用于超声定量血流成像的仿血流的表征
本研究按IEC-61685标准制备三种仿血流(BMF 1-3),用新品质因数评估粒子浮力,并测量密度、黏度、声速、衰减与背向散射系数。BMF 2符合全部IEC-61685参考值,在5 mm/s下两种成像模式均获最高对比噪声比。
要点
- 按IEC-61685标准可制备密度、黏度、声速、衰减与背向散射系数可表征的仿血流。
- 新品质因数能追踪粒子分布并帮助获得中性浮力仿血流。
- 过滤显著影响背向散射系数(p < 0.05),不影响其他测量性质。
核心结论
- 建立了仿血流制备与表征的综合方法,据此识别出与IEC-61685标准一致的配方,可能为开发可重复、表征良好的仿血流奠定基础,促进超声定量血流成像技术发展。
研究设计与样本
- 三种仿血流配方(BMF 1-3)按国际电工委员会(IEC)-61685标准制备,动态黏度与粒子浮力不同,在校准的自制微流模体中于5和20 mm/s用功率多普勒与超声造影(CEUS)评估,以对比噪声比(CNR)为性能指标。
诊断标准与关键数值
- BMF 2符合全部IEC-61685参考值(p > 0.05);BMF 1动态黏度最低,BMF 3背向散射系数最高;5 mm/s时BMF 2在两种成像模式CNR均最高;20 mm/s时BMF 3在功率多普勒CNR最高,CEUS中各仿血流表现相似;BMF 2在两种成像模式各流速下CNR相似。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: This study aimed at characterizing blood-mimicking fluids (BMFs), focusing on properties relevant to standardizing quantitative imaging biomarkers in vascular and microvascular ultrasound.
METHODS: Three BMF formulations (BMFs 1-3) were prepared following the International Electrotechnical Commission (IEC)-61685 standard, varying in dynamic viscosity and particle buoyancy. A novel figure of merit (FOM) assessed particle buoyancy over time. Density, dynamic viscosity, speed of sound, attenuation, and backscatter coefficient (BSC) were measured for unfiltered and filtered BMFs and compared to IEC-61685 reference values. BMFs were evaluated using power Doppler and contrast-enhanced ultrasound (CEUS) imaging in a calibrated custom-made microflow phantom at 5 and 20 mm/s, using contrast-to-noise ratio (CNR) as a performance metric.
RESULTS: The FOM accurately tracked particle distribution and enabled obtaining neutrally buoyant BMFs. Filtration significantly impacted BSC (p < 0.05) but did not affect other properties. BMF 2 matched all the IEC-61685 reference values (p > 0.05). BMF 1 exhibited the lowest dynamic viscosity, while BMF 3 had the highest BSC. BMF 2 yielded the highest CNR at 5 mm/s in both imaging modes. At 20 mm/s, BMF 3 yielded the highest CNR in power Doppler imaging, while all BMFs performed similarly in CEUS. BMF 2 provided similar CNRs across flow velocities in both imaging modes.
CONCLUSION: A comprehensive methodology for BMF preparation and characterization was developed, which enabled identifying a formulation that aligned with the IEC-61685 standard. This methodology could establish a foundation for developing reproducible and well-characterized BMFs, facilitating the advance of quantitative flow imaging techniques with ultrasound.
引用
- Ayala-Dominguez L, Baiu C, Castaneda-Martinez L, et al. Characterization of Blood-Mimicking Fluids for Quantitative Flow Imaging with Ultrasound. Ultrasound Med Biol. 2025. PMID: 40383693. doi:10.1016/j.ultrasmedbio.2025.04.012
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。