone SLCT using a fixed CM dose and low-kVp protocol (80-100 kVp) prior to study initiation, was retrospectively included. Quantitative (signal-to-noise ratio [SNR] and contrast-to-noise ratio [CNR]) and qualitative (subjective image quality and appendiceal visibility) assessments were compared. Diagnostic performance was evaluated using positive predictive value (PPV) for acute appendicitis and alternative diagnoses, based on surgical findings, clinical follow-up, or an Alvarado score ≥ 7.
RESULTS: DLCT significantly reduced iodine dose (0.383 ± 0.049 g I/kg vs. 0.589 ± 0.085 g I/kg; P < 0.001) while maintaining comparable image noise and appendiceal visibility. Although SNR and CNR were lower in the DLCT group, diagnostic performance including PPVs for acute appendicitis (63.6% vs. 72.7%) and alternative diagnoses (100% vs. 86.7%) did not significantly differ between groups.
CONCLUSION: An iodine dose-reducing DLCT protocol tailored to tube voltage and body weight is feasible in the ED setting and offers substantial CM dose saving without compromising diagnostic performance for acute appendicitis or alternative diagnoses.