指南与共识头颈技术与安全CT

牙科锥形束CT:临床建议与适应证相关技术特征

Cone beam computed tomography in dentistry: Clinical recommendations and indication-specific features.
指南 / 共识 · J Dent · 2025-04-23

本文综述了商用CBCT设备的技术特征及其对图像质量和诊断性能的影响,基于文献与厂商资料汇总了来自11个国家34家厂商的87款设备,并针对口腔各亚专科常见适应证开发了决策支持图表,以指导临床选择视野、体素大小、空间分辨率及曝光参数等采集与重建参数。

要点

  • 管电压范围50–120 kV,管电流多在1–17 mA之间。
  • 剂量面积乘积范围10–5600 mGy·cm²,典型有效剂量估计为3–500 µSv。
  • 决策支持图表覆盖修复与种植、牙周、口腔颌面外科、儿童牙科与正畸及牙髓病学。

核心结论

  • 本文概述了商用CBCT设备的技术特征,指出重建参数与剂量考虑方面存在信息缺口,强调应根据患者具体需求个体化调整CBCT采集,以优化诊断准确性并降低生物学风险。

研究设计与样本

  • 叙述性综述,通过检索已发表文献、访问厂商网站、参加2023年至2024年的学术与商业活动,并查询网站与手册直至2025年2月,汇总商用CBCT设备信息,对数据进行描述性与定量分析,并为各牙科亚专科开发决策支持图表。

诊断标准与关键数值

  • 共识别来自11个国家34家厂商的87款商用CBCT设备,但仅10家公司的信息得到确认。
  • 视野与体素大小为最常报告的参数,缺失数据因参数而异。
  • 管电压范围为50–120 kV,管电流主要在1–17 mA之间变化。
  • 剂量面积乘积范围为10–5600 mGy·cm²,典型有效剂量估计为3–500 µSv。
  • 决策支持图表列出修复与种植牙科、牙周病学、口腔颌面外科、儿童牙科与正畸及牙髓病学所需的图像采集与重建相关参数。

临床应用与阅片要点

  • 作者认为基于患者具体需求的定制化建议至关重要,临床医生应个体化CBCT采集以优化诊断准确性并最小化生物学风险。
  • 决策支持图表可帮助临床医生针对不同牙科亚专科选择最佳技术参数。

局限性

  • 仅10家公司的信息得到确认,缺失数据因参数而异,重建参数与剂量考虑方面存在信息缺口。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To provide a narrative overview of commercially available CBCT devices, highlight the wide range of technical features and their impact on image quality and diagnostic performance, and develop decision support charts to guide clinicians in selecting appropriate CBCT parameters for different dental specialties.

DATA AND SOURCES: Data collection involved reviewing published articles on CBCT devices and accessing manufacturer websites. Additional manufacturers were identified through a literature search and by attending scientific and commercial events held in 2023 and 2024. Information on CBCT features was obtained from websites, manuals, and inquiries until February of 2025. Data were categorized and analyzed descriptively and quantitatively to recommend device requirements for dental specialties. Decision support charts were developed for each specialty based on common indications and technical parameters (e.g., field-of-view, scout need, spatial resolution, X-ray exposure parameters, and post-processing tools).

RESULTS: Eighty-seven commercially available CBCT devices from 34 manufacturers across 11 countries were identified, with information confirmed for only 10 companies. Missing data varied, however FOV and voxel size were commonly reported. Tube voltage ranged from 50-120 kV, and tube current mostly varied between 1-17 mA. Dose considerations varied widely, with dose-product-area ranging from 10 to 5600 mGy.cm2 and typical effective dose estimates from 3 to 500 µSv. Decision support charts were created, outlining necessary image acquisition and reconstruction-related parameters for prosthodontics and implant dentistry, periodontology, oral and maxillofacial surgery, pediatric dentistry and orthodontics, and endodontics.

CONCLUSION: This study provides an overview of the technical features of commercially available CBCT devices, highlighting gaps in reconstruction parameters and dose considerations. Tailored recommendations based on patient-specific needs are essential, with clinicians individualizing CBCT acquisition to optimize diagnostic accuracy and minimize biological risks.

CLINICAL SIGNIFICANCE: This study provides valuable insights into commercially available CBCT devices in terms of their features and provides decision support chart to help clinicians select optimal technical parameters tailored to specific dental specialties.

来源:J Dent,查看原文。版权归原作者所有。

引用

  1. Fontenele RC, Gaêta-Araujo H, Jacobs R, et al. Cone beam computed tomography in dentistry: Clinical recommendations and indication-specific features. J Dent. 2025. PMID: 40280537. doi:10.1016/j.jdent.2025.105781

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。