MRI评估腹部器官铁浓度:松弛测量法之间的比较研究
本研究在1.5T MRI上比较商用3D定量Dixon序列(qDixon)与2D多梯度回波序列(GRE)对腹部多器官铁定量的效果,共纳入171例患者的211次MR检查。结果显示两种方法测量的肝、胰、脾、肾R2*值一致性好,肝脏铁过载诊断总体一致率98.5%、Kappa 0.95,qDixon可作为优秀的辅助诊断方法。
要点
- qDixon与GRE测量肝、胰、脾、肾R2*值的ICC为0.99、0.95、0.96、0.97,一致性高。
- 肝脏铁过载诊断总体一致率98.5%、Kappa 0.95;肾脏Kappa 0.69为四器官中最低。
- qDixon可作为铁过载评估的优秀辅助诊断方法,与GRE结果高度吻合。
核心结论
- qDixon与GRE在评估腹部器官铁过载时测量的R2*值具有良好一致性和显著正相关,qDixon是优秀的辅助诊断方法。
研究设计与样本
- 在1.5T MRI上采集171例患者(86男、85女,中位年龄21岁)共211次MR检查的GRE与qDixon数据,比较肝、胰、脾、肾R2*值,采用Bland-Altman、ICC、CCC和线性回归分析,并以总体一致率和Kappa系数评估铁过载诊断一致性。
诊断标准与关键数值
- 肝、胰、脾、肾GRE与qDixon的R2*值CCC分别为0.98、0.94、0.96、0.95,ICC分别为0.99、0.95、0.96、0.97,线性回归决定系数分别为0.96、0.89、0.93、0.92(均P<0.001)。
临床应用与阅片要点
- qDixon与GRE测量的R2*值高度一致,可用于腹部器官铁过载评估;肾脏铁过载诊断Kappa为0.69,低于肝、胰、脾。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To compare magnetic resonance imaging (MRI) commercial 3D quantitative Dixon sequence (qDixon) and 2D multi-gradient recalled echo sequence (GRE) for iron quantification in multiple abdominal organs.
METHODS: 1.5T MRI GRE and qDixon data were collected on patients with 211 MR exams from 171 patients (86 males, 85 females; median age: 21 years). Compare the R2* values of liver, pancreas, spleen, and kidneys using Bland-Altman, intraclass correlation coefficients (ICC), concordance correlation coefficients (CCC), and linear regression. Iron overload (IO) diagnostic concordance was assessed using overall agreement and the Kappa coefficient.
RESULTS: Bland-Altman analysis of liver, pancreas, spleen, and kidney R2* values between GRE and qDixon, resulted in a bias (absolute mean difference) of -11.3 1/s (LoA: 77.7 and -100.3), -11.2 1/s (LoA: 123.9 and -146.3), 5.1 1/s (LoA: 117.3 and -107.1), and 1.9 1/s (LoA: 14.7 and -10.9). The CCCs between GRE and qDixon R2* values for liver, pancreas, spleen, and kidneys were 0.98, 0.94, 0.96, and 0.95, the ICCs were 0.99, 0.95, 0.96, and 0.97, respectively. Linear regression analysis correlating abdominal organs R2* values of GRE and qDixon resulted in a coefficient of determination of 0.96, 0.89, 0.93, and 0.92 (all P < 0.001). The overall agreement was 98.5 %, 94.8 %, 92.6 %, and 90.7 %; the Kappa value was 0.95, 0.89, 0.84, and 0.69 (all P < 0.001).
CONCLUSION: The qDixon and GRE showed good agreement and significant positive in measuring R2* values for IO assessment in abdominal organs, with qDixon being an excellent adjunctive diagnostic method.