使用实验性硅基光子计数CT区分斑块与支架:与传统能量积分探测器CT的实验和临床比较
研究将人动脉支架标本置入琼脂水模体,用硅基光子计数CT(Si-PCCT)与能量积分探测器CT(EIDCT)在CTDIvol 9 mGy下扫描,并在150 mm与50 mm视野重建。Si-PCCT在两种视野下均改善斑块-支架对比、减少 blooming(P<0.001),直径测量准确性仅在50 mm视野改善(P=0.003)。单例临床颈动脉造影病例的评估结果与实验发现一致。
要点
- Si-PCCT在150 mm和50 mm视野下均较EIDCT改善斑块-支架对比并减少blooming(P<0.001)。
- Si-PCCT支架直径测量准确性仅在50 mm视野下较EIDCT改善(P=0.003)。
- 150 mm视野下Si-PCCT在支架外观、斑块外观、支架-斑块区分和blooming方面均优于EIDCT。
核心结论
- Si-PCCT在实验模体和一例临床病例中显示出改善图像对比度和空间分辨率的潜力,有助于改善斑块和支架的评估。
研究设计与样本
- 实验研究:人切除并置入支架的动脉嵌入琼脂水模体中心,用Si-PCCT和EIDCT以相似扫描技术扫描,目标CTDIvol为9 mGy,在150 mm和50 mm视野下以等效设置重建;另比较一例临床颈动脉造影病例的EIDCT与Si-PCCT图像。
关键影像表现/征象
- Si-PCCT在150 mm和50 mm视野下均较EIDCT改善斑块-支架对比并减少blooming(P<0.001)。
- 150 mm视野下Si-PCCT图像在支架外观(P=0.001)、斑块外观(P<0.001)、支架-斑块区分(P=0.002)和blooming(P=0.001)方面均优于EIDCT图像。
- 50 mm视野下Si-PCCT未观察到进一步改善。
- 50 mm视野EIDCT图像较其150 mm视野图像改善斑块-支架对比(P=0.046)、减少blooming(P<0.001)并改善支架外观(P=0.025)。
诊断标准与关键数值
- 斑块-支架区分、支架直径测量准确性和blooming采用5点Likert量表定量和定性评估。
- Si-PCCT与EIDCT比较:两种视野下斑块-支架对比和blooming差异均P<0.001;支架直径测量准确性仅在50 mm视野差异P=0.003。
- 定量和定性评估的组内相关系数(ICC)分别为0.93和0.96。
临床应用与阅片要点
- 在实验模体和一例临床颈动脉造影病例中,Si-PCCT较EIDCT改善了斑块-支架对比、减少了blooming并改善了支架外观和支架-斑块区分。
- Si-PCCT支架直径测量准确性的改善仅在50 mm视野下观察到,150 mm视野下未报告该改善。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To compare plaque-stent interface using an experimental Silicon-based photon-counting CT (Si-PCCT) with a conventional energy-integrating detector CT (EIDCT).
METHODS: A human-resected and stented artery was embedded centrally into an agar-water phantom and scanned in a Si-PCCT and an EIDCT using a similar scan technique targeting a CT volumetric dose index (CTDIvol) of 9 mGy. Images were reconstructed using equivalent settings at 150 mm and 50 mm field-of-view (150FOV, 50FOV). Plaque-stent distinction, stent diameter measurement accuracy and blooming were quantitatively and qualitatively assessed using a 5-point Likert scale. Result comparisons between systems and FOVs were evaluated with a t-test and a Wilcoxon signed-rank test with inter-reader agreement assessed using intra-class correlation (ICC). Additionally, EIDCT and Si-PCCT images from a single clinical carotid angiography case were compared.
RESULTS: Si-PCCT enabled improved plaque-stent contrast and reduced blooming (p < 0.001) at both FOVs, whereas diameter measurement accuracy was only improved at 50FOV (p = 0.003), compared to EIDCT. 150FOV Si-PCCT images were preferred over EIDCT images forstent appearance (p = 0.001), plaque appearance (p < 0.001), stent-plaque distinction (p = 0.002) and blooming (p = 0.001). 50FOV showed similar trends. 50FOV EIDCT images enabled improved plaque-stent contrast (p = 0.046), reduced blooming (p < 0.001), and improved stent appearance (p = 0.025). For Si-PCCT, no improvements were observed at 50FOV. Quantitative and qualitative ICC was 0.93 and 0.96, respectivelyThe experimental findings are reflected in the evaluation of the patient case.
CONCLUSION: This study demonstrates the potential of Si-PCCT to improve image contrast and spatial resolution, enabling improved assessment of plaque and stent in an experimental phantom and a clinical patient case.