基于冠状动脉疾病报告与数据系统2.0的本地自动化冠状动脉CT血管成像分析的诊断准确性
本回顾性研究纳入1032例因胸痛行冠状动脉CT血管成像(CCTA)的患者共1041次检查,评估本地人工智能(AI)软件基于CAD-RADS 2.0自动分析钙化积分、狭窄分级和斑块负荷的诊断效能。AI与专家在狭窄分级上呈实质性一致(加权κ=0.73),排除阻塞性冠心病的阴性预测值达98%,但阳性预测值仅39%。
要点
- AI对CAD-RADS≥3或≥4A的每扫描AUC为0.90,阴性预测值98%,可有效排除阻塞性冠心病。
- AI与专家CAD-RADS狭窄分级一致性为加权κ=0.73,斑块负荷评分一致性加权κ=0.97(基于CACS)和0.79(基于SIS)。
- AI对CAD-RADS≥4A的阳性预测值仅39%,阳性结果需人工复核。
核心结论
- 本地AI可准确排除CCTA上的阻塞性冠心病,并在CAD-RADS 2.0狭窄严重程度和斑块负荷方面与人类专家达到实质性至近乎完美的一致性。
研究设计与样本
- 回顾性研究,纳入2017年1月至2021年10月在一家三级学术中心连续行CCTA冠状动脉分析的患者1032例(1041次检查),使用来自三家厂商的四台扫描仪;排除有支架、旁路移植、异常或不可诊断检查的患者;以原始CCTA报告为参考标准,AI与报告不一致处由两名盲法III级读片医师(分别有8年和5年CCTA经验)裁决。
诊断标准与关键数值
- AI与专家CAD-RADS狭窄严重程度分类的加权κ=0.73;每扫描AUC为0.90(95%CI:0.87,0.92)用于CAD-RADS≥3或≥4A;CAD-RADS≥4A的阴性预测值98%(95%CI:97,99),阳性预测值39%(95%CI:32,45);基于CACS的斑块负荷评分加权κ=0.97,基于SIS的斑块负荷评分加权κ=0.79。
临床应用与阅片要点
- AI可在5分钟内输出CACS、CAD-RADS分类和SIS,可用于排除阻塞性冠心病(阴性预测值98%),但CAD-RADS≥4A的阳性预测值仅39%,阳性结果需结合原始报告和专家判读。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Background Chest pain is a leading cause of outpatient and emergency department visits; advancements in artificial intelligence (AI) could improve coronary CT angiography (CCTA) workflows for these patients. Purpose To evaluate the performance of an on-premise AI-based coronary artery calcium scoring (CACS) and CCTA analysis software against expert interpretation based on Coronary Artery Disease Reporting and Data System (CAD-RADS) 2.0. Materials and Methods This retrospective study included consecutive patients undergoing CCTA for coronary analysis at a tertiary academic center between January 2017 and October 2021 across four scanners from three vendors. Patients with stents, bypass grafts, anomalies, or nondiagnostic studies were excluded. On-premise AI output included CACS, CAD-RADS category, and segment involvement score (SIS) within less than 5 minutes. Original CCTA reports were used as the reference, and discrepancies between AI and reports were further adjudicated by two blinded level-III readers with 8 and 5 years of CCTA experience. Agreement among CACS risk categories, CAD-RADS categories, and plaque burden scores was measured with the weighted κ. The area under the receiver operating characteristic curve, positive predictive value, and negative predictive value were used to evaluate diagnostic performance. Bootstrapping was used to estimate 95% CIs. Results A total of 1032 patients (median age, 62 [IQR, 54-69] years; 581 female) with 1041 CCTA images were included: 361 of the 1041 images (35%) were classified as CAD-RADS 0, 274 (26%) as CAD-RADS 1, 186 (18%) as CAD-RADS 2, 101 (10%) as CAD-RADS 3, 95 (9%) as CAD-RADS 4A, 11 (1%) as CAD-RADS 4B, and 13 (1%) as CAD-RADS 5. There was substantial agreement between AI and expert CAD-RADS stenosis severity categories (weighted κ = 0.73). AI demonstrated high performance (per-scan area under the receiver operating characteristic curve, 0.90; 95% CI: 0.87, 0.92) for CAD-RADS greater than or equal to 3 or greater than or equal to 4A and high negative predictive value (98%; 95% CI: 97, 99) but low positive predictive value (39%; 95% CI: 32, 45) for CAD-RADS greater than or equal to 4A. AI-based plaque burden scores derived from CACS reached near-perfect agreement with experts (weighted κ = 0.97), whereas those derived from SIS showed substantial agreement (weighted κ = 0.79). Conclusion On-premise AI accurately ruled out obstructive coronary artery disease at CCTA and achieved substantial to near-perfect agreement with human experts for CAD-RADS 2.0 stenosis severity and plaque burden. © RSNA, 2025 Supplemental material is available for this article. See also the editorial by van Assen and De Cecco in this issue.
引用
- Kay FU, Canan A, Kukkar V, et al. Diagnostic Accuracy of On-Premise Automated Coronary CT Angiography Analysis Based on Coronary Artery Disease Reporting and Data System 2.0. Radiology. 2025. PMID: 40358444. doi:10.1148/radiol.242087
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。