抗散射滤线栅对体重低于10公斤儿童心导管检查辐射剂量和图像质量的影响
本研究通过模拟新生儿胸厚的体模和30例简单先天性心脏病婴儿的临床研究,评估移除抗散射滤线栅对辐射剂量和图像质量的影响。体模研究中移除滤线栅使剂量面积乘积和空气比释动能降低25.0%至55.6%;临床研究中无滤线栅时剂量面积乘积总和与空气比释动能总和均显著低于有滤线栅时。结论认为移除抗散射滤线栅不影响检查准确性并降低辐射剂量,可能对简单先天性心脏病小患者有益。
要点
- 无抗散射滤线栅时剂量面积乘积总和1.22±0.58 Gy·cm2,显著低于有滤线栅的1.83±0.64 Gy·cm2(P=0.01)。
- 无抗散射滤线栅时空气比释动能总和16.22±6.98 mGy,显著低于有滤线栅的23.23±8.33 mGy(P=0.02)。
- 体模研究中移除抗散射滤线栅使剂量面积乘积和空气比释动能降低25.0%至55.6%。
核心结论
- 移除抗散射滤线栅不影响检查准确性,并在体模和临床研究中均实现更低的辐射剂量;无抗散射滤线栅的心导管检查可能对简单先天性心脏病小患者是有益选择。
研究设计与样本
- 包含模拟新生儿胸厚的自制体模初步研究和30例简单先天性心脏病婴儿的临床研究,评估剂量面积乘积、空气比释动能及视觉与物理图像质量评估。
关键影像表现/征象
- 无抗散射滤线栅的透视模式下,较长的源-图像距离改善了导丝和造影导管的对比噪声比。
临床应用与阅片要点
- 作者认为无抗散射滤线栅的心导管检查可能对简单先天性心脏病小患者是有益选择。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Anti-scatter grids are commonly used for cardiac catheterization in children with heart disease. However, the effect of the grid on the radiation dose or image quality has not been sufficiently demonstrated.
OBJECTIVE: To investigate the effect of the removal of an anti-scatter grid on radiation dose and image quality in phantom and clinical settings.
MATERIALS AND METHODS: This research consists of a preliminary study using a manufactured phantom simulating a newborn's chest thickness and a clinical study of 30 infants with simple congenital heart disease. We evaluated the exposure dose by dose area product (DAP), air kerma (AK), and image quality using visual and physical assessments.
RESULTS: In the phantom study, removal of the anti-scatter grid decreased DAP and AK by 25.0% to 55.6%. In the clinical study, the DAP (sum) and AK (sum) without an anti-scatter grid were significantly lower than with the grid (1.22±0.58 Gy·cm2 vs. 1.83±0.64 Gy·cm2, P=0.01; 16.22±6.98 mGy vs. 23.23±8.33 mGy, P=0.02, respectively). In fluoroscopy mode without an anti-scatter grid, longer source-image-distances improved the contrast-to-noise ratio for guidewires and angiographic catheters. The accuracy of the examination was not affected in the clinical setting.
CONCLUSIONS: The removal of the anti-scatter grid did not affect the accuracy of the examination and achieved a lower radiation dose in both phantom and clinical studies. Cardiac catheterization without an anti-scatter grid may be a beneficial option for small patients with simple congenital heart disease.