评估自由呼吸人工智能电影序列结合运动校正的左心室与右心室功能:与标准多次屏气电影序列的比较
该单中心横断面研究前瞻性纳入72名受试者,在3.0特斯拉下比较自由呼吸人工智能电影序列结合运动校正(FB AI MOCO)与标准屏气电影序列。FB AI MOCO全层采集时间14.7秒±1.9秒,显著短于标准屏气序列的82.6秒±11.9秒(P<0.001)。两种序列的双心室功能参数无显著差异,所有参数相关性r>0.900、一致性ICC>0.900,提示FB AI MOCO在缩短扫描时间的同时可获得与标准屏气序列相当的图像质量和双心室功能参数。
要点
- FB AI MOCO全层采集时间14.7秒±1.9秒,标准屏气序列82.6秒±11.9秒(P<0.001)。
- FB AI MOCO的eCNR高于标准屏气序列(191.6±54.0 vs. 155.8±68.4,P<0.001)。
- 双心室功能参数两序列间无显著差异,相关性r>0.900、一致性ICC>0.900。
核心结论
- 与多次屏气电影序列相比,FB AI MOCO电影序列在更短扫描时间内获得相当的图像质量和双心室功能参数,提示其具有临床应用前景。
研究设计与样本
- 单中心前瞻性横断面研究,共纳入72名受试者(平均年龄38.3±15.4岁,男性40名),在3.0特斯拉下先后行标准屏气电影序列和FB AI MOCO电影序列扫描。
关键影像表现/征象
- FB AI MOCO电影序列全层采集时间为14.7秒±1.9秒,标准屏气电影序列为82.6秒±11.9秒(P<0.001)。
- FB AI MOCO电影序列的eCNR为191.6±54.0,高于标准屏气电影序列的155.8±68.4(P<0.001)。
- FB AI MOCO电影序列血池-心肌对比评分为4.6±0.5,高于标准屏气电影序列的4.4±0.6(P=0.003)。
- 心内膜边缘定义评分FB AI MOCO为4.2±0.5 vs. 标准屏气4.3±0.7(P=0.123),伪影存在评分4.3±0.6 vs. 4.1±0.8(P=0.085),整体图像质量评分4.4±0.4 vs. 4.3±0.6(P=0.448),均无显著差异。
诊断标准与关键数值
- 图像质量采用5点有序Likert量表评估,基于血池-心肌对比、心内膜边缘定义和伪影三项标准,整体质量评分为三项标准的等权平均值。
- aSNR在FB AI MOCO与标准屏气序列间无显著差异(76.7±20.7 vs. 79.8±20.7,P=0.193)。
- 左心室射血分数降低(<50%,n=24)亚组中,两序列间所有双心室功能指标均无显著差异(P均>0.05)。
- 心率升高(≥80次/分,n=17)亚组中,两序列间所有双心室功能指标均无显著差异(P均>0.05)。
临床应用与阅片要点
- FB AI MOCO电影序列可在自由呼吸下完成采集,全层采集时间14.7秒±1.9秒,较标准屏气序列的82.6秒±11.9秒明显缩短(P<0.001)。
- 左心室射血分数降低(<50%)或心率升高(≥80次/分)的亚组中,FB AI MOCO与标准屏气序列的双心室功能指标均无显著差异(P均>0.05)。
- 两序列所有双心室功能参数间呈强相关(r>0.900)和极好一致性(ICC>0.900)。
- 作者认为FB AI MOCO电影序列在缩短扫描时间的同时可获得与多次屏气电影序列相当的图像质量和双心室功能参数,提示其具有临床应用前景。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To assess the image quality and biventricular function utilizing a free-breathing artificial intelligence cine method with motion correction (FB AI MOCO).
METHODS: A total of 72 participants (mean age 38.3 ± 15.4 years, 40 males) prospectively enrolled in this single-center, cross-sectional study underwent cine scans using standard breath-holding (BH) cine and FB AI MOCO cine at 3.0 Tesla. The image quality of the cine images was evaluated with a 5-point Ordinal Likert scale based on blood-pool to myocardium contrast, endocardial edge definition, and artifacts, and overall quality score was calculated by the equal weight average of all three criteria, apparent signal to noise ratio (aSNR), estimated contrast to noise ratio (eCNR) were assessed. Biventricular functional parameters including Left Ventricular (LV), Right Ventricular (RV) End-Diastolic Volume (EDV), End-Systolic Volume (ESV), Stroke Volume (SV), Ejection Fraction (EF), and LV End-Diastolic Mass (LVEDM) were also assessed. Comparison between two sequences was assessed using paired t-test and Wilcoxon signed-rank test, correlation using Pearson correlation. The agreement of quantitative parameters was assessed using intraclass correlation coefficient (ICC) and Bland-Altman analysis. P < 0.05 was statistically significant.
RESULTS: The total acquisition time of the entire stack for FB AI MOCO cine (14.7 s ± 1.9 s) was notably shorter than that for standard BH cine (82.6 s ± 11.9 s, P < 0.001). The aSNR between FB AI MOCO cine and standard BH cine has no significantly difference (76.7 ± 20.7 vs. 79.8 ± 20.7, P = 0.193). The eCNR of FB AI MOCO cine was higher than standard BH cine (191.6 ± 54.0 vs. 155.8 ± 68.4, P < 0.001), as was the scores of blood-pool to myocardium contrast (4.6 ± 0.5 vs. 4.4 ± 0.6, P = 0.003). Qualitative scores including endocardial edge definition (4.2 ± 0.5 vs. 4.3 ± 0.7, P = 0.123), artifact presence (4.3 ± 0.6 vs. 4.1 ± 0.8, P = 0.085), and overall image quality (4.4 ± 0.4 vs. 4.3 ± 0.6, P = 0.448), showed no significant differences between the two methods. Representative RV and LV functional parameters - including RVEDV (102.2 (86.4, 120.4) ml vs. 104.0 (88.5, 120.3) ml, P = 0.294), RVEF (31.0 ± 11.1 % vs. 31.2 ± 11.0 %, P = 0.570), and LVEDV (106.2 (86.7, 131.3) ml vs. 105.8 (84.4, 130.3) ml, P = 0.450) - also did not differ significantly between the two methods. Strong correlations (r > 0.900) and excellent agreement (ICC > 0.900) were found for all biventricular functional parameters between the two sequences. In subgroups with reduced LVEF (<50 %, n = 24) or elevated heart rate (≥80 bpm, n = 17), no significant differences were observed in any biventricular functional metrics (P > 0.05 for all) between the two sequences.
CONCLUSION: In comparison to multiple BH cine, the FB AI MOCO cine achieved comparable image quality and biventricular functional parameters with shorter scan times, suggesting its promising potential for clinical applications.
引用
- Ran L, Yan X, Zhao Y, et al. Assessment of biventricular cardiac function using free-breathing artificial intelligence cine with motion correction: Comparison with standard multiple breath-holding cine. Eur J Radiol. 2025. PMID: 40373673. doi:10.1016/j.ejrad.2025.112133
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。