文献速递心血管外周血管MR

商用宽孔径0.55-T系统上非对比增强三维MR血管成像的可行性:与1.5-T MR血管成像的比较

Feasibility of Noncontrast 3D MR Angiography on a Commercial Wide-Bore 0.55-T System: Comparison with 1.5-T MR Angiography.
PubMed 文献 · Radiol Cardiothorac Imaging · 2025-01-01

研究在宽孔径0.55-T商用低场系统上评估心电图触发、导航门控、三维平衡稳态自由进动非对比增强MR血管成像研究序列,并与1.5-T图像比较图像质量和主动脉尺寸测量。10名健康志愿者和10名扩张性胸主动脉受试者均接受两种场强扫描;0.55-T图像质量足够,定量测量和诊断准确性与1.5-T无显著差异。

要点

  • 0.55-T非对比增强三维平衡稳态自由进动MRA图像质量中位评分与1.5-T相当。
  • 0.55-T与1.5-T的主动脉尺寸测量组内和组间一致性均极好(ICC>0.9)。
  • 扩张性胸主动脉诊断在两种场强间无显著差异,0.55-T可用于评估。

核心结论

  • 0.55-T非对比增强三维平衡稳态自由进动MRA图像质量足够,定量测量和诊断准确性与1.5-T相比无显著差异。

研究设计与样本

  • 研究纳入10名健康志愿者(28.8岁±9.0;4名男性)和10名评估扩张性胸主动脉的受试者(53.6岁±10.7;6名男性)。
  • 所有受试者在1.5-T和0.55-T宽孔径商用低场系统上均接受非对比增强心电图触发、导航门控、三维平衡稳态自由进动MRA研究序列扫描。
  • 采用Wilcoxon检验、组内相关系数和Bland-Altman图评估两组受试者不同场强间的整体图像质量、主动脉尺寸评定者内和评定者间一致性及诊断准确性。

关键影像表现/征象

  • 志愿者和受试者的图像质量中位评分在不同场强间保持相当。
  • 志愿者升主动脉(P=.01)和受试者升主动脉(P=.02)的血流信号变异性在1.5-T更大。
  • 志愿者和受试者升主动脉和近端主动脉弓的血-心肌对比度在1.5-T显著更高(P<.05)。

诊断标准与关键数值

  • 0.55-T和1.5-T的评定者内和评定者间一致性均极好(组内相关系数>0.9)。
  • 主动脉尺寸测量组内相关系数>0.9,主动脉扩张诊断未显示显著场强间差异。

临床应用与阅片要点

  • 在宽孔径0.55-T商用低场系统上,可采用非对比增强三维平衡稳态自由进动MRA评估胸主动脉尺寸和扩张。
  • 阅片时注意1.5-T在升主动脉和近端主动脉弓的血-心肌对比度显著更高、升主动脉血流信号变异性更大。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Purpose To evaluate a noncontrast electrocardiographically triggered, navigator-gated, three-dimensional, balanced steady-state free precession MR angiography (MRA) research sequence on a wide-bore 0.55-T commercial low-field-strength system, by comparing image quality and aortic dimension measurements against images obtained at 1.5 T. Materials and Methods Ten healthy volunteers (28.8 years ± 9.0 [SD]; four male) and 10 participants being evaluated for dilated thoracic aorta (53.6 years ± 10.7; six male) underwent noncontrast MRA scans at both 1.5 T and 0.55 T. Overall image quality, intrarater and interrater agreement in aortic dimensions, and diagnostic accuracy were evaluated between field strengths for both groups of individuals using Wilcoxon test, intraclass correlation coefficient, and Bland-Altman plots. Results Median image quality scores remained comparable between volunteers and participants across field strengths. Blood signal variability was greater at the ascending aorta at 1.5 T for volunteers (P = .01) and participants (P = .02). Blood-myocardium contrast was significantly higher at the ascending aorta and proximal arch at 1.5 T for volunteers and participants (P < .05). Excellent intra- and interrater agreement (intraclass correlation coefficient > 0.9) was demonstrated at 0.55 T and 1.5 T. Aortic dimensions (intraclass correlation coefficient > 0.9) and diagnosis of aortic dilatation did not demonstrate significant interfield differences. Conclusion Noncontrast, three-dimensional, balanced steady-state free precession MRA at 0.55 T demonstrated adequate image quality and no significant differences in quantitative measurements or diagnostic accuracy compared with 1.5 T. Keywords: Noncontrast MR Angiography, Low Field, 0.55 T, Obesity, Thoracic Aorta © RSNA, 2025.