腰椎金属植入物患者低剂量锡滤过CT联合迭代金属伪影减少算法与双能CT的对比

Low-dose tin-filtered computed tomography (CT) with iterative metal artifact reduction algorithm versus dual-energy CT for patients with lumbar metal implants.
PubMed 文献 · Eur J Radiol · 2025-04-04

本研究采用混合设计纳入70例腰椎金属植入物患者,前瞻性低剂量锡滤过CT组35例与回顾性双能CT组35例匹配,评估图像质量与辐射剂量。低剂量锡滤过CT有效剂量较双能CT降低26%(7.34±1.45 mSv vs. 9.93±2.00 mSv,P<0.05),客观与主观图像质量指标无显著差异。作者推荐该方案用于腰椎植入物患者的常规随访。

要点

  • 低剂量锡滤过CT联合iMAR有效剂量较双能CT iMAR降低26%(7.34±1.45 mSv vs. 9.93±2.00 mSv)。
  • 低剂量锡滤过CT与双能CT虚拟单能+在SNR、CNR、FOM上无显著差异(P=0.238-0.310)。
  • 主观图像质量及金属伪影评分两组间无统计学显著差异(P=0.120-0.763)。

核心结论

  • 低剂量锡滤过CT联合iMAR方案在图像质量与辐射剂量之间达到最佳平衡,推荐用于腰椎植入物患者的常规随访。

研究设计与样本

  • 混合设计研究,共纳入70例腰椎金属植入物患者,前瞻性低剂量锡滤过CT队列35例与回顾性双能CT队列35例匹配。
  • 评估包括双能CT混合图像、噪声优化虚拟单能成像130 keV(DE Mono+)以及低剂量锡滤过CT联合或不联合iMAR。
  • 客观评估指标为信噪比(SNR)、对比噪声比(CNR)和品质因数(FOM),主观评估采用5分评分系统评价整体图像质量、软组织显示、骨结构和金属伪影程度。
  • 记录辐射剂量指标包括CT剂量指数(CTDIvol)、剂量长度乘积(DLP)和有效剂量(ED)。

关键影像表现/征象

  • 低剂量锡滤过CT联合iMAR的主观评分范围为2.00至2.11,双能CT虚拟单能+联合iMAR评分范围为1.83至2.00(P=0.120-0.763)。
  • 整体图像质量、软组织显示、骨结构和金属伪影程度的主观评估组间仅见轻微差异,无统计学显著性。

诊断标准与关键数值

  • 低剂量锡滤过CT方案有效剂量为7.34±1.45 mSv,双能CT方案为9.93±2.00 mSv,低剂量锡滤过CT较双能CT降低26%(P<0.05)。
  • 低剂量锡滤过CT与DE Mono+在SNR、CNR和FOM上无显著差异(P=0.238-0.310)。

临床应用与阅片要点

  • 作者推荐低剂量锡滤过CT联合iMAR方案用于腰椎金属植入物患者的常规随访。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: This study aimed to evaluate a low-dose tin-filtered (LD Sn) CT protocol with iterative metal artifact reduction (iMAR) to reduce metal artifacts and radiation dose compared to dual-energy CT (DECT) with iMAR in patients with lumbar metal implants.

METHODS: The study included 70 patients, comprising 35 patients in the prospective LD Sn CT cohort and 35 matched patients in the retrospective DECT cohort, utilizing a hybrid design. Evaluations included DECT Mixed images, noise-optimized virtual monoenergetic imaging at 130 keV (DE Mono+), and LD Sn CT with and without iMAR. Objective assessments included the signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and figure of merit (FOM). A 5-point scoring system was used for subjective evaluations of image quality, including overall image quality, soft tissue visualization, bone architecture, and the degree of metal artifacts. Radiation dose metrics, including CT dose index (CTDIvol), dose length product (DLP), and effective dose (ED), were recorded.

RESULTS: The LD Sn CT protocol achieved a 26 % lower effective dose than the DECT protocol (7.34 ± 1.45 mSv vs. 9.93 ± 2.00 mSv, P < 0.05). No significant differences were observed in SNR, CNR, or FOM between LD Sn and DE Mono+ (P = 0.238-0.310). Subjective assessments, including overall image quality, soft tissue visualization, bone architecture, and the degree of metal artifact, showed slight differences without statistical significance. LD Sn iMAR scores ranged from 2.00 to 2.11, while DE Mono+ iMAR scores ranged from 1.83 to 2.00 (P = 0.120-0.763).

CONCLUSIONS: The LD Sn iMAR protocol achieved the optimal balance between image quality and radiation dose. It is recommended for routine follow-up in patients with lumbar implants.

来源:Eur J Radiol,查看原文。版权归原作者所有。