深度学习重建对加速增强颈部MRI图像质量的影响
研究对比深度学习重建TSE(DL-TSE)与常规插值TSE(Conv-TSE)在颈部增强MRI中的图像质量,纳入106例患者,DL-TSE在保持面内分辨率的同时将轴位和冠状位扫描时间分别缩短15.9%和52.6%,整体图像质量、锐利度、口咽及下咽黏膜显示和病灶检出信心均提高。
要点
- DL-TSE在轴位整体图像质量(4.61 vs 3.32)和锐利度(4.40 vs 3.11)上均优于Conv-TSE(P<0.001)。
- DL-TSE口咽冠状位黏膜FWHM 1.08 mm vs Conv-TSE 1.52 mm,下咽1.26 mm vs 1.91 mm(均P<0.001)。
- DL-TSE体外伪影评分更高(4.32 vs 3.90),但重叠解剖结构的伪影稍明显(3.78 vs 3.95,P<0.001)。
核心结论
- DL重建用于加速颈部MRI可提高整体图像质量、锐利度、易运动区域黏膜显示及病灶检出信心,虽重叠解剖结构的伪影稍明显,但大幅缩短扫描时间并提升诊断性能。
研究设计与样本
- 采用3T扫描仪对106例患者行增强T1WI DL-TSE与Conv-TSE成像,DL-TSE使用闭源工作进展版算法(WIP No. 1062,iTSE,version 10;Siemens Healthineers)进行插值和去噪。
关键影像表现/征象
- 静止体模FWHM在DL-TSE持续更锐利;流动体模外信号鬼影DL-TSE较低(0.06% vs 0.14%),体模内较高(8.92% vs 1.75%)。
诊断标准与关键数值
- 患者图像由两名神经放射科医生采用5分Likert量表对整体质量、锐利度、区域黏膜显示、伪影和病灶进行视觉评分。
临床应用与阅片要点
- DL-TSE在轴位和冠状位扫描时间分别缩短15.9%和52.6%的同时,口咽和下咽黏膜显示更清晰,但需注意重叠解剖结构的伪影稍明显。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To compare the quality of deep learning-reconstructed turbo spin-echo (DL-TSE) and conventionally interpolated turbo spin-echo (Conv-TSE) techniques in contrast-enhanced MRI of the neck.
MATERIALS AND METHODS: Contrast-enhanced T1-weighted DL-TSE and Conv-TSE images were acquired using 3T scanners from 106 patients. DL-TSE employed a closed-source, 'work-in-progress' (WIP No. 1062, iTSE, version 10; Siemens Healthineers) algorithm for interpolation and denoising to achieve the same in-plane resolution (axial: 0.26 × 0.26 mm²; coronal: 0.29 × 0.29 mm²) while reducing scan times by 15.9% and 52.6% for axial and coronal scans, respectively. The full width at half maximum (FWHM) and percent signal ghosting were measured using stationary and flow phantom scans, respectively. In patient images, non-uniformity (NU), contrast-to-noise ratio (CNR), and regional mucosal FWHM were evaluated. Two neuroradiologists visually rated the patient images for overall quality, sharpness, regional mucosal conspicuity, artifacts, and lesions using a 5-point Likert scale.
RESULTS: FWHM in the stationary phantom scan was consistently sharper in DL-TSE. The percent signal ghosting outside the flow phantom was lower in DL-TSE (0.06% vs. 0.14%) but higher within the phantom (8.92% vs. 1.75%) compared to Conv-TSE. In patient scans, DL-TSE showed non-inferior NU and higher CNR. Regional mucosal FWHM was significantly better in DL-TSE, particularly in the oropharynx (coronal: 1.08 ± 0.31 vs. 1.52 ± 0.46 mm) and hypopharynx (coronal: 1.26 ± 0.35 vs. 1.91 ± 0.56 mm) (both P < 0.001). DL-TSE demonstrated higher overall image quality (axial: 4.61 ± 0.49 vs. 3.32 ± 0.54) and sharpness (axial: 4.40 ± 0.56 vs. 3.11 ± 0.53) (both P < 0.001). In addition, mucosal conspicuity was improved, especially in the oropharynx (axial: 4.41 ± 0.67 vs. 3.40 ± 0.69) and hypopharynx (axial: 4.45 ± 0.58 vs. 3.58 ± 0.63) (both P < 0.001). Extracorporeal ghost artifacts were reduced in DL-TSE (axial: 4.32 ± 0.60 vs. 3.90 ± 0.71, P < 0.001) but artifacts overlapping anatomical structures were slightly more pronounced (axial: 3.78 ± 0.74 vs. 3.95 ± 0.72, P < 0.001). Lesions were detected with higher confidence in DL-TSE.
CONCLUSION: DL-based reconstruction applied to accelerated neck MRI improves overall image quality, sharpness, mucosal conspicuity in motion-prone regions, and lesion detection confidence. Despite more pronounced ghost artifacts overlapping anatomical structures, DL-TSE enables substantial scan time reduction while enhancing diagnostic performance.