基于临床评分与图像质量评估的AI加速T2加权脑MRI评价
本研究纳入100例连续患者,在同一次检查中采集深度学习重建的T2加权图像(T2(DR))与常规T2加权图像(T2(CN)),由四名神经放射科医生盲法评价图像质量、诊断安全性和伪影,并计算信噪比。结果显示两者在不可诊断评级的分布上无显著差异,但T2(CN)在整体图像质量和伪影方面显著更优;T2(DR)的有效信噪比(按采集时间校正)更高。结论提示T2(DR)可能适用于诊断场景,但质量不完全等同于高标准常规T2(CN)采集。
要点
- T2(DR)与T2(CN)在不可诊断评级分布上无显著差异(伪影p=0.108,整体质量p=0.700,诊断安全p=0.652)。
- 在全评级谱上T2(CN)的整体图像质量(p=0.003)和伪影(p<0.001)显著优于T2(DR)。
- T2(CN)信噪比更高(157.9 vs 112.8,p<0.001),但T2(DR)按采集时间校正的有效信噪比更高(14.1 vs 10.8,p<0.001)。
核心结论
- 结果提示T2(DR)可能适用于诊断场景,但其质量不完全等同于高标准的常规T2(CN)采集。
研究设计与样本
- 研究纳入100例连续患者,在同一次检查中于Siemens Vida 3 T扫描仪配合64通道头线圈上采集T2(DR)与T2(CN)两种图像。
关键影像表现/征象
- 由四名神经放射科医生盲法评价整体图像质量、诊断安全性和图像伪影;并计算空气、灰质、白质和脑脊液的信噪比及按采集时间校正的有效信噪比。
诊断标准与关键数值
- T2(CN)采集时间为3.33 min,T2(DR)采集时间为1.04 min;T2(CN)信噪比为157.9(SD 123.4),T2(DR)为112.8(SD 82.7),p<0.001。
临床应用与阅片要点
- 作者结论认为T2(DR)可能适用于诊断场景,但质量不完全等同于高标准常规T2(CN)采集。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To compare clinical ratings and signal-to-noise ratio (SNR) measures of a commercially available Deep Learning-based MRI reconstruction method (T2(DR)) against conventional T2- turbo spin echo brain MRI (T2(CN)).
MATERIALS AND METHODS: 100 consecutive patients with various neurological conditions underwent both T2(DR) and T2(CN) on a Siemens Vida 3 T scanner with a 64-channel head coil in the same examination. Acquisition times were 3.33 min for T2(CN) and 1.04 min for T2(DR). Four neuroradiologists evaluated overall image quality (OIQ), diagnostic safety (DS), and image artifacts (IA), blinded to the acquisition mode. SNR and SNReff (adjusted for acquisition time) were calculated for air, grey- and white matter, and cerebrospinal fluid.
RESULTS: The mean patient age was 43.6 years (SD 20.3), with 54 females. The distribution of non-diagnostic ratings did not differ significantly between T2(CN) and T2(DR) (IA p = 0.108; OIQ: p = 0.700 and DS: p = 0.652). However, when considering the full spectrum of ratings, significant differences favouring T2(CN) emerged in OIQ (p = 0.003) and IA (p < 0.001). T2(CN) had higher SNR (157.9, SD 123.4) than T2(DR) (112.8, SD 82.7), p < 0.001, but T2(DR) demonstrated superior SNReff (14.1, SD 10.3) compared to T2(CN) (10.8, SD 8.5), p < 0.001.
CONCLUSION: Our results suggest that while T2(DR) may be clinically applicable for a diagnostic setting, it does not fully match the quality of high-standard conventional T2(CN), MRI acquisitions.
引用
- Nonninger JN, Kienast P, Pogledic I, et al. Assessment of AI-accelerated T2-weighted brain MRI, based on clinical ratings and image quality evaluation. Eur J Radiol. 2025. PMID: 40315626. doi:10.1016/j.ejrad.2025.112123