光子计数探测器CT对脑膜中动脉的可视化:与能量积分探测器CT的比较

Visualization of the middle meningeal artery on photon-counting detector CT: Comparison with energy-integrating detector CT.
PubMed 文献 · Eur J Radiol · 2025-05-07

回顾性分析30例患者头颈CT血管造影图像,比较光子计数探测器CT超高分辨率模式下0.2mm层厚、1024×1024矩阵(PCD-1024)与512×512矩阵(PCD-512)及能量积分探测器CT 512×512矩阵(EID-512)对脑膜中动脉的可视化能力。定量与定性结果均显示PCD-1024对脑膜中动脉的可视化优于PCD-512和EID-512,且0.2mm重建在视觉上优于虚拟单能图像。

要点

  • PCD-1024的脑膜中动脉FWHM为1.41±0.17,小于PCD-512的1.48±0.15和EID-512的1.77±0.12。
  • PCD-1024的最大强度为407.5±71.7,高于PCD-512的386.0±72.5和EID-512的218.2±58.2。
  • 定性评分中PCD-1024的整体图像质量、锐利度和伪影评分均优于PCD-512和EID-512(P<0.05)。

核心结论

  • PCD-1024对脑膜中动脉的可视化优于PCD-512和EID-512,0.2mm重建在视觉上优于虚拟单能图像。

研究设计与样本

  • 回顾性分析2023年4月至2024年7月接受PCD-CT超高分辨率模式头颈CT血管造影的患者30例(男19例、女11例,中位年龄65岁),比较同患者PCD-1024、PCD-512与既往EID-512图像。

关键影像表现/征象

  • PCD-1024的FWHM为1.41±0.17,PCD-512为1.48±0.15,EID-512为1.77±0.12。
  • PCD-1024的最大强度为407.5±71.7,PCD-512为386.0±72.5,EID-512为218.2±58.2(PCD-1024 vs EID-512及PCD-512 vs EID-512的P均<0.001)。
  • 定性评分中PCD-1024、PCD-512、EID-512的整体图像质量分别为4(3-4)、3(3-4)、2(2-3),锐利度分别为4(3-4)、3(3-3)、2(2-3),伪影分别为3(3-3)、3(2-3)、2(2-2),组间差异P<0.05。

诊断标准与关键数值

  • 定量指标包括信噪比(SNR)、对比噪声比(CNR)、对比率(CR)、半高全宽(FWHM)及血管线轮廓最大强度。
  • PCD-1024、PCD-512、EID-512的SNR分别为16.9±5.6、20.3±6.2、23.2±7.8(PCD-1024 vs EID-512的P<0.01)。
  • PCD-1024、PCD-512、EID-512的CNR分别为16.6±4.5、20.2±5.5、44.4±13.9。
  • PCD-1024、PCD-512、EID-512的CR分别为0.74±0.06、0.74±0.06、0.68±0.08。
  • 定性评估采用四点Likert量表,评估整体图像质量、锐利度和伪影。

临床应用与阅片要点

  • 在需要显示脑膜中动脉等颅内细小动脉时,PCD-CT超高分辨率模式0.2mm层厚、1024×1024矩阵重建较512×512矩阵和EID-CT提供更好的图像锐利度和血管可视化。
  • 虚拟单能图像分析中45keV的SNR、CNR、CR和最大强度最高,但FWHM组间无差异;定性上0.2mm重建在所有评估方面优于虚拟单能图像(P<0.05)。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: The middle meningeal artery (MMA) is the target arteries for the treatment of intracranial diseases. The purpose of this study was to investigate the ability of photon-counting detector CT (PCD-CT) to improve the visualization of the MMA compared to energy-integrated detector CT (EID-CT).

METHODS: We retrospectively analyzed head CT angiography images from patients who underwent PCD-CT with ultra-high-resolution (UHR) mode between April 2023 and July 2024. We compared the images reconstructed with a slice thickness 0.2 mm and the matrix sizes 1024 × 1024 ('PCD-1024') and 512 × 512 ('PCD-512') with the same-patient images previously obtained by EID-CT with 0.75- or 1.0- mm slice thickness and the matrix size 512 × 512 ('EID-512'). Quantitatively, a region of interest (ROI) was placed on the maxillary artery near the origin of the MMA, and the signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and contrast ratio (CR) were measured. Image sharpness was evaluated by measuring the full width at half maximum (FWHM) and the maximum intensity from the MMA vessel line profile. Qualitatively, the overall image quality, sharpness, and artifacts were evaluated using four-point Likert scales. The same evaluation was performed using PCD-CT with 0.2 mm ('0.2 mm') and virtual monoenergetic images (VMIs) ('45 keV', '55 keV', and '70 keV').

RESULTS: 30 patients (19 males, 11 females; median age 65 years) were included. Quantitatively, the SNR (mean ± SD) for PCD-1024, PCD-512, and EID-512 were 16.9 ± 5.6, 20.3 ± 6.2, and 23.2 ± 7.8, respectively (P < 0.01 for PCD-1024 vs. EID-512). The CNR were 16.6 ± 4.5, 20.2 ± 5.5, and 44.4 ± 13.9. The CR were 0.74 ± 0.06, 0.74 ± 0.06, and 0.68 ± 0.08. The FWHM were 1.41 ± 0.17, 1.48 ± 0.15, and 1.77 ± 0.12, and the maximum intensity were 407.5 ± 71.7, 386.0 ± 72.5, and 218.2 ± 58.2 (P < 0.001 for PCD-1024 vs. EID-512 and PCD-512 vs. EID-512, respectively). Qualitatively, significant differences (P < 0.05) were identified among the PCD-1024, PCD-512, and EID-512 groups regarding the respective scores (median [IQR]) for overall image quality (4 [3-4], 3 [3-4], and 2 [2-3]), sharpness (4 [3-4], 3 [3-3], and 2 [2-3]), and artifacts (3 [3-3], 3 [2-3], and 2 [2-2]). In VMI analysis, quantitatively, 45 keV showed the highest values for SNR, CNR, CR, and maximum intensity; however, there were no differences in FWHM among the groups. Qualitatively, 0.2 mm was superior to the other groups in all evaluated aspects (P < 0.05).

CONCLUSION: PCD-1024 visualized the MMA better than both PCD-512 and EID-512. The 0.2 mm reconstruction was visually superior to the VMIs.

来源:Eur J Radiol,查看原文。版权归原作者所有。