文献速递腹部技术与安全US

提高机械指数可改善剪切波弹性成像:信号增强的初步研究

Increased Mechanical Index Improves Shear Wave Elastography: Pilot Study of Signal Enhancement.
PubMed 文献 · Ultrasound Med Biol · 2025-04-09

这项初步研究在22名按体重指数分层的健康志愿者中,比较传统机械指数1.4与提高至2.5的推送脉冲对剪切波弹性成像信号质量的影响,发现较高机械指数使信噪比提高4.6 dB、变异系数降低13%,并提高检查成功率,尤其对体重指数超过30者,血清生物标志物未见生物效应或伤害迹象。

要点

  • 推送脉冲机械指数从1.4提高至2.5可改善剪切波弹性成像信号质量。
  • 较高机械指数使信噪比提高4.6 dB、变异系数降低13%。
  • 体重指数超过30的受试者检查成功率提升更明显。

核心结论

  • 提高推送脉冲机械指数至2.5可改善剪切波弹性成像的诊断效用,尤其对体重指数较高者,且未引入显著额外风险。

研究设计与样本

  • 初步研究纳入22名健康志愿者,按体重指数分层,分别采用传统机械指数1.4与提高至2.5的推送脉冲进行剪切波弹性成像,数据用传统剪切波弹性成像算法处理。

诊断标准与关键数值

  • 较高机械指数使信噪比提高4.6 dB(p<10-4,95%置信区间:3.4-5.8 dB)。
  • 较高机械指数使测量变异系数降低13%(p<10-4,95%置信区间:5.8%-20.3%)。
  • 较高机械指数提高剪切波弹性成像检查成功率,尤其对体重指数超过30的受试者。

临床应用与阅片要点

  • 作者认为提高推送脉冲机械指数至2.5可改善剪切波弹性成像的诊断效用,尤其对体重指数较高者。
  • 作者认为该方法可进一步增强剪切波弹性成像在人群规模慢性肝病监测中的重要作用。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Monitoring liver stiffness is essential for managing chronic liver disease, which poses a major public health challenge. Shear wave elastography (SWE), a non-invasive ultrasound-based technique, is commonly used to quantify liver stiffness. However, its performance can be compromised in individuals with higher body mass indices (BMIs) due to increased ultrasound absorption and distortion. Increasing the intensity of the ultrasound push beam could potentially improve signal quality, but regulatory limits currently restrict this due to safety concerns. This pilot study investigated the efficacy of increasing the push pulse mechanical index (MI) from a conventional value of 1.4 to 2.5 toward improving signal quality, and reducing measurement variability and failure rates.

METHODS: Healthy volunteers (N=22) stratified by BMI underwent SWE with conventional and increased MI push pulses. The resulting data were processed with conventional SWE algorithms, and the signal and measurement quality of the results were analyzed.

RESULTS: We found that the higher MI improved the signal-to-noise ratio by 4.6 dB (p<10-4, 95% confidence interval: 3.4-5.8 dB) and reduced the measurement's coefficient of variation by 13% (p<10-4, 95% confidence interval: 5.8%-20.3%), enhancing the success rate of SWE examinations, especially for subjects with a BMI over 30. Liver function tests before and after the SWE examinations showed no signs of bioeffects or harm based on serum biomarkers.

CONCLUSION: These results suggest that increasing the push pulse MI to 2.5 improves the diagnostic utility of SWE, particularly for individuals with a higher BMI, without introducing significant additional risk. This approach could further enhance SWE's vital role in the monitoring of chronic liver disease at a population scale.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Schoen S, Wang M, Dayavansha S, et al. Increased Mechanical Index Improves Shear Wave Elastography: Pilot Study of Signal Enhancement. Ultrasound Med Biol. 2025. PMID: 40204561. doi:10.1016/j.ultrasmedbio.2025.03.003

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。