文献速递胸部技术与安全CT

胸部定量CT:双能虚拟平扫与真实平扫的区域差异

Quantitative chest computed tomography: regional differences in dual-energy-derived virtual vs. true non-contrast scans.
PubMed 文献 · Clin Radiol · 2025-03-26

这项双中心回顾性研究纳入48例患者,比较对比增强双能CT生成的虚拟平扫(VNC)与真实平扫(TNC)在肺组织定量分析中的一致性,重点关注区域差异。全肺平均肺密度、低衰减容积和第15百分位数的组内相关系数均大于0.9,高衰减容积的一致性稍低(ICC 0.84)。上、中、下三分之一的平均肺密度平均差异分别为-5.71 HU、1.62 HU和5.77 HU,虽存在轻微区域差异,但VNC图像可可靠用于定量分析。

要点

  • VNC与TNC在全肺平均肺密度、低衰减容积及第15百分位数上高度一致(ICC均大于0.9)。
  • 高衰减容积的一致性稍低,ICC为0.84(95%CI:0.73,0.91)。
  • 肺密度存在区域差异,上、中、下三分之一平均差异分别为-5.71 HU、1.62 HU和5.77 HU。

核心结论

  • 尽管观察到轻微的区域差异,使用VNC图像可可靠地进行定量分析。

研究设计与样本

  • 双中心回顾性、IRB批准的研究,纳入2018年4月至2022年12月间一年内接受对比增强双能CT和平扫胸部CT的连续患者48例(26例女性,中位年龄63岁,四分位距58-69岁),图像共配准后半手动分割为全肺、中心/外周、腹侧/背侧半及上/中/下三分之一等感兴趣区。

关键影像表现/征象

  • 全肺平均肺密度、低衰减容积和第15百分位数的Pearson相关系数和组内相关系数均大于0.9,显示高度相关和一致性。
  • 高衰减容积的一致性稍低,组内相关系数为0.84(95%CI:0.73,0.91)。
  • 平均肺密度的平均差异(HUVNC减HUTNC)在上、中、下三分之一分别为-5.71 HU(95%CI:-11.07,-0.35)、1.62 HU(95%CI:-2.98,6.23)和5.77 HU(95%CI:-0.47,12.01)。

诊断标准与关键数值

  • 定量参数包括平均肺密度、高衰减容积、低衰减容积和第15百分位数,在VNC和TNC图像上对每个感兴趣区计算。
  • 采用Pearson相关系数、组内相关系数和Bland-Altman分析评估相关性和一致性,亚组分析涉及体重指数和肺气肿标志物。

临床应用与阅片要点

  • VNC图像可用于胸部CT定量分析,但需注意高衰减容积的一致性稍低(ICC 0.84)。
  • 肺密度在上、中、下三分之一存在区域差异,上三分之一VNC较TNC低5.71 HU,下三分之一VNC较TNC高5.77 HU,阅片和报告时应考虑区域差异。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

AIM: Quantitative analysis of chest computed tomography (CT) usually requires non-contrast scans. Multi-energy CT-derived virtual non-contrast (VNC) images may obviate additional true non-contrast (TNC) scans. We aimed to quantitatively compare virtual and true non-contrast images, with a focus on regional differences in pulmonary tissue.

MATERIALS AND METHODS: This bi-center retrospective, IRB-approved study included consecutive patients who received contrast-enhanced dual-energy and non-contrast chest CT scans within one year between April 2018 and December 2022. Scans were co-registered and semi-manually segmented into various volumes of interest: whole lung, center/periphery, ventral/dorsal half and upper/middle/lower third. The mean lung density, high and low attenuation volumes and the lower 15th percentile were computed from the VNC and TNC scans for each volume. Correlation and agreement were assessed using Pearson's and intraclass correlation coefficients and Bland-Altman analysis. Subgroup analyses involved body mass index and markers of pulmonary emphysema.

RESULTS: Forty-eight patients were included (26 females, median age: 63 years, interquartile range: 58-69). Pearson's and intraclass correlation coefficients > 0.9 demonstrated high correlation and agreement for the mean lung density, low attenuation volume and the lower 15th percentile across the whole lung. The agreement for high attenuation volume was slightly less pronounced (intraclass correlation coefficient, 0.84 [95% confidence interval: 0.73, 0.91]). Regional differences were observed, for example, concerning the mean lung density with mean differences (HUVNC minus HUTNC) of -5.71 HU [-11.07, -0.35], 1.62 HU [-2.98, 6.23] and 5.77 HU [-0.47, 12.01] for the upper, middle and lower third, respectively.

CONCLUSION: Although slight regional differences were observed, quantitative analysis can be reliably performed using VNC images.

来源:Clin Radiol,查看原文。版权归原作者所有。