基于体重指数调整双能CT方案降低CT肺动脉造影辐射剂量与对比剂用量
前瞻性纳入90例疑似肺栓塞患者并随机分为标准方案A、常规双能方案B和基于体重指数的双能方案C,每组30例。方案C肺动脉CT值、信噪比和对比噪声比高于方案B,方案B和C的剂量长度乘积与对比剂用量均低于方案A,三组诊断准确性和主观图像质量无显著差异。
要点
- 基于体重指数的双能CT肺动脉造影方案在BMI<30 kg/m²患者中可降低辐射剂量与碘负荷且不影响图像质量和诊断准确性。
- 方案C肺动脉CT值、信噪比和对比噪声比均高于方案B。
- 方案B和C的剂量长度乘积与对比剂用量均显著低于方案A。
核心结论
- 与常规CT肺动脉造影相比,基于体重指数的双能CT方案可降低辐射暴露和碘负荷,在BMI低于30 kg/m²的患者中不影响图像质量和诊断准确性。
研究设计与样本
- 前瞻性随机对照研究,纳入90例疑似肺栓塞患者(平均年龄54.1±13.3岁,57名男性),随机分为三组各30例:方案A为标准方案(100 kV/140 mAs/50 mL碘对比剂),方案B为常规双能方案(80-140 kV切换/自动管电流/体重依赖对比剂用量),方案C为基于体重指数的双能方案(80-140 kV切换/基于BMI的管电流/体重依赖对比剂用量、基于BMI的对比剂注射速率);由两名放射科医师独立进行主观和客观图像分析并评估肺栓塞诊断准确性。
关键影像表现/征象
- 方案C肺动脉CT值为358.2±50.5,高于方案B的324.9±57.2(p=0.047)。
- 方案C肺动脉信噪比为17.8±3.2,高于方案B的15.3±2.9(p=0.010)。
- 方案C肺动脉对比噪声比为19.2±4.1,高于方案B的15.7±3.8(p=0.014)。
- 三组间主观图像质量无显著差异。
诊断标准与关键数值
- 三组间肺栓塞诊断准确性无显著差异。
- 方案A剂量长度乘积为287.9±34.0 mGy·cm,方案B为177.2±39.2 mGy·cm,方案C为183.8±23.1 mGy·cm(p<0.001)。
- 方案A对比剂用量为50.0±0.0 mL,方案B为22.1±3.0 mL,方案C为23.2±4.0 mL(p<0.001)。
临床应用与阅片要点
- 对BMI低于30 kg/m²的疑似肺栓塞患者,可采用基于体重指数的双能CT肺动脉造影方案以降低辐射剂量和对比剂用量。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIM: To evaluate the image quality and diagnostic value of a body mass index (BMI)-based dual-energy computed tomography (CT) protocol in computed tomography pulmonary angiography (CTPA) for reduction of radiation dose and contrast medium (CM) volume.
MATERIALS AND METHODS: Patients suspected of having pulmonary embolism were prospectively included and randomly assigned to one of three protocols: Protocol A (a standard protocol, 100 kV/140 mAs/50 mL iodinated CM), B (a conventional dual-energy protocol, 80-140 kV switching/automatic tube current/weight-dependent CM volume) and C (a BMI-based dual-energy protocol, 80-140 kV switching/BMI-based tube current/weight-dependent CM volume, BMI-based CM injection rate). Subjective and objective image analysis were performed by two radiologists independently. Diagnostic accuracy of pulmonary embolism were evaluated. Inter-group comparison was performed.
RESULTS: Ninety patients (mean age 54.1 ± 13.3 years, 57 men) were included, with 30 patients in each group. The CT values, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) of the pulmonary arteries in group C were significantly higher than those in group B (358.2 ± 50.5 vs 324.9 ± 57.2, p=0.047, 17.8 ± 3.2 vs 15.3 ± 2.9, p=0.010, 19.2 ± 4.1 vs 15.7 ± 3.8, p=0.014). The dose length product (DLP) and CM volume were significantly reduced in groups B and C (A vs B vs C, DLP, 287.9 ± 34.0 vs 177.2 ± 39.2 vs 183.8 ± 23.1 mGy·cm), p<0.001, CM, 50.0 ± 0.0 vs 22.1 ± 3.0 vs 23.2 ± 4.0 ml, p<0.001). There were no significant differences in diagnostic accuracy or subjective image quality among the three groups.
CONCLUSION: The BMI-based dual-energy CT protocol can reduce radiation exposure and iodine burden compared to routine CTPA without affecting the image quality and diagnostic accuracy in patients with a BMI under 30 kg/m2.