研究在6名健康志愿者口服0.8 g/kg [6,6']-2H-葡萄糖后,于7T采用密度加权bSSFP结合快速三维同心环轨迹采样行全脑动态氘代谢成像,bSSFP-CRT的SNR较传统spoiled FID-CRT 2H-MRSI最高提高3倍,使空间分辨率提高2倍,并在0.75 mL与0.36 mL各向同性分辨率下70分钟时检出相当的2H-Glc、2H-Glx和2H-水浓度。

要点

  • bSSFP-CRT较传统spoiled FID-CRT 2H-MRSI的SNR最高提高3倍,空间分辨率提高2倍。
  • 口服2H-Glc后约70分钟,两种采集方案在0.75 mL和0.36 mL分辨率下检出相当的2H-Glc、2H-Glx和2H-水浓度。
  • CGM间质液Glc平均AUC为509±65 mM·min,为DMI脑Glc平均AUC 149±43 mM·min的3.4倍。

核心结论

  • 本研究在7T成功实现bSSFP采集方案用于全脑动态人脑DMI,较传统spoiled采集SNR提高3倍,使空间分辨率较传统FID-CRT DMI提高2倍;系统连续血糖测量结合动态DMI显示出临床应用的显著潜力,可以微创且生理的方式将脑葡萄糖代谢与时间分辨的外周葡萄糖水平联系起来。

研究设计与样本

  • 6名健康志愿者(2女/4男)口服0.8 g/kg [6,6']-2H-Glc后,于7T采用新型密度加权bSSFP采集结合快速3D同心环轨迹k空间采样行全脑动态DMI。

关键影像表现/征象

  • 口服2H标记Glc后约80分钟内采集时间分辨全脑DMI数据集,每个数据集7分钟,空间分辨率为0.75 mL和0.36 mL各向同性。
  • bSSFP-CRT的SNR较传统spoiled FID-CRT 2H-MRSI最高提高3倍,空间分辨率提高2倍。
  • 口服示踪剂后70分钟,两种采集方案在常规和高空间分辨率(0.75 mL和0.36 mL各向同性)下检出相当的2H-Glc、2H-Glx和2H-水浓度。

诊断标准与关键数值

  • CGM传感器测得间质液Glc平均AUC为509±65 mM·min,DMI测得脑Glc平均AUC为149±43 mM·min,前者为后者的3.4倍。
  • 线性混合模型评估CGM与脑2H-Glc关系,灰质斜率β1=0.47(P=0.01),白质斜率β1=0.36(P=0.03),均达统计学显著。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Deuterium ( 2 H) metabolic imaging (DMI) is an emerging magnetic resonance technique to non-invasively map human brain glucose (Glc) uptake and downstream metabolism following oral or intravenous administration of 2 H-labeled Glc. The achievable spatial resolution is limited due to inherently low sensitivity of DMI. This hinders potential clinical translation. The purpose of this study was to improve the signal-to-noise ratio (SNR) of 3D DMI via a balanced steady-state free precession (bSSFP) acquisition scheme combined with fast non-Cartesian spatial-spectral sampling to enable high-resolution dynamic imaging of neural Glc uptake and glutamate+glutamine (Glx) synthesis of the human brain at 7T.

MATERIALS AND METHODS: Six healthy volunteers (2 f/4 m) were scanned after oral administration of 0.8 g/kg [6,6']- 2 H-Glc using a novel density-weighted bSSFP acquisition scheme combined with fast 3D concentric ring trajectory (CRT) k-space sampling at 7T. Time-resolved whole brain DMI datasets were acquired for approximately 80 minutes (7 minutes per dataset) after oral 2 H-labeled Glc administration with 0.75 mL and 0.36 mL isotropic spatial resolution and results were compared to conventional spoiled Free Induction Decay (FID) 2 H-MRSI with CRT readout at matched nominal spatial resolution. Dynamic DMI measurements of the brain were accompanied by simultaneous systemic Glc measurements of the interstitial fluid using a continuous Glc monitoring (CGM) sensor (on the upper arm). The correlation between brain and interstitial Glc levels was analyzed using linear mixed models.

RESULTS: The bSSFP-CRT approach achieved SNRs that were up to 3-fold higher than conventional spoiled FID-CRT 2 H-MRSI. This enabled a 2-fold higher spatial resolution. Seventy minutes after oral tracer uptake comparable 2 H-Glc, 2 H-Glx, and 2 H-water concentrations were detected using both acquisition schemes at both, regular and high spatial resolutions (0.75 ml and 0.36 mL isotropic). The mean Areas Under the Curve (AUC) for interstitial fluid Glc measurements obtained using a CGM sensor was 509 ± 65 mM·min. This is 3.4 times higher than the mean AUC of brain Glc measurements of 149 ± 43 mM·min obtained via DMI. The linear mixed models fitted to assess the relationship between CGM measures and brain 2 H-Glc yielded statistically significant slope estimates in both GM (β 1 = 0.47, P = 0.01) and WM (β 1 = 0.36, P = 0.03).

CONCLUSIONS: In this study we successfully implemented a balanced steady-state free precession (bSSFP) acquisition scheme for dynamic whole-brain human DMI at 7T. A 3-fold SNR increase compared to conventional spoiled acquisition allowed us to double the spatial resolution achieved using conventional FID-CRT DMI. Systemic continuous glucose measurements, combined with dynamic DMI, demonstrate significant potential for clinical applications. This could help improve our understanding of brain glucose metabolism by linking it to time-resolved peripheral glucose levels. Importantly, these measurements are conducted in a minimally invasive and physiological manner.

来源:Invest Radiol,查看原文。版权归原作者所有。