楔形滤过器在诊断性脑血管造影中降低晶状体辐射剂量的有效性:体模与前瞻性研究

Effectiveness of wedge filter application in reducing lens radiation dose during diagnostic cerebral angiography: a phantom and prospective study.
PubMed 文献 · Eur Radiol · 2025-04-26

本研究通过体模实验确定楔形滤过器的最佳深度,并前瞻性纳入20例颅内动脉瘤患者,对一侧颈内动脉使用楔形滤过器降低晶状体剂量方案、对侧使用标准方案。结果显示楔形滤过器使晶状体剂量降低47.2%(中位数1.06 mGy vs 0.56 mGy,p < 0.001),且图像噪声与质量无显著差异。

要点

  • 楔形滤过器使脑血管造影中晶状体剂量降低47.2%,图像质量无显著差异。
  • 体模中前后位投照4 cm深度剂量降低25.5%,侧位3 cm深度降低28.8%。
  • 两种方案下左晶状体剂量均显著高于右晶状体。

核心结论

  • 楔形滤过器可显著降低脑血管造影中的晶状体辐射剂量,且不影响图像质量。

研究设计与样本

  • 体模实验确定楔形滤过器最佳深度,随后前瞻性纳入20例颅内动脉瘤患者,同一患者一侧颈内动脉使用楔形滤过器降低剂量方案、对侧使用标准方案,用光致发光玻璃剂量计测量晶状体剂量。

诊断标准与关键数值

  • 楔形滤过器降低方案 vs 标准方案:晶状体剂量中位数0.56 mGy vs 1.06 mGy,降低47.2%,p < 0.001。
  • 标准方案:左晶状体剂量1.37 mGy vs 右晶状体0.70 mGy,p < 0.001。
  • 降低方案:左晶状体剂量0.75 mGy vs 右晶状体0.36 mGy,p < 0.001。
  • 体模实验:前后位投照4 cm深度剂量降低25.5%,侧位投照3 cm深度降低28.8%。

临床应用与阅片要点

  • 在常规神经血管造影中可使用楔形滤过器降低晶状体辐射暴露,操作简单且保持图像质量。
  • 使用楔形滤过器时左右晶状体剂量均显著低于标准方案,左晶状体剂量在两种方案下均高于右侧。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Diagnostic cerebral angiography poses a risk of direct radiation exposure to the eye lens. This study aimed to evaluate the effectiveness of using a wedge filter to reduce lens exposure during examination.

METHODS: A phantom study was initially conducted to evaluate the efficacy and determine the optimal depth of the wedge filter. Twenty patients with intracranial aneurysms scheduled for diagnostic cerebral angiography were prospectively enrolled. For each patient, the lens dose reduction protocol (involving a wedge filter) was used on one internal carotid artery (ICA), whereas the standard protocol was applied to the other ICA. Photoluminescent glass dosimeters were used to measure the lens dose. Quantitative noise measurements and qualitative analysis were performed to assess image quality.

RESULTS: The phantom study demonstrated that the wedge filter reduced the radiation dose to the eye lens in anteroposterior (25.5% reduction at 4 cm depth) and lateral projections (28.8% reduction at 3 cm depth). In the patient study (20 patients, 60% female, mean age 57.5 years), the lens dose reduction protocol reduced the dose by 47.2% compared with the standard protocol (median 1.06 mGy vs 0.56 mGy, p < 0.001). In both protocols, the left lens dose was significantly higher than the right (standard: 1.37 mGy vs 0.70 mGy, p < 0.001; reduction: 0.75 mGy vs 0.36 mGy, p < 0.001). No significant differences were observed in image noise or quality.

CONCLUSION: The wedge filter significantly reduced the lens radiation dose during cerebral angiography without affecting image quality.

KEY POINTS: Question The radiation dose to the eye lens during cerebral angiography remains unclear; a targeted method is needed to selectively reduce exposure to this radiosensitive organ. Findings This study directly measured eye lens radiation dose and found that using a wedge filter reduces exposure by nearly half while maintaining diagnostic image quality. Clinical relevance Using a wedge filter during routine neuroangiography effectively reduces eye lens radiation exposure with a simple operation, preserving image quality and potentially lowering the risk of radiation-induced cataracts.

来源:Eur Radiol,查看原文。版权归原作者所有。