文献速递技术与安全介入US

新型E4D介入成像探头的超声图像质量评估

Ultrasound Image Quality Assessment for Novel E4D Interventional Imaging Probe.
PubMed 文献 · Ultrasound Med Biol · 2025-06-20

本研究旨在表征一种新型磁共振兼容实时二维矩阵阵列(4D)线阵超声探头(E4DL)的图像质量,该探头设计用于超声引导介入。在体模中评估了穿透深度、横向分辨率和4 mm球形病灶可检测性,并与两种商用探头(4V心脏探头和C1-6曲阵腹部探头)比较,同时采集了在体肝脏图像。E4DL与4V的穿透深度分别为147.0 ± 1.4 mm和164.8 ± 3.2 mm,C1-6穿透深度更优(180+ mm);横向分辨率在125 mm深度内相当;在各自焦点附近比较病灶信噪比时,E4DL优于两种商用探头。结论认为E4DL的图像性能指标与商用探头相当,可能有助于介入中的图像引导和运动管理。

要点

  • E4DL穿透深度147.0 ± 1.4 mm,4V为164.8 ± 3.2 mm,C1-6达180+ mm。
  • 横向分辨率在125 mm深度内三种探头相当。
  • 病灶信噪比在各自焦点附近比较时E4DL优于两种商用探头。

核心结论

  • E4DL探头的图像性能指标与商用探头相当,由于其图像质量和设计特点,可能有助于医疗介入中的图像引导和运动管理。

研究设计与样本

  • 研究论文;在体模中评估E4DL(3 MHz原型换能器)与4V心脏探头(GE Vivid E95扫描仪)及C1-6曲阵腹部探头(GE LOGIQ E10扫描仪)的图像质量,并采集在体肝脏图像。

关键影像表现/征象

  • E4DL穿透深度147.0 ± 1.4 mm,4V穿透深度164.8 ± 3.2 mm。
  • C1-6穿透深度更优,达180+ mm。
  • 横向分辨率在125 mm深度内所有探头相当。
  • C1-6的病灶信噪比仰角焦区约在50 mm,深于E4DL的35 mm和4V的40 mm。
  • 在各自焦点附近比较病灶信噪比时,E4DL优于两种商用探头。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Ultrasound for guidance during therapy procedures often is augmented when image features of targeted regions are correlated with features derived from other cross-sectional images such as X-ray, CT and MRI. This work aimed to characterize image quality of a novel magnetic resonance-compatible, real-time two-dimensional matrix array (4D) linear ultrasound probe (E4DL) designed for ultrasound-based image-guided interventions. Performance of this 3 MHz prototype transducer is compared to that of two commercial probes-a 4V cardiac probe and a C1-6 curvilinear abdominal probe (GE Healthcare, Milwaukee, WI).

METHODS: Image quality measurements were performed on a GE Vivid E95 scanner for the E4DL and 4V and on a GE LOGIQ E10 scanner for the C1-6. Depth of penetration (DOP), lateral resolution and 4 mm spherical lesion detectability versus depth (lesion signal-to-noise ratio [LSNR]) were assessed in phantoms. Additionally, in vivo liver images were acquired.

RESULTS: The E4DL and 4V probes achieved DOPs of 147.0 ± 1.4 mm and 164.8 ± 3.2 mm, respectively. The C1-6 displayed superior DOP (180+ mm). The lateral resolution was comparable for all probes up to 125 mm. From the LSNR analysis, the C1-6 elevational focal zone appears to be around 50 mm; this was at a greater depth than that of the E4DL (35 mm) or 4V (40 mm) probes. When comparing the LSNR near each focal zone, the E4DL probe outperformed both commercial probes.

CONCLUSION: The E4DL probe yielded image performance metrics comparable to other commercial probes. Due to its image quality and design features, this probe may be useful for image guidance and motion management during medical interventions.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Jupitz S, Zagzebski J, Dekker S, et al. Ultrasound Image Quality Assessment for Novel E4D Interventional Imaging Probe. Ultrasound Med Biol. 2025. PMID: 40544102. doi:10.1016/j.ultrasmedbio.2025.05.019

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。