优化的束硬化伪影校正算法在急诊头颅平扫CT中改善图像质量与诊断确定性的研究

Improved Beam-hardening Reduction in CT - Impact on Image Quality and Diagnostic Certainty in Emergency Imaging of the Brain.
PubMed 文献 · Acad Radiol · 2025-06-12

回顾性纳入2023年连续100例急诊头颅平扫CT患者,比较标准与优化迭代束硬化校正算法(iBHC)。优化算法幕上SNR降低但脑桥SNR升高,后颅窝伪影指数显著降低,四位阅片者多数认为其图像质量与诊断确定性更优。

要点

  • 优化iBHC算法使后颅窝伪影指数从6.4降至5.5(P<.0001),幕上及幕下图像被多数阅片者评为更优。
  • 优化算法幕上SNR降低(额叶10.5 vs 12.9),但脑桥SNR升高(5.9 vs 5.5),伪影减少与噪声变化需结合区域判读。
  • 四位阅片者在99%–100%的幕上病例和86%–100%的幕下病例中选择优化算法为优。

核心结论

  • 优化的iBHC算法在急诊头颅平扫CT中显著改善图像质量、减少伪影并提高诊断确定性。

研究设计与样本

  • 回顾性研究,连续纳入2023年因排除创伤性脑损伤行急诊头颅平扫CT的100例患者(55例女性,年龄64.1±20岁),比较标准与优化迭代束硬化校正算法重建图像。

关键影像表现/征象

  • 优化算法CT衰减值显著低于标准算法(所有P<.0001)。
  • 优化算法幕上SNR显著降低(额叶区10.5 vs 12.9,P<.0001),脑桥SNR显著升高(5.9 vs 5.5,P<.0001)。
  • 优化算法后颅窝伪影指数显著降低(5.5 vs 6.4,P<.0001),而颅下伪影指数无显著差异(P=0.304)。

诊断标准与关键数值

  • 后颅窝伪影指数(PFAI)和颅下伪影指数(SAI)用于量化伪影,优化算法PFAI显著降低而SAI无显著差异。
  • 两位神经放射科医师和两位急诊放射科医师采用5点Likert量表及侧对侧比较独立评价伪影与诊断确定性。
  • 优化算法在幕上病例中被阅片者1至4分别以100%、100%、99%、99%的比例选为优;在幕下病例中分别以100%、99%、99%、86%的比例选为优。

临床应用与阅片要点

  • 在急诊头颅平扫CT中,优化iBHC算法可减少后颅窝伪影并提高诊断确定性,阅片时可优先选用。
  • 优化算法使幕上SNR降低,判读幕上结构时需注意噪声变化。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: To assess the impact of an optimized beam-hardening artifact reduction algorithm on image quality and diagnostic certainty in emergency unenhanced brain CT.

MATERIALS AND METHODS: Retrospective study of consecutive patients referred for unenhanced brain CT to rule out traumatic brain injuries in 2023. Images were reconstructed using both (A) a standard and (B) an optimized iterative beam-hardening correction algorithm (iBHC). CT attenuation, image noise and SNR were measured in the cortex of supratentorial and infratentorial regions and in the pons. Posterior Fossa Artifact Index (PFAI) and Subcalvarial Artifact Index (SAI) were calculated. Two neuroradiologists and two emergency radiologists independently compared artifacts and diagnostic certainty between both algorithms using 5-point Likert scales and side-by-side comparisons. A paired Wilcoxon-test with correction for multiple testing was used.

RESULTS: 100 consecutive patients (55 women; 64.1 ± 20 years) were included. CT attenuation was significantly lower for B (all P <.0001). SNR was significantly lower supratentorial (frontal region: 10.5 vs. 12.9, P<.0001) and significantly higher in the pons (5.9 vs. 5.5, P <.0001) for B. PFAI was significantly reduced for B (5.5 vs. 6.4, P <.0001), while there was no significant difference in SAI (P = 0.304). The optimized algorithm was selected as superior in 100%, 100%, 99%, 99% of supratentorial and in 100%, 99%, 99%, 86% of infratentorial cases for readers 1 to 4, respectively.

CONCLUSION: An optimized iBHC algorithm demonstrated significantly improved image quality, reduced artifacts and improved diagnostic certainty in emergency unenhanced brain CT.

来源:Acad Radiol,查看原文。版权归原作者所有。