简化肝脏铁评估:3D qDixon MRI定量肝脏铁过载的准确性与局限性

Streamlining liver iron assessment: accuracy and limitations of 3D qDixon MRI for liver iron overload quantification.
PubMed 文献 · Eur J Radiol · 2025-06-12

前瞻性纳入161例慢性输血依赖患者,以FerriScan和1.5 T多回波GRE为参考标准评估qDixon定量肝铁浓度的准确性。qDixon与ME-GRE在正常、轻度、中度铁过载中一致性良好(CCC 0.986、0.982、0.956),重度铁过载一致性差(CCC 0.216)。分段回归确定qDixon准确定量LIC的上限阈值为28.47 mg/g干重,低于该阈值时与FerriScan线性相关强(r=0.82,P<0.001)。

要点

  • qDixon在LIC低于28.47 mg/g干重时与FerriScan一致性良好(CCC 0.984),可用于定量评估。
  • 重度肝铁过载时qDixon与ME-GRE一致性差(CCC 0.216),平均差异达6.8 mg/g干重。
  • qDixon-R2*超过约892.86 s-1(对应LIC 28.47 mg/g干重以上)时定量不可靠。

核心结论

  • 1.5 T qDixon序列可在特定范围(LIC低于28.47 mg/g干重)内定量评估肝铁浓度,简化临床工作流程。

研究设计与样本

  • 前瞻性研究,纳入161例慢性输血依赖患者(男83例、女43例,平均年龄21.04岁),均行ME-GRE和qDixon扫描,其中67例行FerriScan检查;以FerriScan测定的LIC和1.5 T ME-GRE为参考标准。

诊断标准与关键数值

  • qDixon-LIC与ME-GRE-LIC平均差异在正常、轻度、中度、重度肝铁过载分别为0.22、0.01、-0.27、6.8 mg/g干重。
  • 对应CCC值分别为0.986、0.982、0.956、0.216。
  • 分段回归确定qDixon准确定量LIC的上限阈值为28.47 mg/g干重,超过该阈值时qDixon-R2*达892.86 s-1。
  • 低于该阈值时qDixon与FerriScan估测LIC线性相关强(r=0.82,P<0.001),平均差异1.68 mg/g干重,CCC为0.984。

临床应用与阅片要点

  • 作者结论认为1.5 T qDixon序列可用于LIC低于28.47 mg/g干重范围内的肝铁定量评估,简化临床工作流程。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To prospectively evaluate the accuracy and limitations of the 1.5 T 3D-Dixon sequence (qDixon) in quantifying liver iron concentration (LIC), using FerriScan technology-determined LIC and the 1.5 T multi-echo GRE sequence (ME-GRE) as reference standards.

METHODS: A total of 161 chronic transfusion-dependent patients, 83 men and 43 women with a mean age of 21.04 years, were scanned using ME-GRE and qDixon sequences; among them, 67 were scanned using FerriScan technology. R2* values (ME-GRE-R2* and qDixon-R2*) and LIC were obtained for both sequences. Comparative analysis was conducted using concordance correlation coefficients (CCC), Bland-Altman plots, and linear regression. A piecewise regression model was constructed to determine the upper limit for quantifying LIC using the qDixon.

RESULTS: The qDixon-LIC and ME-GRE-LIC showed mean differences of 0.22, 0.01, -0.27, and 6.8 mg/g dry weight (dw) in normal, mild, moderate, and severe liver iron overload with corresponding CCC values of 0.986, 0.982, 0.956, and 0.216, respectively. The piecewise regression model established a 28.47 mg/g dw threshold for accurate liver iron quantification using qDixon, beyond which qDixon-R2* reached 892.86 s-1. Below this threshold, qDixon demonstrated a strong linear correlation with FerriScan in estimating LIC (r = 0.82, p < 0.001), with a negligible mean difference of 1.68 mg/g dw and a CCC of 0.984.

CONCLUSION: The 1.5 T qDixon sequence can be used to quantitatively assess LIC within a specific range (LIC < 28.47 mg/g dw), simplifying clinical workflow procedures.

来源:Eur J Radiol,查看原文。版权归原作者所有。