新鲜与福尔马林固定脑组织的基本超声特性以参数图形式显示

Fundamental Ultrasonic Properties of Fresh and Formalin Fixed Brain Visualized as Parametric Images.
PubMed 文献 · Ultrasound Med Biol · 2025-06-03

研究用5 MHz聚焦换能器浸没扫描系统对9个牛脑的23个1 cm厚脑切片行四种超声参数测量,新鲜组织声速1535±2 m/s、衰减频率斜率0.546±0.037 dB/cm/MHz、积分背散射系数0.402±0.165×10⁻³ cm⁻¹sr⁻¹、表观积分背散射-60.1±1.1 dB;白质区声速和衰减频率斜率更高、表观积分背散射和积分背散射系数更低;福尔马林固定使四项参数均发生显著变化。

要点

  • 白质区声速和衰减频率斜率高于其他区域,表观积分背散射和积分背散射系数低于其他区域。
  • 福尔马林固定后声速最多增加0.6%、表观积分背散射最多增加2%、衰减频率斜率最多增加20%、积分背散射系数最多增加55%。
  • 多数参数图中组织结构和白质可清晰区分。

核心结论

  • 多数参数图可清晰区分脑组织结构和白质;福尔马林固定使四种超声参数均发生显著变化。

研究设计与样本

  • 离体研究:取9个牛脑的矢状面和冠状面制备23个1 cm厚脑组织切片,用5 MHz聚焦换能器浸没扫描系统以615 μm步进测量,比较新鲜与福尔马林固定后的声速、衰减频率斜率、积分背散射系数和表观积分背散射。

关键影像表现/征象

  • 白质区表现为声速和衰减频率斜率较高、表观积分背散射和积分背散射系数较低。
  • 多数参数图中组织结构和白质可清晰区分。

诊断标准与关键数值

  • 新鲜组织声速为1535±2 m/s(均值±标准差,代表标本间均值变异)。
  • 新鲜组织衰减频率斜率为0.546±0.037 dB/cm/MHz。
  • 新鲜组织积分背散射系数为0.402±0.165×10⁻³ cm⁻¹sr⁻¹。
  • 新鲜组织表观积分背散射为-60.1±1.1 dB,以平面玻璃反射体为参照。
  • 福尔马林固定后声速最多增加0.6%、表观积分背散射最多增加2%、衰减频率斜率最多增加20%、积分背散射系数最多增加55%(对所有标本所有测量取平均)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: The brain is a complex organ with multiple tissue types and a complicated morphology of lobes, folds, ventricles and other structures. The goal of this study was to create detailed parametric images of brain tissue before and after formalin fixation for four ultrasonic parameters: speed of sound (SOS), frequency slope of attenuation (FSA), integrated backscatter coefficient (IBC) and apparent integrated backscatter (AIB).

METHODS: Twenty-three, 1-cm thick slices of brain tissue were prepared from the sagittal and coronal planes of nine bovine brains. Ultrasonic measurements were performed using an immersion scanning system equipped with a 5 MHz focused transducer moved in 615 μm steps.

RESULTS: Measured values, reported as mean ± standard deviation (representing variation between specimen means) averaged over all measurements on all specimens of fresh tissue, were (1535 ± 2) m/s for SOS, (0.546 ± 0.037) dB/cm/MHz for FSA, (0.402 ± 0.165) × 10-3 cm-1 str-1 for IBC and (-60.1 ± 1.1) dB for AIB measured relative to a planar glass reflector. Regions of white matter were characterized by higher values of SOS and FSA, and lower values of AIB and IBC. Formalin fixation caused up to a 0.6% increase in SOS, up to a 2% increase in AIB, up to a 20% increase in FSA and up to a 55% increase in IBC averaged over all measurements on all specimens.

CONCLUSION: Tissue structures and white matter were clearly distinguishable in most parametric images. Formalin fixation produced significant changes in all four ultrasonic parameters.