基于钨对比剂的光子计数CT肝脏铁定量
在商用光子计数CT上,用两种或三种材料分解法对含碘或钨对比剂的肝脏体模进行铁定量。双材料分解无对比剂时铁定量准确度达1.4 mg/mL,钨使铁浓度高估,碘影响更大;三材料分解可同时分离铁、肝和钨,钨存在时准确度1.5–1.6 mg/mL。三材料分解有望在增强CT中检测和定量关键肝铁水平,钨对比剂优于碘对比剂。
要点
- 三材料分解可同时分离铁、肝和钨,钨存在时铁定量准确度1.5–1.6 mg/mL。
- 双材料分解在碘存在时铁定量准确度>18–>37 mg/mL,影响远大于钨。
- 双材料分解需严重病理才能检测出铁肝,三材料分解提示临床可行。
核心结论
- 基于4组光谱数据的三材料分解在增强CT扫描中提示可临床可行地检测和定量关键肝铁水平,钨对比剂优于碘对比剂。
研究设计与样本
- 在商用光子计数CT系统上进行实验,单次扫描可同时采集最多4组光谱数据;检查2个腹部体模,内含液体肝组织替代物、脂肪、铁及碘或钨对比剂样本,以模拟增强CT中不同肝脏组成;通过重建光谱CT图像的材料分解定量铁。
关键影像表现/征象
- 双材料分解基于2组光谱数据,在无对比剂样本的铁图像中提供铁和肝的材料图像,准确度1.4 mg/mL。
- 钨影响铁定量:材料样本中2和4 mgW/mL时,铁浓度被高估(P < 0.05),准确度分别为2.7和4.7 mg/mL。
- 三材料分解基于4组光谱数据,提供铁、肝和钨的材料图像,无钨图像准确度1.4 mg/mL,2和4 mgW/mL时铁图像准确度分别为1.5 mg/mL(无显著差异,P > 0.07)和1.6 mg/mL(高估,P > 0.03)。
- 碘对铁定量的影响大于钨:在双材料和三材料分解中,材料样本含2和4 mgI/mL时,测得铁浓度更高(P < 0.05),准确度分别>18和>37 mg/mL。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: This study explores the potential of quantifying hepatic iron in computed tomography (CT) scans in the presence of iodine (I)- or tungsten (W)-based contrast media (CM).
MATERIALS AND METHODS: Experimental work was performed on a commercial photon-counting CT system able to simultaneously acquire up to 4 spectral data sets in a single scan. We examined 2 anthropomorphic abdominal phantoms with material samples of liquid liver tissue surrogate, fat, iron, and I- or W-based CM to mimic different liver compositions in an enhanced CT scan. Iron was quantified by material decomposition of reconstructed spectral CT images.
RESULTS: Two-material decomposition based on 2 spectral data sets provided material images of iron and liver with an accuracy of 1.4 mg/mL in the iron image of CM-free samples. The presence of W affected the iron quantification: For 2 and 4 mgW/mL in the material samples, the iron concentration was overestimated ( P < 0.05) with accuracies of 2.7 and 4.7 mg/mL, respectively. Three-material decomposition based on 4 spectral data sets provided material images of iron, liver, and W, with an accuracy of 1.4 mg/mL in the images without W and 1.5 (nonsignificant difference, P > 0.07) and 1.6 mg/mL (overestimation, P > 0.03) in the iron image at 2 and 4 mgW/mL, respectively. The presence of I affected the iron quantification more than W in both 2- and 3-material decomposition: For 2 and 4 mgI/mL in the material samples, the measured iron concentration was even higher ( P < 0.05), with accuracies >18 and >37 mg/mL, respectively.
CONCLUSIONS: The accuracy of iron quantification from a 3-material decomposition suggested clinically feasible detection and quantification of critical hepatic iron levels in enhanced CT scans with suitable CM. In a 2-material decomposition, severe pathology is required to detect an iron liver. W-based CM was superior to I-based CM.
引用
- Neumann J, Haubold J, Jergas F, et al. Photon-Counting Computed Tomography-Based Hepatic iron Quantification Using a Tungsten-Based Contrast Agent. Invest Radiol. 2025. PMID: 40232874. doi:10.1097/RLI.0000000000001189
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。