7T脑MRI动态并行发射的临床实施:SPACE序列改善均匀性与对比度

Clinical Implementation of Dynamic Parallel Transmission in 7T Brain MRI: Improved Homogeneity and Contrast Using SPACE Sequences.
PubMed 文献 · AJNR Am J Neuroradiol · 2025-11-03

在连续临床7T脑MRI患者中评估原型动态并行发射(pTx)用于T2-SPACE与FLAIR-SPACE序列的表现,并与单发射模式(sTx)比较。结果显示pTx显著改善图像信号与对比度(P = .001),颞叶白质与皮层对比度亦显著改善(P = .001),信号均匀性在较大临床队列中表现一致,结论为pTx在临床患者中可行且性能稳健。

要点

  • 动态pTx较sTx显著改善T2-SPACE与FLAIR-SPACE的信号与对比度(P = .001)。
  • 颞叶白质与皮层对比度在pTx下亦显著改善(P = .001)。
  • 较大临床队列中CoV未发现信号均匀性离群病例。

核心结论

  • 原型动态pTx在T2-SPACE与FLAIR-SPACE序列中具有时间效率可行的实施方式,较sTx显著增强全脑信号与对比度,并在较大临床患者队列中表现一致且稳健。

研究设计与样本

  • 连续临床患者接受7T脑MRI检查,每个序列各有11例患者同时接受sTx与pTx扫描,比较临床队列每个序列各有40例患者接受pTx;采用变异系数(CoV)评估信号均匀性,采用跨皮层进入白质的8 mm线图像强度剖面评估组织对比度,并比较pTx与sTx序列的头部预测局部比吸收率(SAR)。

关键影像表现/征象

  • pTx较sTx在T2-SPACE与FLAIR-SPACE上均显著改善图像信号与对比度(P = .001)。
  • 颞叶白质与皮层之间的组织对比度在pTx下显著改善(P = .001)。
  • 较大临床队列中CoV未发现离群病例,信号均匀性表现一致。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: The adoption of routine clinical 7T MRI has been constrained by several challenges, with heterogeneity of the transmit field (B1+) being among the most notable. Dynamic parallel transmission (pTx) presents a promising strategy to enhance B1+ transmit homogeneity; however, associated technical challenges have limited its routine use. We assess performance of a prototype dynamic pTx implementation for the 3D sampling perfection with application-optimized contrasts by using different flip angle evolution (SPACE) sequence and hypothesize that signal homogeneity and tissue contrast will improve versus single transmit mode (sTx).

MATERIALS AND METHODS: Data from consecutive clinical patients undergoing 7T brain MRI for various indications were utilized. Signal homogeneity was assessed by using coefficient of variation (CoV). Tissue contrast was assessed by using an image intensity profile along an 8-mm line crossing the cortex into underlying white matter. Additionally, stability of signal homogeneity across a larger cohort of clinical patients was assessed with CoV. Predicted local specific absorption rate (SAR) for the head between pTx and sTx sequences was also compared.

RESULTS: For each sequence, 11 patients had both sTx and pTx scans. The comparison clinical cohort had 40 patients with pTx for each sequence. Image signal and contrast were significantly improved with pTx versus sTx for both T2-SPACE and FLAIR-SPACE (P = .001). Tissue contrast between white matter and cortex was also significantly improved in the temporal lobe with pTx (P = .001). CoV did not reveal any outlier cases across a large clinical cohort, demonstrating consistency in signal homogeneity. Despite increased SAR, T2-SPACE pTx consistently operated in first-level controlled mode, while FLAIR-SPACE pTx scans generally operated in normal mode.

CONCLUSIONS: We demonstrate the feasibility of a time-efficient prototype dynamic pTx implementation in T2-SPACE and FLAIR-SPACE sequences, which significantly enhances signal and contrast across the brain compared with sTx, while exhibiting consistent and robust performance in a large cohort of clinical patients.