文献速递胸部技术与安全CT

手臂无法上举患者胸部CT的辐射剂量降低与图像质量提升:一项对比研究

Radiation dose reduction and image quality enhancement for patients unable to elevate their arms in chest CT: A comparative study.
PubMed 文献 · Eur J Radiol · 2025-04-17

研究对比手臂上举(A组)、使用自制气隙装置手臂下垂(B组)及无装置手臂下垂(C组)胸部CT的辐射剂量与图像质量,B组DLP较C组降低33.0%且图像质量与A组相似,作者认为气隙装置可在维持图像质量的同时降低辐射剂量。

要点

  • 手臂无法上举者使用气隙装置行胸部CT,DLP较无装置手臂下垂降低33.0%。
  • B组图像质量与A组相似,CNRDR为30.2±11.1 vs 26.6±10.9(p=0.046)。
  • 体重超过70 kg者三种体位DLP均超过胸部CT诊断参考水平(324.2 mGy×cm)。

核心结论

  • 手臂无法上举患者行胸部CT时,使用气隙装置手臂下垂可降低辐射剂量并维持图像质量。
  • 定制装置有助于安全高效的检查流程,尤其对体重较高的患者。

研究设计与样本

  • 对比研究,胸部CT采集三种体位:手臂上举(对照组,A组)、使用自制气隙装置手臂下垂(B组)、无装置手臂下垂(C组)。
  • 数据按体重分类,辐射剂量以剂量长度乘积(DLP)评估,定量分析图像噪声、对比噪声比(CNR)及CNR与剂量比(CNRDR),定性评估采用Likert量表。

关键影像表现/征象

  • C组平均DLP为473.7±142.9 mGy×cm,A组最低为267.3±93.3 mGy×cm。
  • B组DLP为317.2±121.4 mGy×cm,较A组高18.7%,较C组低33.0%。
  • 体重超过70 kg患者中,所有体位DLP均超过胸部CT诊断参考水平(324.2 mGy×cm)。
  • 体重低于70 kg(即<50 kg、50-70 kg)者中,B组DLP低于上述诊断参考水平。
  • 定性评估A组评分最高为4.96,B组为4.51,C组为3.47。

诊断标准与关键数值

  • 定量分析示B组图像质量与A组相似,CNRDR为30.2±11.1 vs A组26.6±10.9,差异边缘显著(p=0.046)。
  • 胸部CT诊断参考水平(DRL)为324.2 mGy×cm,体重超过70 kg者各体位DLP均超过该水平。

临床应用与阅片要点

  • 对无法上举手臂的患者,采用气隙装置手臂下垂行胸部CT可在维持图像质量的同时降低辐射剂量。
  • 体重超过70 kg者即使使用气隙装置,DLP仍超过诊断参考水平,需关注辐射剂量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To propose a method to effectively reduce radiation dose while enhancing image quality in chest CT for patients unable to elevate their arms utilizing the air-gap technique.

METHODS: Chest CT images were acquired in three positions: arms raised (control group, position A), arms lowered with the air-gap technique utilizing an in-house device (position B), and arms lowered without the device (position C). Data were categorized by body weight. Radiation dose was assessed using dose-length product (DLP) values, with quantitative analyses of image noise, contrast-to-noise ratio (CNR), and CNR-to-dose ratio (CNRDR). Qualitative assessment utilized a Likert scale.

RESULTS: Average DLP was 473.7 ± 142.9 mGy × cm in position C, with the lowest in position A at 267.3 ± 93.3 mGy × cm. Position B had a DLP of 317.2 ± 121.4 mGy × cm, 18.7 % higher than position A but 33.0 % lower than position C. In patients over 70 kg, all positions exceeded the thoracic CT diagnostic reference level (DRL) (324.2 mGy × cm). For those under 70 kg (i.e., <50 kg, 50-70 kg), position B showed lower DLPs compared to the above DRL. Quantitative analysis indicated that position B maintained image quality similar to position A, with CNRDR of 30.2 ± 11.1 versus 26.6 ± 10.9 in position A, showing marginal significance (p = 0.046). The qualitative evaluation indicated that position A had the highest rating at 4.96, followed by position B at 4.51, and position C at 3.47.

CONCLUSION: The air-gap device with lowered arms during chest CT for patients unable to elevate their arms can reduce radiation dose while maintaining image quality. A custom-designed device facilitates safe and efficient procedures, especially for patients with higher body weight.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Lee KB, Kim SM, Yoon JT, et al. Radiation dose reduction and image quality enhancement for patients unable to elevate their arms in chest CT: A comparative study. Eur J Radiol. 2025. PMID: 40286552. doi:10.1016/j.ejrad.2025.112120

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。