文献速递心血管MR

在缺血性与非缺血性心肌病患者中评估3.0 T多任务成像衍生的自由呼吸、非心电触发磁共振T1 mapping

Evaluation of free-breathing, non-ECG-triggered magnetic resonance T1 mapping derived from multitasking imaging at 3.0 T in patients with ischaemic and non-ischaemic cardiomyopathy.
PubMed 文献 · Br J Radiol · 2025-01-01

研究比较自由呼吸非心电触发多任务T1 mapping与传统MOLLI序列在缺血性与非缺血性心肌病中的图像质量、T1值、病灶面积及ROC曲线,体外模体显示高度一致,体内多任务与MOLLI强相关,在屏气困难及心律不齐患者中图像质量和诊断准确性更优,但受BMI影响。

要点

  • 多任务T1 mapping与MOLLI在健康志愿者、缺血性及非缺血性心肌病中T1值强相关(R²=0.69、0.71、0.81,均P<.001)。
  • 在屏气困难和不利心率患者中,多任务图像质量优于MOLLI(P<.001)。
  • 稳定患者中两序列诊断准确性相当(P>.05),屏气困难和心律不齐亚组多任务更优(P<.05)。

核心结论

  • 多任务序列在心肌病中提供准确的T1定量和病灶勾画,对屏气不佳或心律不齐患者 robustness 增强;BMI相关限制需技术改进。

研究设计与样本

  • 体外研究比较4个模体的CMR多任务与MOLLI图像T1值;体内研究纳入108例患者(38例缺血性心肌病,70例非缺血性心肌病)和31名健康志愿者,均行中间心室短轴CMR多任务(自由呼吸)和MOLLI(屏气)。

关键影像表现/征象

  • 多任务序列在屏气困难和不利心率患者中图像质量优于MOLLI(P<.001);MOLLI图像质量随屏气困难、心律不齐和高BMI下降(P<.05),多任务受影响较小但仍受BMI影响(P<.05)。

诊断标准与关键数值

  • 模体T1值与参考值高度一致(多任务 vs 参考:R²=0.99,P<.001);体内多任务与MOLLI原生T1在健康志愿者(R²=0.69)、缺血性(R²=0.71)和非缺血性组(R²=0.81)强相关(均P<.001);病灶面积两序列无差异(P>.05);ROC分析以LGE为参考,稳定患者两序列诊断准确性相当(P>.05),屏气困难和心律不齐亚组多任务更优(P<.05)。

临床应用与阅片要点

  • 对于屏气困难或心律不齐的患者,多任务T1 mapping可提供优于MOLLI的图像质量和诊断准确性;但BMI较高时仍需注意其影响。

局限性

  • BMI相关限制需技术改进。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To evaluate the image quality of the free-breathing, non-ECG-multitasking T1 mapping and to compare T1 values, lesion area, and ROC curve with those derived from the traditional MOLLI sequence in ischaemic and non-ischaemic cardiomyopathy.

METHODS: In vitro, T1 values from CMR multitasking and MOLLI images were compared in 4 phantoms. In vivo studies included 108 patients (38 ischaemic cardiomyopathy, 70 non-ischaemic cardiomyopathy) and 31 healthy volunteers. All underwent mid-ventricular short-axis CMR Multitasking (free-breathing) and MOLLI (breath-held). Image quality was scored (4-point scale). T1 values and lesion areas were compared via linear regression, Bland-Altman analysis, and ROC curves (LGE as reference).

RESULTS: Phantom T1 values showed excellent agreement (Multitasking versus reference: R2 = 0.99, P < .001). In vivo, multitasking strongly correlated with MOLLI for native T1 in healthy subjects (R2 = 0.69), ischaemic (R2 = 0.71), and non-ischaemic groups (R2 = 0.81, all P < .001). Multitasking demonstrated superior image quality in patients with breath-holding difficulties and adverse heart rates (P < .001). MOLLI image quality declined with breath-holding difficulties, arrhythmia, and high BMI (P < .05); multitasking was less affected but still impacted by BMI (P < .05). Lesion areas showed no inter-sequence differences (P > .05). ROC analysis revealed comparable diagnostic accuracy between sequences in stable patients (P > .05), but superior multitasking performance in breath-holding difficulties and arrhythmia subgroups (P < .05).

CONCLUSION: Multitasking provides accurate T1 quantification and lesion delineation in cardiomyopathy, with enhanced robustness for patients with poor breath-holding or arrhythmia. BMI-related limitations warrant technical refinement.

ADVANCES IN KNOWLEDGE: Free-breathing, non-ECG CMR multitasking achieves accurate T1 mapping in ischaemic/non-ischaemic cardiomyopathy, overcoming ECG/respiratory limitation.